The Journal of Thoracic and Cardiovascular Surgery, Vol 97, 95-97, Copyright © 1989 by The American Association for Thoracic Surgery and The Western Thoracic Surgical Association
D Lindblom, L Rodriguez and VO Bjork
The early production series of the 29 to 31 mm 70-degree convexo- concave
(CC70) Bjork-Shiley heart valves constitutes a subgroup of CC70 values with
an extra high risk of outlet strut fracture. The hazard function for
mechanical failure among these valves has remained almost constant over a
7-year follow-up period, and the 7-year actuarial incidence of mechanical
failure among these valves is 12.5%. Selected patients with these valves
will be considered for prophylactic rereplacement.
ARTICLES
Mechanical failure of the Bjork-Shiley valve. Updated follow-up and considerations on prophylactic rereplacement
Department of Thoracic Surgery, Karolinska Hospital, Stockholm, Sweden.
This article has been cited by other articles:
![]() |
L. Hansen, M. Danne, B. Hoffmann, and F.-C. Riess Structural valve failure with every beat regurgitation observed using the Medtronic Advantage aortic valve. J. Thorac. Cardiovasc. Surg., November 1, 2007; 134(5): 1344 - 1345. [Full Text] [PDF] |
||||
![]() |
W. J. Blot, M. A. Ibrahim, T. D. Ivey, D. E. Acheson, R. Brookmeyer, A. Weyman, J. Defauw, J. K. Smith, and D. Harrison Twenty-Five-Year Experience With the Bjork-Shiley Convexoconcave Heart Valve: A Continuing Clinical Concern Circulation, May 31, 2005; 111(21): 2850 - 2857. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. L. Gott, D. E. Alejo, and D. E. Cameron Mechanical heart valves: 50 years of evolution Ann. Thorac. Surg., December 1, 2003; 76(6): S2230 - 2239. [Abstract] [Full Text] [PDF] |
||||
![]() |
R Z Omar, L S Morton, D A Halliday, E M Danns, M T Beirne, W J Blot, and K M Taylor Outlet strut fracture of Bjork-Shiley convexo concave heart valves: the UK cohort study Heart, July 1, 2001; 86(1): 57 - 62. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. W. Steyerberg, M. Kallewaard, Y. Van Der Graaf, L. A. Van Herwerden, and J.D. F. Habbema Decision Analyses for Prophylactic Replacement of the Bjork-Shiley Convexo-concave Heart Valve:: An Evaluation of Assumptions and Estimates Med Decis Making, January 1, 2000; 20(1): 20 - 32. [Abstract] [PDF] |
||||
![]() |
B. K. Podesser, G. Khuenl-Brady, E. Eigenbauer, S. Roedler, A. Schmiedberger, E. Wolner, and A. Moritz Long-term results of heart valve replacement with the Edwards Duromedics bileaflet prosthesis: A prospective ten-year clinical follow-up J. Thorac. Cardiovasc. Surg., May 1, 1998; 115(5): 1121 - 1129. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kallewaard, A. Algra, J. Defauw, Y. d. Graaf, and B.-S. S. Group Prophylactic Replacement Of BjOrk-Shiley Convexo-Concave Valves At Risk Of Strut Fracture J. Thorac. Cardiovasc. Surg., March 1, 1998; 115(3): 577 - 581. [Abstract] [Full Text] |
||||
![]() |
D. Lindblom Valve thrombosis and strut fracture with the Bjork-Shiley valve J. Thorac. Cardiovasc. Surg., July 1, 1994; 108(1): 177 - 178. [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |