J Thorac Cardiovasc Surg 1998;115:1111-1115
© 1998 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
From the Section of Pediatric Cardiology and Cardiovascular Surgery, Veterans General Hospital, Taipei, the Department of Chemical Engineering, National Central University, and the Department of Pathology, Pig Research Institute, Taipei, Taiwan.
Received for publication Dec. 24, 1996. Revisions requested April 15, 1997; revisions received Sept. 24, 1997. Accepted for publication Sept. 30, 1997. Address for reprints: Jen-Her Lu, MD, Section of Pediatric Cardiology, Veterans General Hospital, Taipei, Taiwan.
Abstract
Objective: Primary concerns about currently available pericardial substitutes include adhesion and epicardial reaction. The purpose of this study is to evaluate host reaction to pericardial substitutes with and without incorporating slow heparin release.
Methods: To avoid biologic variation among these pericardial patches, we made a composite of six membranes. The composite membrane consisted of epoxy-fixed patches with (1) or without (2) ionically bound heparin, a glutaraldehyde-fixed patch with (3) or without (4) ionically bound heparin, an expanded polytetrafluoroethylene patch (5), and a polyester polymeric patch (6). Ten recipient dogs weighing from 12 to 19 kg (mean 13.6 kg) were used to assess the composite membranes as pericardial substitutes. The implanted composite membranes were retrieved 1 week (one dog), 2 weeks (one dog), 4 weeks (one dog), 8 weeks (one dog), and 12 weeks (six dogs) after implantation.
Results: Overall, the synthetic patches had a more notable inflammatory reaction than the biologic patches with or without ionically bound heparin. The heparin-bound patches caused significantly less inflammation than their nonheparinized counterparts. The heparinized porcine patches cross-linked with different compounds were found to have less fibrous formation than the nonheparinized patches and the synthetic patches.
Conclusions: Heparinized pericardial substitutes may cause less adhesion and inflammatory reaction than nonheparinized material.
Primary closure of the pericardium is recommended because it is believed to reduce the formation of mediastinal adhesion and postoperative complications. The use of pericardial substitutes is advocated when primary closure is not possible. Several materials can be used as pericardial substitutes, including fascia lata,
1 silicone rubber,
2-4 rubber-coated Dacron
2 or polyester polymer,
5 and polytetrafluoroethylene (PTFE),
2 as well as mammalian pericardium.
6-11 These materials have been tested in the past to facilitate reoperations, with inconclusive results.
The primary concerns about currently available pericardial substitutes include adhesion and epicardial reaction. The synthetic patches, made from expanded PTFE (ePTFE) or polyester polymer (Dacron), are widely used as pericardial substitutes. Several groups
2-6 reported the presence of a tissue-free space around the synthetic patch that facilitates the opening of the chest and diminishes damage to the heart. However, the synthetic patch brought on severe inflammatory reaction and diffuse fibrosis. The epicardium beneath the synthetic patch became covered by a dense layer of fibrotic tissue. In a canine model, Meus and associates
5 reported that recognition of the coronary artery was greatly impeded by this fibrotic tissue, which was very difficult to separate from the underlying myocardium and vessels. The glutaraldehyde-fixed patches, made from heterologous porcine,
5,6 equine,
7,8 and bovine pericardium,
9-11 were often used as pericardial substitutes. However, glutaraldehyde tended to stiffen biologic tissue and enhance fibrosis, adhesion, and calcification.
11 To address the aforementioned deficiencies, a cross-linking technique using an epoxy compound as a fixative has recently been developed.
12,13 Furthermore, heparinization of biomaterials to enhance the antithrombogenicity is now recognized as a decisive factor in the development of antiadhesive membranes.
14 The purpose of this study was to evaluate host reaction to pericardial substitute with and without incorporating slow heparin release. In addition, two commercially available synthetic pericardial substitutes (ePTFE and polyester polymeric patch) were used as controls.
Methods
Pericardium harvesting
Fresh porcine pericardia procured from a slaughterhouse were used as the raw materials. The procured porcine pericardia were transported in cold physiologic saline solution. On return, the pericardia were first gently rinsed with fresh saline solution to rid them of excess blood on the tissue. Adherent fat was then carefully trimmed from the pericardial surface. The maximum time period between retrieval and initiation of tissue fixation was less than 10 hours.
Fixation process
For the epoxy-fixed pericardial patches without ionically bound heparin, the porcine pericardia were first fixed in an epoxy compound solution (ethylene glycol diglycidyl ether, Denacol EX-810, Nagase Chemicals, Ltd., Osaka, Japan) at room temperature for 6 days. The solution was buffered with sodium carbonate/sodium bicarbonate (0.21/0.02 mol/L, pH 10.5). The amount of solution used in fixation was approximately 200 ml for a 6 x 6 cm piece of pericardium. The fixed porcine pericardia were subsequently sterilized with the same epoxy compound solution (2%) at 37° C for 7 days.
For the epoxy-fixed pericardial patches with ionically bound heparin, the fresh porcine pericardia were first immersed in a 1% protamine sulfate solution for l hour. The porcine pericardia were then fixed in the epoxy compound solution as described at room temperature for 6 days. Finally, the fixed pericardia were heparinized in a 0.5% sodium heparin solution (from porcine intestinal mucosa, Sigma Chemical Co., St. Louis, Mo.) at 37° C for 1 hour. The same sterilization procedure as used for the epoxy-fixed patch without ionically bound heparin was applied.
For the glutaraldehyde-fixed pericardial patches without ionically bound heparin, the fresh porcine pericardia were first fixed in a 0.625% glutaraldehyde solution at room temperature for 10 hours. The solution was buffered with HEPES buffer (0.02 mol/L, pH 7.4). Subsequently, the fixed pericardia were sterilized in a glutaraldehyde-based sterilant in a phosphate buffer at 45° C for 2 days. The heparinization process for the glutaraldehyde-fixed pericardial patches with ionically bound heparin was similar to that for their epoxy-fixed counterparts: the impregnation of protamine and heparinization were performed just before and after glutaraldehyde fixation, respectively.
Composite membranes as pericardial substitutes
To avoid biologic variation among these pericardial patches, we used a composite membrane. Preparation of the composite patches was completed by suturing these patches with 6-0 polypropylene sutures in the following order: (1) an epoxy-fixed patch with ionically bound heparin, (2) a glutaraldehyde-fixed patch with ionically bound heparin, (3) an ePTFE patch,*
(4) a polyester polymeric patch (Cooley low porosity polyester fabric, Meadox Medicals, Inc., Oakland, N.J.), (5) an epoxy-fixed patch without ionically bound heparin, and (6) a glutaraldehyde-fixed patch without ionically bound heparin. Each individual patch was 3 cm in length and 1 cm in width, which made a 3 cm x 6 cm composite membrane. Care was taken to ensure that the smooth side of this biologic patch was in contact with the epicardial surface of the recipient hearts.
Surgery
Ten recipient dogs weighing from 12 to 19 kg (mean 13.6 kg) were used to assess the composite membranes. An intravenous line, an arterial line in the femoral artery, and an electrocardiographic monitor were established. Anesthesia was induced via the external jugular vein with 2.5% thiamylal sodium (Surital) at a dose of 17.5 mg/kg. Anesthesia was maintained in the operating room with halothane and nitrous oxide in oxygen. Mechanical ventilation was instituted with an approximate tidal volume set at 10 ml/kg body weight with 100% oxygen at a rate of 8 to 14 cycles per minute. Anesthesia was maintained at stage III, planes 1 to 2, with a mixture of halothane in oxygen. Ventilation rates were adjusted to maintain arterial blood gases in the normal physiologic range.
The heart was exposed via a left lateral thoracotomy incision through the fourth intercostal space. A 2 x 5 cm segment of the midportion of the pericardium was removed and the composite patch was anchored to the pericardial defect with 6-0 Prolene interrupted sutures (Ethicon, Inc., Somerville, N.J.). Care was taken to avoid contact with the epicardial surface of the heart. Prophylactic antibiotics were given intravenously just before and after the operation and were administered intramuscularly for 5 days after the operation. Animal care was in compliance with the "Principles of Laboratory Animal Care" formulated by the National Society for Medical Research and the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources and published by the National Institutes of Health (NIH Publication No. 86-23, revised 1985).
Pathologic examination
The implanted composite membranes were retrieved at the end of 1 week (one dog), 2 weeks (one dog), 4 weeks (one dog), 8 weeks (one dog), and 12 weeks (six dogs) after the operation. The en bloc heart, including pericardium and left lung, was removed through a left-sided thoracotomy. The retrieved composite membranes were then examined and photographed. The development of adhesions and the epicardial reaction were graded (Table I). Subsequently, the retrieved patches were fixed in 10% phosphate-buffered formalin for at least 3 days. Histologic specimens of the patches, suture lines, and epicardia were obtained. All specimens for histologic examination, including the samples taken after harvesting, were embedded in paraffin, sectioned at 5 µm, and stained with hematoxylin-eosin. All biologic patches were stained with van Gieson technique additionally to demonstrate the collagen component. A system of assessing the microscopic status of the individual patches by assigning them arbitrary scores was devised and is described in Table II.
|
|
All the animals survived implantation and were evaluated as scheduled. From the sequential observation of the specimens, the epicardial and pleural reactions to the patch could be classified macroscopically with regard to pliability of the patch, adhesions to the epicardium, and epicardial fibrous reaction (Tables III and IV). Microscopic evaluation included acute and chronic active inflammatory reaction and fibrous reaction (Tables V and VI).
|
|
|
|
Microscopic findings
The acute inflammatory reaction was chiefly noted in the first several weeks. This reaction consisted of ecchymosis with neutrophil infiltration into the epicardium as a sign of postoperative mechanical stimulation from a foreign body. The degree of acute inflammatory reaction decreased after 2 weeks of implantation. The chronic active inflammatory reaction consisted of a large amount of lymphocyte aggregation, histiocytes, macrophages, giant cells, and an association with necrotic cells. The fibroblast reaction was represented as focal fibrosis, hyalinization, fibrin deposit, and dense fibrous connective tissue. The severity of mononuclear lymphocyte infiltration with fibrosis as a sign of chronic active inflammatory reaction gradually increased after 4 weeks of implantation (Tables V
and VI
). Compatible with macroscopic findings, the inflammatory and fibrous reaction on the rough surface of biologic patch (pleural side) was more severe than that observed on the smooth surface (epicardial side).
Epoxy-fixed pericardial substitutes
On gross examination, all epoxy-fixed patches appeared soft and pliable. The patches with heparin caused minimal to mild acute and chronic inflammatory reaction between the patch and the epicardial surface before 12 weeks. Histologic studies showed only sparse neutrophils and plasma cell infiltration. Also, on gross examination, the external (pulmonic) layer of the patch showed trivial fibrin deposits. The fibrous reaction was minimal, and only scanty adhesion with fibrin deposits to the epicardium and suture lines was observed before 12 weeks. However, the adhesion was aggravated in those patches retrieved at 12 weeks. Histologic studies showed a thin fibrous tissue layer. There were mild chronic inflammatory reactions in the patches with heparin at 12 weeks after implantation (Fig. 1). Observations in the patches without heparin were similar to those in the heparin-bound patches, although the inflammatory reaction and epicardial adhesion were more severe (Fig. 2).
|
|
|
|
|
Polyester pericardial substitutes
The polyester patch (Dacron) brought on the most severe inflammatory reaction and fibrosis between the pleura and epicardium. Diffuse fibrosis was present in all Dacron patches. On gross examination all patches appeared shrunken and thickened (Fig. 6, left). Two weeks after implantation, moderate adhesions were observed for these polyester patches. Severe adhesions were found at 4 weeks. These structural changes occurred on both surfaces of the patch (Tables IV
and V
). This layer could not be dissected away from the epicardium because of its thickness and adherence. There was no tissue-free space around the Dacron patch. The underlying structures were not identifiable. Histologic examination showed growth of dense fibrous connective tissue. The Dacron fabric itself showed a dense fibrous ingrowth of native collagen between the fiber of the Dacron patch at the sites of contact with the epicardial wall. The epicardial surface showed severe active chronic inflammation at 12 weeks (Fig. 6
, top left and bottom left), which was localized directly below the patch.
|
Reoperation is a risk predominantly because fibrosis and scarring increase the danger of injury to the heart, great vessels, or grafts during a repeat sternotomy.
15 Reoperations are prolonged by the takedown of adhesions and are associated with increased operative bleeding and complications. The presence of epicardial fibrosis obscures the underlying anatomic structures. This is particularly important in patients undergoing repeat coronary artery revascularization, because the identification of the coronary arteries is tedious and hazardous.
Research on heterologous pericardial substitutes had focused on the glutaraldehyde-treated pericardium. Meus and coworkers
5 reported that glutaraldehyde-treated biologic patches were excellent substitutes and had better antithrombogenicity than synthetic pericardial substitutes. Using a growing rat model, Okoshi's group
16 reported that the epoxy-fixed collagen gel disks implanted subcutaneously showed remarkably less calcium deposition than their glutaraldehyde-fixed counterparts. To overcome the thrombogenic problem commonly observed in the prosthetic vascular grafts, Noishiki and Miyata
14 developed a simple method to ionically bind heparin to the biologic arteries fixed with glutaraldehyde or epoxy compounds. In their animal studies, they found that the heparinized vascular grafts were significantly superior in antithrombogenicity to their nonheparinized counterparts. Although heparinized biomaterials, such as cardiac valve substitutes,
16,17 vascular grafts, or conduits,
18,19 have been throughly examined, no reports on these biomaterials in use as pericardial substitutes have been published. Inspired by these encouraging results, we undertook the development of a pericardial patch fixed in different compounds with or without ionically bound heparin.
In the development of biologic pericardial substitutes,
12 it was discovered that the porcine pericardia fixed with an epoxy compound were softer and more pliable that those preserved with glutaraldehyde. Additionally, the pericardia cross-linked with polyepoxy compound appeared whiter and more appealing than their glutaraldehyde counterparts. The superiority in pliability of the pericardia fixed with epoxy compound might be due to the flexible ether joint in the epoxy compound.
16 The experimental results of porcine pericardia with heparinization used to close pericardial defects in dogs are very encouraging. The porcine patches with ionically bound heparin were found to have significantly less adhesion formation than those without ionically bound heparin at each time period assessed. Although minimal adhesions developed in heparinized patches, these were easily dissected. At retrieval, under microscopic observation, minimal chronic active inflammatory changes were observed in the polyepoxy compoundcross-linked patches in comparison with their glutaraldehyde counterparts. Both ePTFE and polyester polymeric patches caused severe epicardial fibrotic adhesion with chronic active inflammatory reactions.
Overall, this observation highlighted the fact that pericardial substitutes bonded with heparin may reduce inflammatory reaction and fibrous formation. However, it is possible that the experimental findings might not be completely reproducible clinically, because the operations performed in this study did not damage the epicardium, cardiopulmonary bypass was not performed, and postoperative bleeding was minimal. The actual clinical condition might be quite different from the study design. Another unsolved problem is that the reduction of heparin on the surface of biologic tissue can reduce its antiadhesive capability. Our results suggested that a small amount of heparin was continuously released over a long period of time. Noishiki and Miyata
14 showed that about 45% of the initial heparin content was released after 1 month's rinsing in vitro, 63% after 2 months, and 76% after 3 months. In this experiment, the adhesion and fibrosis in patches with heparin are progressively increased at 12 weeks. The release rate of heparin in the pericardial substitutes with ionically bound heparin was unknown. Experiments in vitro with a longer follow-up are underway to clarify the release rate of the ionically bound heparin in pericardial substitutes.
Footnotes
*Gore-Tex surgical membrane, W. L. Gore & Associates, Inc., Flagstaff, Ariz. ![]()
References
This article has been cited by other articles:
![]() |
I. Yoshioka, Y. Saiki, K. Sakuma, A. Iguchi, T. Moriya, Y. Ikada, and K. Tabayashi Bioabsorbable Gelatin Sheets Latticed With Polyglycolic Acid Can Eliminate Pericardial Adhesion Ann. Thorac. Surg., September 1, 2007; 84(3): 864 - 870. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Masuda, Y. Saiki, S. Kawatsu, I. Yoshioka, H. Fujiwara, S. Kawamoto, S. Sai, A. Iguchi, N. Sakamoto, T. Ohashi, et al. Trial of new vascular clips for aortic anastomosis in a canine model J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 723 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Sakuma, A. Iguchi, Y. Ikada, and K. Tabayashi Closure of the Pericardium Using Synthetic Bioabsorbable Polymers Ann. Thorac. Surg., November 1, 2005; 80(5): 1835 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Kajihara, M. Eto, Y. Oishi, N. Boku, K. Kuwahara, N. Nishiguchi, C. Kotani, and S. Morita Three-layered synthetic pericardial substitutes reduce postoperative pericardial adhesions J. Thorac. Cardiovasc. Surg., January 1, 2005; 129(1): 18 - 24. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Saha, W. McEwan, F. A Frizelle, and H. Singh Limiting Pericardial Adhesions after Coronary Artery Bypass: Experimental Study Asian Cardiovasc Thorac Ann, March 1, 2004; 12(1): 61 - 64. [Abstract] [Full Text] [PDF] |
||||
![]() |