J Thorac Cardiovasc Surg 2002;123:783-787
© 2002 The American Association for Thoracic Surgery
Evolving Technology (ET) |
From the Division of Cardiothoracic Surgery, The Ohio State University Medical Center, Columbus, Ohio.
Received for publication July 12, 2001. Revisions requested Aug 28, 2001; revisions received Sept 10, 2001. Accepted for publication Sept 12, 2001. Address for reprints: Randall K. Wolf, MD, Division of Cardiothoracic Surgery, The Ohio State University Medical Center, 410 W 10th Ave, Doan HallRoom N816, Columbus OH 43210 (E-mail: wolf-4{at}medctr.osu.edu).
Objective: We are in the midst of development of several new anastomotic devices for use in coronary artery bypass grafting surgery. This study was designed to examine one of these devices (a new self-closing clip) for left internal thoracic artery-left anterior descending coronary artery anastomosis. Its feasibility and the quality of anastomosis were evaluated.
Methods: Fourteen patients who underwent first-time elective coronary artery bypass surgery were enrolled between July and December 2000. The left internal thoracic artery was anastomosed to the left anterior descending coronary artery in an interrupted manner with Coalescent U-clips (Coalescent Surgical, Inc, Sunnyvale, Calif). Immediate patency was checked with a transit-time flowmeter. Selective angiography was performed 6 months after surgical intervention.
Results: Five patients underwent on-pump coronary bypass grafting, 9 on the beating heart. One patient was excluded from the study intraoperatively because of a poor target site necessitating a 2-cm-long anastomosis. Left internal thoracic artery-left anterior descending artery anastomoses were created with an average of 11.8 clips in 15.9 minutes. Mean graft flow was 45.6 mL/min. Neither conversion to standard suture technique nor revision of anastomosis was necessary. There tended to be a learning curve in the anastomosis on the beating heart. Postoperative lengths of stay in the intensive care unit and the hospital were 20.7 hours and 3.9 days, respectively. Neither death nor major complication was seen, except for temporary atrial fibrillation in 2 patients. Graft patency at 6 months was 100% (FitzGibbon grade A).
Conclusion: Left internal thoracic arteryleft anterior descending artery anastomoses can be created safely and effectively with new self-closing clips on the beating, as well as the arrested, heart. Midterm patency was shown to be perfect by means of angiography.
This article has been cited by other articles:
![]() |
U. Klima, H. MacVaugh III, E. Bagaev, M. Maringka, S. Kirschner, J. Beilner, and A. Haverich Magnetic Vascular Port in Minimally Invasive Direct Coronary Artery Bypass Grafting Circulation, September 14, 2004; 110(11_suppl_1): II-55 - II-60. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Carrel, L. Englberger, D. Keller, S. Windecker, B. Meier, and F. Eckstein Clinical and angiographic results after mechanical connection for distal anastomosis in coronary surgery J. Thorac. Cardiovasc. Surg., June 1, 2004; 127(6): 1632 - 1640. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Carrel, F. S. Eckstein, L. Englberger, P. A. Berdat, and J. Schmidli Clinical experience with devices for facilitated anastomoses in coronary artery bypass surgery Ann. Thorac. Surg., March 1, 2004; 77(3): 1110 - 1120. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Ngaage Off-pump coronary artery bypass grafting: the myth, the logic and the science Eur. J. Cardiothorac. Surg., October 1, 2003; 24(4): 557 - 570. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. D'Ancona, T. A. Salerno, P. S. Aoukar, and H. Karamanoukian Use of nitinol-U clips and flow characteristics of LIMA-LAD anastomoses Interactive CardioVascular and Thoracic Surgery, September 1, 2003; 2(3): 237 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Wolf, E. L. Alderman, M. P. Caskey, A. R. Raczkowski, M. K. Dullum, D. C. Lundell, A. C. Hill, N. Wang, and M. A. Daniel Clinical and six-month angiographic evaluation of coronary arterial graft interrupted anastomoses by use of a self-closing clip device: a multicenter prospective clinical trial J. Thorac. Cardiovasc. Surg., July 1, 2003; 126(1): 168 - 177. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Elkouri, P. Gloviczki, and N. W. Chbat Minimally Invasive Vascular Surgery and the Evolution of Vascular Anastomosis Techniques Perspectives in Vascular Surgery and Endovascular Therapy, January 1, 2002; 15(2): 127 - 153. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |