Continuous thermodilution cardiac output measurement in sheep
M Yelderman, MD Quinn, RC McKown, RC Eberhart and ML Dollar
InterFlo Medical, Inc., Plano, Tex. 75074.
A technique has been developed to continuously measure cardiac output by
means of the principles of thermodilution. Pulmonary artery catheters were
modified by placing a 10 cm filament near the usual injectate port. Small
amounts of heat were infused according to a randomly repeating binary
on-off sequence. The distal blood temperature was recorded and
cross-correlated with the heat waveform to produce a dilution curve and
calculate cardiac output. The technique was compared with bolus
thermodilution in seven sheep. Cardiac output ranged from 1.5 to 13.2
L/min, and heart rate varied from 59 to 180 beats/min. The linear
regression between the data obtained by the two methods is represented by
the equation y = 1.00x + 0.13; the correlation coefficient, R, is 0.97, and
the p value is less than 0.0001.