|
|
||||||||
J Thorac Cardiovasc Surg 1998;115:1287-1291
© 1998 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Read at the Twenty-third Annual Meeting of The Western Thoracic Surgical Association, Napa, Calif., June 25-28, 1997.
Received for publication July 8, 1997. Revisions requested Oct. 10, 1997; revisions received Dec. 1, 1997. Accepted for publication Dec. 30, 1997. Address for reprints: Giovanni Battista Luciani, MD, Division of Cardiac Surgery, University of Verona, O. C. M. Piazzale Stefani 1, Verona, 37126, Italy
Abstract
Objective: The ideal substitute for the diseased aortic valve is yet to be found. For the assessment and comparison of the midterm results after aortic valve replacement with three different types of freehand stentless xenografts, all patients who underwent the operation between October 1992 and April 1997 were reviewed.
Methods: Of 231 patients undergoing aortic valve replacement, 106 patients (group 1) were given the Biocor PSB (Biocor Industria e Pesquisa Ltda, Belo Horizonte, MG, Brazil); 76 patients (group 2) were given the Toronto SPV (St. Jude Medical, Inc., St. Paul, Minn.), and 49 patients (group 3) were given the O'Brien-Angell valve (Bravo Cardiovascular model 300, Cryolife, Inc., Marietta, Ga.). The first two xenografts require inflow and outflow suturelines; the third xenograft needs a single-sutureline implantation. Mean age (70 ± 6 years; 70 ± 7 years; 72 ± 9 years; p = 0.6), prevalence of male sex (56 patients, 53%; 37 patients, 49%; 22 patients, 45%; p = 0.7), of aortic stenosis (72 patients, 68%; 54 patients, 71%; 37 patients, 73%; p = 0.6), and need for associated procedures (51 patients, 48%; 30 patients, 40%; 21 patients, 43%; p = 0.1) were comparable among groups. Mean aortic crossclamp time was shorter in group 3 (96 ± 24 minutes; 100 ± 23 minutes; 88 ± 25 minutes; p = 0.01).
Results: Early deaths were 3 of 106 (3%) in group 1, 2 of 76 (3%) in group 2, and 2 of 49 (4%) in group 3. Follow-up of survivors ranged from 1 to 54 months (mean 32 ± 13 months). Survival at 4 years was 90% ± 3% in group 1, 95% ± 3% in group 2, 85% ± 8% in group 3 (p = 0.3). At 4 years, freedom from valve-related events was 95% ± 6%, 100%, 70% ± 8% (p = 0.004), while freedom from valve deterioration was 99% ± 1%, 100%, 73% ± 8% (p = 0.001), in group 1, 2, and 3, respectively (p = 0.001). At follow-up, reintervention on the xenograft was necessary in one patient (endocarditis) in group 1, none in group 2, and six in group 3 (technical cause, group 3; valve tear, group 2; pannus, group 1). Regression analysis showed O'Brien-Angell type of xenograft to be predictive of valve-related events (p = 0.02), valve deterioration (p = 0.001), and reoperation (p = 0.001) during follow-up.
Conclusions: Midterm survival after stentless aortic valve replacement is good with all three xenografts. Freedom from valve-related events, valve deterioration, and reoperation are excellent with the Biocor PSB or the Toronto SPV stentless valves but less satisfactory with the O'Brien-Angell valve.
The ideal aortic valve substitute is yet to be found. Despite the low valve-related morbidity associated with stented porcine xenografts, the evidence of limited durability and suboptimal hemodynamics, even with the latest generation of bioprostheses, continues to represent major shortcomings.
1,4
The observation of more durable results and favorable hemodynamic performance obtained with freehand when compared with stent-mounted aortic homografts
5 has prompted a revived interest in a new generation of stentless porcine xenografts.
6 Because of the structural similarity with freehand homograft valves, stentless prostheses adapt to the aortic root to reproduce the anatomy of the native aortic valve. The expectation is that by offering optimal hemodynamic performance and reducing the mechanical stress on the valve leaflets, degeneration of the bioprosthesis may be slowed and stentless xenografts may prove more durable than commonly used stented valves.
6
Although only the test of time over 10 years after implantation will prove the value of this hypothesis, initial experience with stentless aortic xenografts already allows identification of some of the limitations with the currently available valve models. We present our clinical experience with aortic valve replacement (AVR) with three types of freehand stentless xenografts in 231 patients.
Patients and methods
Patient population.
