J Thorac Cardiovasc Surg 2003;125:353-360
© 2003 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology (CSP) |
From the Department of Cardiovascular Surgery, Graduate School of Medicine,a and the Department of Pharmacoepidemiology, School of Public Health,b Kyoto University, and the Department of Mechanical Engineering, Faculty of Engineering,c Setsunan University, Kyoto, Japan.
Received for publication Nov 26, 2001. Revisions requested Dec 26, 2001; revisions received March 14, 2002. Accepted for publication May 9, 2002. Address for reprints: Kazunobu Nishimura, MD, PhD, Department of Cardiovascular Surgery, Graduate School of Medicine, Kyoto University, Sakyo-ku, Kyoto 606-8507, Japan (E-mail: kas{at}kuhp.kyoto-u.ac.jp).
Objectives: The purpose of this study was to investigate how the inflow cannulation site of the left ventricular assist system with a centrifugal pump would influence cardiac function on failing heart models.
Methods: In 10 sheep, a left ventricular assist system was instituted by an outflow cannula in the descending aorta, two inflow cannulas in the left atrium and the left ventricle, and connecting those cannulas to a magnetically suspended centrifugal pump. A conductance catheter and a tipped micromanometer for monitoring the pressure-volume loop were also inserted into the left ventricle. Myocardial oxygen consumption was directly measured. Heart failure was induced by injection of microspheres into the left main coronary artery. The assist rate was varied from 0% to 100% at each inflow cannulation site.
Results: The pump flow with left ventricular cannulation increased during the systolic phase and decreased during the diastolic phase, whereas it was constant with left atrial cannulation. Ejection fraction with left atrial cannulation decreased as the assist rate increased, whereas that with left ventricular cannulation was maintained up to 75% assist. The external work with left atrial cannulation decreased gradually as the assist rate increased, whereas the external work with left ventricular cannulation did not decrease until the assist rate reached 75%. The myocardial oxygen consumption in both cannulations decreased proportionally as the assist rate increased; they were significantly less with left ventricular cannulation at the 100% assist rate than with left atrial cannulation.
Conclusion: Left ventricular cannulation during left ventricular assistance maintains ejection fraction and effectively reduces oxygen consumption.
This article has been cited by other articles:
![]() |
Y. Ootaki, K. Kamohara, M. Akiyama, F. Zahr, M. W. Kopcak Jr., R. Dessoffy, and K. Fukamachi Phasic coronary blood flow pattern during a continuous flow left ventricular assist support Eur. J. Cardiothorac. Surg., November 1, 2005; 28(5): 711 - 716. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |