J Thorac Cardiovasc Surg 2002;124:785-789
© 2002 The American Association for Thoracic Surgery
Evolving Technology (ET) |
From the Department of Cardiology, Heart-lung Center Utrecht, University Medical Center,a Utrecht, and St Antonius Hospital,b Nieuwegein, The Netherlands.
Received for publication Aug 6, 2001. Revisions requested Oct 31, 2001; revisions received Nov 15, 2001. Accepted for publication Dec 27, 2001. Address for reprints: Cornelius Borst, MD, PhD, Professor of Experimental Cardiology, University Medical Center Utrecht (Room G02.523), PO Box 85500, 3508 GA Utrecht, The Netherlands (E-mail: c.borst{at}hli.azu.nl).
Objective: In off-pump coronary artery bypass surgery, both plaque and calcifications and torrential back flow from side branches in the isolated coronary segment may hamper coronary anastomosis suturing. The epicardial application of an ultrasonographic minitransducer with color Doppler scanning modality was studied in off-pump coronary surgery to detect septal perforating side branches; to assess the location, size, and quality of the coronary vessel; and to visualize the anastomotic orifice before chest closure.
Methods: Thirteen patients were selected for multivessel off-pump bypass. The investigation was limited to the left anterior descending coronary artery. The anastomotic target site was chosen on preoperative and intraoperative findings. The conventionally designated site was scanned with a novel 10-MHz ultrasonographic miniprobe (15 x 6 x 9 mm) that fitted in between the cardiac stabilizer pods.
Results: In 11 of the 13 cases, the course of the left anterior descending coronary artery could be properly identified. In 3 cases detection of perforating side branches caused a change in anastomotic site. A sufficiently dry field was obtained in all 13 cases. In 11 cases the anastomotic orifice was adequately visualized.
Conclusion: During off-pump coronary surgery, 10-MHz ultrasonographic images from a minitransducer aided in the intraoperative choice of anastomotic site.
This article has been cited by other articles:
![]() |
P. K. Hol, K. Andersen, H. Skulstad, P. S. Halvorsen, P. S. Lingaas, R. Andersen, J. Bergsland, and E. Fosse Epicardial Ultrasonography: A Potential Method for Intraoperative Quality Assessment of Coronary Bypass Anastomoses? Ann. Thorac. Surg., September 1, 2007; 84(3): 801 - 807. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Stein, J. M. Smith, J. R. Robinson, and M. R. Katz Target vessel detection and coronary anastomosis assessment by intraoperative 12-MHz ultrasound. Ann. Thorac. Surg., September 1, 2006; 82(3): 1078 - 1084. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, P. F. A. Bakker, R. Meijer, C. Borst, and P. F. Grundeman Ultrasound Mini-Transducer with Malleable Handle for Coronary Artery Surgery Ann. Thorac. Surg., January 1, 2006; 81(1): 322 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, W. J.L. Suyker, P. T.W. Suyker, C. W.J. Verlaan, R. Meijer, C. Borst, and P. F. Grundeman Quality assessment of distal S2AS connector anastomosis by 13MHz epicardial ultrasound Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 833 - 837. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, R. Meijer, T. C. Dessing, C. Borst, and P. F. Grundeman Detection of construction errors in ex vivo coronary artery anastomoses by 13-MHz epicardial ultrasonography J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 1078 - 1083. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, T. C. Dessing, R. Meijer, P. F.A. Bakker, C. Borst, and P. F. Grundeman Robot-assisted 13 MHz epicardial ultrasound for endoscopic quality assessment of coronary anastomoses Interactive CardioVascular and Thoracic Surgery, December 1, 2004; 3(4): 616 - 620. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Dessing, R. P.J. Budde, R. Meijer, P. F.A. Bakker, C. Borst, and P. F. Grundeman Geometry assessment of coronary artery anastomoses with construction errors by epicardial ultrasound Eur. J. Cardiothorac. Surg., August 1, 2004; 26(2): 257 - 261. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P. J. Budde, R. Meijer, P. F. A. Bakker, C. Borst, and P. F. Grundeman Endoscopic localization and assessment of coronary arteries by 13 MHz epicardial ultrasound Ann. Thorac. Surg., May 1, 2004; 77(5): 1586 - 1592. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Wolf and V. Falk Intraoperative assessment of coronary artery bypass grafts J. Thorac. Cardiovasc. Surg., September 1, 2003; 126(3): 634 - 637. [Full Text] [PDF] |
||||
![]() |
Y. Suematsu and S. Takamoto Future technologic innovations for intraoperative visualization of native coronary artery and graft anastomoses J. Thorac. Cardiovasc. Surg., July 1, 2003; 126(1): 304 - 304. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |