JTCS Email Content Delivery
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Spence, P. A.
Right arrow Articles by Salerno, T. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Spence, P. A.
Right arrow Articles by Salerno, T. A.

The Journal of Thoracic and Cardiovascular Surgery, Vol 89, 264-268, Copyright © 1985 by The American Association for Thoracic Surgery and The Western Thoracic Surgical Association


ARTICLES

Pulmonary artery balloon counterpulsation in the management of right heart failure during left heart bypass

PA Spence, RD Weisel, J Easdown, KA Jabr and TA Salerno

Pulmonary artery balloon counterpulsation was instituted in 10 pigs when right ventricular failure limited cardiac output. Global myocardial depression was produced by infusion of propranolol, and the left ventricle was fully supported by left heart bypass. A stable model of failure was achieved in six pigs. Following application of pulmonary artery balloon counterpulsation right atrial pressure decreased from 18.2 +/- 2.1 to 15.9 +/- 2.5 mm Hg (p less than 0.05). Cardiac output increased from 416 +/- 94 to 758 +/- 127 ml/min (p less than 0.001). Right ventricular stroke work increased from 0.29 +/- 0.07 to 0.65 +/- 0.12 gm X m. (p less than 0.05). There was no cardiac output before or after institution of balloon counterpulsation in four pigs studied during ventricular fibrillation or asystole. We conclude that pulmonary artery balloon counterpulsation improved cardiac output and right ventricular stroke work in a model of right ventricular failure where the pulmonary circulation was unaltered and the left ventricle supported by left heart bypass. Balloon counterpulsation was not effective during ventricular fibrillation or asystole. Pulmonary artery balloon counterpulsation should be considered when right ventricular failure limits cardiac output during left heart bypass.


This article has been cited by other articles:


Home page
PerfusionHome page
J. Chalmers and T. Graham
Review article : Mechanical ventricular support in the management of postcardiotomy cardiogenic shock
Perfusion, October 1, 1989; 4(4): 243 - 253.