JTCS St. Jude Medical
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Cornelius Borst
Paul F. Gründeman
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Budde, R. P.J.
Right arrow Articles by Gründeman, P. F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Budde, R. P.J.
Right arrow Articles by Gründeman, P. F.
Related Collections
Right arrow Minimally invasive surgery

J Thorac Cardiovasc Surg 2005;129:1078-1083
© 2005 The American Association for Thoracic Surgery


Evolving Technology

Detection of construction errors in ex vivo coronary artery anastomoses by 13-MHz epicardial ultrasonography

Ricardo P.J. Budde, MSc, Rudy Meijer, Thomas C. Dessing, MSc, Cornelius Borst, MD, PhD, Paul F. Gründeman, MD, PhD*

Heart Lung Center Utrecht, University Medical Center, Utrecht, The Netherlands.

Received for publication July 4, 2004; revisions received September 1, 2004; accepted for publication September 7, 2004.

* Address for reprints: Paul F. Gründeman, MD, PhD, Associate Professor, Cardiothoracic and Vascular Research, Large Animals, University Medical Center Utrecht (Room G02.523), Heidelberglaan 100, 3584 CX Utrecht, The Netherlands. (E-mail: p.f.grundeman{at}hli.azu.nl).

OBJECTIVE: Intraoperative detection of suboptimal coronary anastomoses allows revision before chest closure. We evaluated an epicardial 13-MHz ultrasound minitransducer as a means to detect three different coronary anastomosis construction errors.

METHODS: In total, 120 internal thoracic arter-t-coronary artery anastomoses were constructed correctly (n = 60) or incorrectly (n = 60) with one technical error: suture crossover, purse-string or deep toe stitch (n = 20 each). Anastomoses were performed on ex vivo pressure-perfused porcine (96 anastomoses) and human hearts (24 anastomoses). Two blinded observers scanned and scored the anastomoses with epicardial ultrasonography. In 24 human and 24 porcine anastomoses, angiograms were made of 24 correct and 24 incorrect anastomoses and scored by two other blinded observers. Angioscopy and cast injection served as a reference.

RESULTS: Overall, 119 of 120 anastomoses were accurately scored as correct or incorrect within a median of 67 seconds (8–381 seconds) by both observers (sensitivity 0.98, specificity 1.00, {kappa} 1.00 (1.00, 1.00, and 1.00 in angiography subset, respectively). One deep toe stitch that induced outflow corner stenosis was spotted by both observers but regarded as insignificant and thus inaccurately scored as correct. In 5 anastomoses, unintended irregularities were detected. By angiography, anastomoses were accurately scored with a sensitivity of 0.75 and a specificity of 0.81 (P < .001 vs ultrasonography) and {kappa} of 0.54. Angioscopy and cast confirmed ultrasonographic findings and did not reveal irregularities other than detected by ultrasonography.

CONCLUSION: Ex vivo epicardial 13-MHz ultrasonography allowed rapid and accurate evaluation of coronary anastomoses and detected technical construction errors with higher sensitivity and specificity than angiography.





This article has been cited by other articles:


Home page
Ann. Thorac. Surg.Home page
P. K. Hol, K. Andersen, H. Skulstad, P. S. Halvorsen, P. S. Lingaas, R. Andersen, J. Bergsland, and E. Fosse
Epicardial Ultrasonography: A Potential Method for Intraoperative Quality Assessment of Coronary Bypass Anastomoses?
Ann. Thorac. Surg., September 1, 2007; 84(3): 801 - 807.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
R. P.J. Budde, P. F. A. Bakker, R. Meijer, C. Borst, and P. F. Grundeman
Ultrasound Mini-Transducer with Malleable Handle for Coronary Artery Surgery
Ann. Thorac. Surg., January 1, 2006; 81(1): 322 - 326.
[Abstract] [Full Text] [PDF]


Home page
Eur. J. Cardiothorac. Surg.Home page
R. P.J. Budde, W. J.L. Suyker, P. T.W. Suyker, C. W.J. Verlaan, R. Meijer, C. Borst, and P. F. Grundeman
Quality assessment of distal S2AS connector anastomosis by 13MHz epicardial ultrasound
Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 833 - 837.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2005 by The American Association for Thoracic Surgery.