This report presents the chest film findings in three patients with Beall
mitral valve prostheses in whom severe disc wear resulted in disc
embolization. In all three cases the worn disc embolized to the thoracic
aorta with catastrophic decompensation, and all three patients were
salvaged with emergency operation. Detection of the embolized disc was
possible with plain chest radiography in each case and may permit removal
of the embolized disc at the time of valve replacement. Early detection of
disc wear and replacement of worn Beall mitral valve prostheses are
advocated to avoid this left-threatening complication.