The Journal of Thoracic and Cardiovascular Surgery, Vol 96, 952-961, Copyright © 1988 by The American Association for Thoracic Surgery and The Western Thoracic Surgical Association
E Knott, H Reul, M Knoch, U Steinseifer and G Rau
The hydrodynamic performance of a series of mechanical heart valve
prostheses is measured in the aortic position. All experiments are
performed in an electrohydraulic, computer-controlled pulse duplicator
simulating the left side of the human circulatory system. Testing
conditions are set according to a Food and Drug Administration
interlaboratory comparison protocol with cardiac outputs of 3.0, 4.5, 6.5,
and 8.0 L/min at a constant heart rate of 70 beats/min. Mean systolic
pressure differences, volume losses, and energy losses, as well as
dimensionless pressure losses and energy loss coefficients, are calculated
from the recorded pressure, volume, and flow tracings. The results of 10
mechanical heart valve prostheses (eight tilting disc valves and two
bileaflet valves) are presented and their clinical relevance is discussed.
ARTICLES
In vitro comparison of aortic heart valve prostheses. Part 1: Mechanical valves
Helmholtz-Institute for Biomedical Engineering, Aachen University of Technology, Federal Republic of Germany.
This article has been cited by other articles:
![]() |
R. Fries, T. Graeter, D. Aicher, H. Reul, C. Schmitz, M. Bohm, and H.-J. Schafers In vitro comparison of aortic valve movement after valve-preserving aortic replacement J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 32 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. De Paulis, C. Schmitz, R. Scaffa, P. Nardi, L. Chiariello, and H. Reul In vitro evaluation of aortic valve prosthesis in a novel valved conduit with pseudosinuses of Valsalva J. Thorac. Cardiovasc. Surg., October 1, 2005; 130(4): 1016 - 1021. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hartrumpf, J. M. Albes, T. Krempl, V. Rudolph, and T. Wahlers The hemodynamic performance of standard bileaflet valves is impaired by a tilted implantation position Eur. J. Cardiothorac. Surg., March 1, 2003; 23(3): 283 - 291. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Niinami, S. Aomi, H. Tomioka, K. Nakano, and H. Koyanagi A comparison of the in vivo performance of the 19-mm St. Jude medical hemodynamic plus and 21-mm standard valve Ann. Thorac. Surg., October 1, 2002; 74(4): 1120 - 1124. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Torregrosa, J. Gomez-Plana, F. J. Valera, J. Caffarena, J. M. Maronas, F. Garcia-Sanchez, J. Peris, R. Frias, and J. M. Caffarena Long-term clinical experience with the omnicarbon prosthetic valve Ann. Thorac. Surg., September 1, 1999; 68(3): 881 - 886. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lund, K. Emmertsen, T. T. Nielsen, F. T. Jensen, C. Flo, H. K. Pilegaard, B. S. Rasmussen, O. K. Hansen, and L. H. Kristensen Impact of Size Mismatch and Left Ventricular Function on Performance of the St. Jude Disc Valve After Aortic Valve Replacement Ann. Thorac. Surg., May 1, 1997; 63(5): 1227 - 1234. [Abstract] [Full Text] |
||||
![]() |
A. A. Mikhail and C. W. Akins Omniscience Valve Evolution and Literature Ann. Thorac. Surg., August 1, 1996; 62(2): 624 - 626. [Full Text] |
||||
![]() |
T. Abe, K. Kamata, K. Kuwaki, K. Komatsu, and S. Komatsu Ten Years' Experience of Aortic Valve Replacement With the Omnicarbon Valve Prosthesis Ann. Thorac. Surg., April 1, 1996; 61(4): 1182 - 1187. [Abstract] [Full Text] |
||||
![]() |
S. A. Blome-Eberwein, D. Mrowinski, J. Hofmeister, and R. Hetzer Impact of Mechanical Heart Valve Prosthesis Sound on Patients' Quality of Life Ann. Thorac. Surg., February 1, 1996; 61(2): 594 - 602. [Abstract] [Full Text] |
||||
![]() |
A. Laske, R. Jenni, M. Maloigne, G. Vassalli, O. Bertel, and M. I. Turina Pressure Gradients Across Bileaflet Aortic Valves by Direct Measurement and Echocardiography Ann. Thorac. Surg., January 1, 1996; 61(1): 48 - 57. [Abstract] [Full Text] |
||||
![]() |
W. Voelker, H. Reul, G. N. Ing, T. S. Ing, B. S. Ing, A. S. Ing, and K. R. Karsch Comparison of Valvular Resistance, Stroke Work Loss, and Gorlin Valve Area for Quantification of Aortic Stenosis : An In Vitro Study in a Pulsatile Aortic Flow Model Circulation, February 15, 1995; 91(4): 1196 - 1204. [Abstract] [Full Text] |
||||
![]() |
O. Lund Late Chronic Hemolysis after Valve Replacement for Aortic Stenosis. Relation to Residual Hypertrophy and Impaired Left Ventricular Function Angiology, October 1, 1990; 41(10): 836 - 847. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |