JTCS Speed Up Your Browser
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Jens Scheewe
Jochen T. Cremer
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Böning, A.
Right arrow Articles by Cremer, J. T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Böning, A.
Right arrow Articles by Cremer, J. T.
Related Collections
Right arrow Extracorporeal circulation

J Thorac Cardiovasc Surg 2004;127:1458-1465
© 2004 The American Association for Thoracic Surgery


Surgery for congenital heart disease

Phosphorylcholine or heparin coating for pediatric extracorporeal circulation causes similar biologic effects in neonates and infants

Andreas Böning, MDa,*, Jens Scheewe, MDa, Thomas Ivers, ECCPa, Christine Friedrich, PhDa, Jürgen Stieh, MDb, Sandra Freitag, PhDc, Jochen T. Cremer, PhDa

a Department of Cardiovascular Surgery, University Hospital, Kiel, Germany,
b Department of Pediatric Cardiology, University Hospital, Kiel, Germany,
c Institute of Medical Informatics and Statistics,c University Hospital, Kiel, Germany

Received for publication April 22, 2003; revisions received August 14, 2003; accepted for publication August 18, 2003.

* Address for reprints: Andreas Böning, MD, Department of Cardiovascular Surgery, University Hospital, Arnold-Heller-Str 7, 24105 Kiel, Germany
aboening{at}kielheart.uni-kiel.de

OBJECTIVE: Cardiac surgery for complex congenital malformations with use of extracorporeal circulation predisposes to an excessive systemic inflammatory response and a consecutive capillary leak syndrome. In a prospective randomized study the influence of 2 oxygenators especially designed for pediatric use on inflammatory markers and clinical outcome was investigated.

METHODS: Forty neonates and infants (body surface area, <0.36 m2) undergoing cardiac surgery with extracorporeal circulation were randomized into one of 3 groups: in the first group (n = 14) the Medtronic Minimax Oxygenator and in the second group (n = 12) the Dideco Lilliput 1 Oxygenator, both with a 750-mL priming volume, were used. In the third group the Dideco Lilliput 1 Oxygenator was filled with a reduced priming volume of 450 mL. Parameters of interest for evaluation of a systemic inflammatory response after extracorporeal circulation were interleukin 6, tumor necrosis factor {alpha}, neutrophil elastase, complement C3, and free hemoglobin. In addition, erythrocyte, leukocyte, and thrombocyte counts and hemoglobin and C-reactive protein values were determined at different measurement points before, during, and after the operation.

RESULTS: In all 3 groups peak values for tumor necrosis factor {alpha} were observed during the operation, whereas interleukin 6, elastase, and free hemoglobin values peaked in the first 4 hours. The highest values for leukocytes and C-reactive protein were obtained between 24 and 72 hours after the operation. Erythrocyte and thrombocyte counts, as well as hemoglobin values, were lowest at extracorporeal circulation onset, normalizing under substitution in the first 4 hours after the operation. By using the Lilliput/750 oxygenator, higher interleukin 6 values 1 and 4 hours after the operation and higher tumor necrosis factor {alpha} values during and 1 hour after the operation could be observed compared with results with the Minimax and Lilliput/450 oxygenators. In spite of our randomization protocol, patients in the Lilliput/750 group were significantly smaller and younger than those in the Minimax group. However, the statistical analysis showed no correlation between age and interleukin 6 or tumor necrosis factor {alpha} values, but it did show a correlation between younger age and the occurrence of capillary leak syndrome. Accordingly, the number of children with clinically complicated course (capillary leak, longer duration of catecholamine therapy, and ventilation) was higher in the Lilliput/750 group than in the Minimax group.

CONCLUSION: By using an adequate priming volume, the systemic inflammatory response is similar after use of the Dideco Lilliput 1 Oxygenator and the Medtronic Minimax Oxygenator. Tip-to-tip surface coating of the extracorporeal circulation with either heparin or phosphorylcholine seems to have similar biologic effects in neonates and infants undergoing cardiac surgery.





This article has been cited by other articles:


Home page
ICVTSHome page
K. Miyaji, T. Miyamoto, S. Kohira, K. Nakashima, N. Inoue, H. Sato, and K. Ohara
Miniaturized cardiopulmonary bypass system in neonates and small infants
Interactive CardioVascular and Thoracic Surgery, February 1, 2008; 7(1): 75 - 79.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
A. M. Draaisma, M. G. Hazekamp, N. Anes, P. H. Schoof, C. E. Hack, A. Sturk, and R. A.E. Dion
Phosphorylcholine Coating of Bypass Systems Used for Young Infants Does Not Attenuate the Inflammatory Response
Ann. Thorac. Surg., April 1, 2006; 81(4): 1455 - 1459.
[Abstract] [Full Text] [PDF]


Home page
PediatricsHome page
S. J. Roth, I. Adatia, G. D. Pearson, and and Members of the Cardiology Group
Summary proceedings from the cardiology group on postoperative cardiac dysfunction.
Pediatrics, March 1, 2006; 117(3 Pt 2): S40 - S46.
[Abstract] [Full Text] [PDF]


Home page
PerfusionHome page
A. Khosravi, C. A Skrabal, B. Westphal, G. Kundt, B. Greim, E. Kunesch, A. Liebold, and G. Steinhoff
Evaluation of coated oxygenators in cardiopulmonary bypass systems and their impact on neurocognitive function
Perfusion, September 1, 2005; 20(5): 249 - 254.