The Journal of Thoracic and Cardiovascular Surgery, Vol 97, 90-94, Copyright © 1989 by The American Association for Thoracic Surgery and The Western Thoracic Surgical Association
W Klepetko, A Moritz, J Mlczoch, H Schurawitzki, E Domanig and E Wolner
Two cases of leaflet fracture in the Edwards-Duromedics valve at 36 and 38
months after implantation are reported. Both patients were immediately
reoperated on and recovered well. In one valve an older housing fracture
with partial tissue ingrowth was noted beside a recent transverse leaflet
fracture. In the other valve the leaflet was fractured near the pivot
mechanism. All larger embolized parts were detected in the iliac artery
region by computed tomographic scan and were subsequently removed. Problems
in diagnosis and the importance of immediate reoperation, even without
exact diagnosis, are discussed. Technical evaluation of the valve revealed
crack growth and arrest, giving evidence of fatigue fracture. Scanning
electron microscopic examination revealed several areas of pitting and
erosion. Although the exact cause of mechanical disruption remains
speculative, pyrolytic carbon seems to have the characteristic of fatigue
fracture as well as erosion damage. A connection between the two might
exist.
ARTICLES
Leaflet fracture in Edwards-Duromedics bileaflet valves
Second Surgical Department, University of Vienna, Austria.
This article has been cited by other articles:
![]() |
R. F. Padera Jr. and F. J. Schoen Pathology of Cardiac Surgery Card. Surg. Adult, January 1, 2008; 3(2008): 111 - 178. [Full Text] |
||||
![]() |
T. F. Cianciulli, J. A. Lax, M. C. Saccheri, H. J. Redruello, S. M. Belforte, V. P. Picone, and H. A. Prezioso Acute mitral valve dysfunction due to leaflet escape in a Tri-technologies bileaflet mechanical valve Eur J Echocardiogr, January 1, 2007; 8(1): 63 - 66. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Butany and M J Collins Analysis of prosthetic cardiac devices: a guide for the practising pathologist J. Clin. Pathol., February 1, 2005; 58(2): 113 - 124. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. A. Ferraris and S. P. Ferraris Assessing the medical literature: let the buyer beware Ann. Thorac. Surg., July 1, 2003; 76(1): 4 - 11. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. J. Schoen and R. F. Padera Jr. Cardiac Surgical Pathology Card. Surg. Adult, January 1, 2003; 2(2003): 119 - 185. [Full Text] |
||||
![]() |
A. Kornberg, S. M. Wildhirt, C. Schulze, and E. Kreuzer Leaflet escape in Omnicarbon monoleaflet valve Eur. J. Cardiothorac. Surg., June 1, 1999; 15(6): 867 - 869. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. K. Podesser, G. Khuenl-Brady, E. Eigenbauer, S. Roedler, A. Schmiedberger, E. Wolner, and A. Moritz Long-term results of heart valve replacement with the Edwards Duromedics bileaflet prosthesis: A prospective ten-year clinical follow-up J. Thorac. Cardiovasc. Surg., May 1, 1998; 115(5): 1121 - 1129. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Vitale, D. Serena, C. Fondacone, A. Agnino, P. Cicala, and L. de Luca COMPLETE INTRAOPERATIVE DISLODGMENT OF A ST. JUDE MEDICAL MECHANICAL HEART VALVE HEMODYNAMIC PLUS* VALVE SEWING CUFF J. Thorac. Cardiovasc. Surg., December 1, 1996; 112(6): 1669 - 1671. [Full Text] |
||||
![]() |