In October 1992, a clinical program of AVR with stentless porcine xenografts was started at the University of Verona. Selection criteria for enrollment in the study were the same applied to patients who received stented porcine xenografts, including (1) age older than 65 years, (2) contraindications to oral warfarin therapy, and (3) deliberate request of a biologic valve by the patient. In the latter two cases, a choice among an aortic homograft, a pulmonary homograft, and a porcine xenograft was offered to the patient. Children less than 18 years of age were excluded from the study because they were only offered a pulmonary autograft. When the decision to replace the aortic valve with a xenograft was reached, the ultimate choice between stented (Hancock II, Medtronic, Inc., Minneapolis, Minn.) or freehand stentless valve was left to the surgeon.
Three different models of stentless porcine bioprostheses were used for the study, including (1) group 1: the Biocor PSB valve (Biocor Industria e Pesquisa Ltda, Belo Horizonte, MG, Brazil); (2) group 2: the Toronto SPV valve (St. Jude Medical, Inc., St. Paul, Minn.); and (3) group 3: the O'Brien-Angell valve (Bravo Cardiovascular model 300, Cryolife, Inc., Marietta, Ga.). The choice of the type of stentless valve to implant was done by the surgeon in a nonrandom fashion.
Two hundred thirty-one consecutive patients were included in the study between October 1992 and April 1997. Retrospective review of the patient population showed a predominantly elderly group of subjects with similar demographic variables among the three groups (Table I). Primary indication for AVR was aortic stenosis in all groups, less commonly regurgitation or a combination of the two. Only five patients had prosthetic valve dysfunction as indication for stentless AVR (Table I
).
|
The Toronto SPV valve is made of a single porcine root that is fixed under low pressure with glutaraldehyde. The coronary sinuses are trimmed down to 1.5 to 2.0 mm to form the base of the leaflets, and the outside wall is thinned and covered with a Dacron cloth.
The O'Brien-Angell valve is a composite valve made of three noncoronary leaflets from three different porcine valves, already fixed with glutaraldehyde under near zero pressure. The leaflets are sutured together along the free edges of the aortic wall at the leaflet commissures.
The technique of implantation adopted for the Biocor PSB and Toronto SPV valve consisted of sizing at the sinotubular junction and grafting with an inflow (with interrupted 4-0 multifilament polyester sutures or continuous running 4-0 polypropylene sutures) and an outflow sutureline with running 4-0 polypropylene sutures.
6 Implantation of the O'Brien-Angell valve included sizing at the level of the aortic anulus, with selection of a prosthesis one size larger than measured, and grafting in a supraannular, infracoronary position with running 3-0 polypropylene sutures.
7
Shorter aortic crossclamp and cardiopulmonary bypass times were recorded in group 3 patients as expected because of the simplified implantation technique of the O'Brien-Angell xenograft (Table II). The need for associated procedures was relatively common in our series and consisted primarily of myocardial revascularization operations (Table II
). When a concomitant procedure on the mitral valve was required, an attempt at repair was done whenever feasible. In case of replacement, a stented porcine bioprosthesis was always implanted. Replacement of the ascending aorta in patients with poststenotic dilatation (G5 cm diameter) and unilateral or bilateral carotid thromboendoarterectomy accounted for the remaining associated procedures. Analysis of the number and type of associated operations performed did not reveal any difference among patient groups (Table II
).
|
Statistical analysis.
Continuous variables were expressed as means ± standard deviations, and discrete variables were expressed as percentage. Comparison among groups was with the one-way analysis of variance for continuous variables and Pearson's
2 test for discrete ones. Actuarial life table estimates were constructed with the Kaplan-Meier method and comparison among curves was carried out with the log-rank test. Multivariable analysis was performed with the Cox proportional hazards method
8 to identify risk factors for time-related occurrence of events after AVR, including (1) any event, (2) valve-related events, (3) reoperation, and (4) structural and nonstructural valve deterioration. Definitions of these events were established according to the recently recommended guidelines.
9 The variables entered in the analysis are listed in the appendix.
Results
Survival.
There were seven (3%) early deaths (<30 days or before hospital discharge). Three (3%) patients in group 1 died of low output syndrome
2 and cerebrovascular accident,
1 respectively, 5, 15, and 45 days after the operation. Two (3%) early deaths in group 2 were due to low output syndrome and cerebrovascular accident, 1 and 28 days after operation. Finally, two (4%) patients in group 3 died early of cerebrovascular accident and of sepsis, 30 and 50 days after AVR. Follow-up of survivors ranged from 1 to 54 months (mean 32 ± 13 months). There was a total of nine (4%) late deaths during the study period. Six (6%) patients in group 1 died of malignancy (n = 2), cardiac cause (n = 2), pneumonia (n = 1), and sudden death (n = 1). Only one (1%) patient in group 2 died of malignancy. Two (4%) patients in group 3 died of malignancy and sudden death. Actuarial 1- and 4-year survival was comparable in the three groups as shown in Fig. 1.
|
Valve-related events.
When only valve-related adverse events, including valve-related death, were considered, a significant difference was noted between group 1 and 2 patients, who had similar freedom from events, and group 3, who more commonly experienced valve-related complications (Fig. 2). Mulitvariable analysis disclosed O'Brien-Angell type of xenograft and female gender as incremental risk factors for a valve-related adverse event during the study period (Table III).
|
|
Two patients in group 1 had cerebral embolism, 1 and 2 years respectively after AVR. None of the patients had undergone concomitant carotid endoarterectomy or had been previously diagnosed with cerebrovascular pathologic evidence; thus the genesis of the embolic episodes was classified as valve related.
Only in one patient in group 1 did Streptococcus bovis prosthetic valve endocarditis develop 3 years after operation. The patient underwent replacement of the Biocor xenograft valve with an aortic homograft and had an uneventful recovery, with no replase of infection 10 months after homograft root replacement.
Valve deterioration.
Structural valve deterioration was encountered only in two (4%) patients in group 3 during follow-up. Both women had moderate to severe periprosthetic regurgitation, confirmed by angiography, respectively 5 and 10 months after AVR, which led to reoperation.
Nonstructural valve deterioration was recognized in one (1%) patient in group 1 and in eight (16%) patients in group 3. The patient with Biocor PSB AVR was diagnosed with a peak transprosthetic pressure gradient of 50 mm Hg 6 months after operation, which remained unchanged at follow-up. Of the eight patients with O'Brien-Angell xenografts who had nonstructural valve deterioration, four underwent reoperation (two early and two late) because of the severity of prosthetic valve obstruction and/or worsening symptoms. The remaining four patients have peak pressure gradients between 52 and 65 mm Hg, which has shown slow progression at follow-up, and are currently asymptomatic.
Actuarial freedom from valve deterioration for any cause was significantly lower in patients in group 3 (Fig. 3). Multivariable analysis identified O'Brien-Angell type of valve and female gender as incremental risk factors for valve deterioration during follow-up (Table IV).
|
|
Six (12%) reoperations were necessary in group 3, all because of valve dysfunction. This was identified as structural in two patients with periprosthetic regurgitation. At reoperation, a tear of the porcine aortic wall support of the O'Brien-Angell xenograft was identified in both patients. In one woman the xenograft had to be replaced with a stented xenograft; in the other woman, because of the limited extension of the tear, the stentless xenograft could be reanchored to the native aortic wall. The remaining four patients underwent reoperation for nonstructural deterioration, which had manifested with prosthetic valve obstruction (n = 3) or regurgitation (n = 1). The cause of the dysfunction was possibly technical in three cases, resulting in either inappropriate sizing or incorrect supraanular positioning of the graft; the dysfunction was related to early pannus ingrowth in one.
10 All patients survived replacement of the stentless bioprosthesis with a stented one, which had to be associated to a Nicks aortic anulus enlargement procedure in one patient because of extensive scarring of the aortic root. Actuarial freedom from reoperation on the prosthesis was significantly lower in group 3 (Fig. 4). Regression analysis demonstrated O'Brien-Angell type of valve and female gender to be incremental risk factors for reintervention on the xenograft (Table V).
|
|
When functional performance of the xenograft was analyzed by means of transthoracic echocardiography, no instance of prosthetic valve regurgitation greater than mild was detected in any of the patients, excluding those undergoing reoperation for valve deterioration. On the contrary, different results were found in terms of peak transprosthetic pressure gradients at 6-month follow-up investigation among patient groups (Table VI). Although peak pressure gradients were higher the smaller the diameter of the valve in each group, as might be expected, the values were also higher for the same valve size in patients in group 3 when compared with patients in both group 1 and group 2. Limited sampling in the very small and in the very large valve size categories may account for p values not reaching statistical significance (Table VI).
|
The concept of AVR with a stentless porcine xenograft began with the pioneering work of Binet and associates
11 and O'Brien and associates
12 in the mid-1960s. The recent revival of attention toward stentless valves moved primarily from the observation of optimal hemodynamic performance and longer durability of freehand when compared with stent-mounted aortic homografts.
5 Because mechanical stress imposed on the leaflets of a stented porcine bioprosthesis has been identified as the major cause of late degeneration,
13,15 it was hypothesized that the limitations even with the latest models of stent-mounted porcine valves
1,3 could be overcome with stentless xenografts.
6 Preliminary reports of ideal hemodynamic function, absolutely comparable with aortic homografts
16-18 and the advantage of ready availability in all sizes promoted a rapid diffusion of a large variety of xenograft models into the market,
6,18-21 possibly delaying the collection of significant data to test the durability of these valves.
Unique to the present study is the simultaneous experience with three types of stentless valves in a single institution. A retrospective comparison of clinical and functional results is thus possible.
Replacement of the aortic valve with a stentless xenograft is a more demanding surgical procedure and will result in longer periods of myocardial ischemia when compared with AVR with stented bioprostheses.
16,17,20 In addition, freehand aortic xenografts, which are implanted with inflow and outflow suture lines, will require longer duration of ischemia compared with simplified-implant xenografts, where a single suture line is needed.
22 It is apparent from the data herein that longer aortic crossclamp time does not adversely affect operative survival, which was absolutely comparable to age-matched populations receiving stented bioprostheses,
1-3 as previously shown by others.
19-23 This consideration holds true in spite of the high prevalence of associated disease conditions requiring concomitant surgical management and thus increasing the operative risk, as evident in our patient population. It is noteworthy that cerebrovascular accident was as common a cause of early death as circulatory failure, possibly because of the advanced mean age of the patients and the prevalence of vascular disorders. Midterm survival after stentless xenograft AVR in the elderly was also satisfactory and in agreement with previously published series with younger patients.
23 As observed with operative survival, no difference in terms of midterm survival could be observed among groups of patients who received the three different prosthetic devices. Advanced age could once again account for malignancy being the leading cause of late death. It would thus be unjustified to reject use of Biocor PSB, Toronto SPV, or O'Brien-Angell stentless xenografts on the basis of a presumptive higher risk of operative or late death.
Prevalence of early and late morbidity was overall low, not exceeding 9% of the entire patient population. Complications unrelated to the insertion or presence of the aortic xenograft were comparable in the three subgroups and commonly caused by malignancy or congestive heart failure. When freedom from all, including valve-related, events was analyzed, this seemed satisfactory both in patients who received Biocor PSB and Toronto SPV valves and comparable with estimates presented after AVR with stented bioprostheses.
1-3 However, a significantly poorer outcome was already evident 4 years after implantation of O'Brien-Angell xenografts. The demonstration that the difference, in terms of event-free survival among patient groups, was solely due to the prevalence of valve-related complications is evident from Fig. 2.
Accordingly, logistic regression analysis disclosed the latter type of xenograft as an incremental risk factor for valve-related events during follow-up. Further assessment showed that prevalence of hemorrhage, thromboembolism, and prosthetic valve endocarditis could not account for the notable difference among the three types of xenografts. Indeed, these complications were overall quite uncommon, and no difference among groups could be disclosed. On the contrary, occurrence of valve deterioration, rarely or never encountered respectively after Biocor PSB or Toronto SPV implantation, was prevalent in patients who received O'Brien-Angell valves. Because the test of time for a porcine bioprosthesis is durability 8 to 10 years after implantation,
1,2,3,6 an actuarial freedom from valve deterioration of 73% at 4 years for the latter xenograft represents a disappointingly low figure (Fig. 3
). Once again, the use of the O'Brien-Angell prosthesis was isolated as a risk factor for valve deterioration after stentless AVR. To date this complication has led 6 of 10 patients in group 3 to reoperation. Despite a slow tendency to progress, reintervention in the near future on the remaining four patients with moderate prosthetic valve obstruction cannot be excluded.
Because of the relatively short period of observation, reoperation on the stentless xenografts was generally uncommon and never in relation with calcific degeneration, the well-recognized leading cause of late failure of stented bioprostheses.
24 Aside from a case of replacement of Biocor PSB xenograft for prosthetic valve endocarditis, reoperation was limited to patients who had received the O'Brien-Angell bioprosthesis, making implantation of this device an incremental risk factor for reintervention during the observation period. Structural failure consisting in a tear of the aortic wall support of the O'Brien-Angell composite xenograft was observed in our series.
10 Similar complications were already reported by Hvass and associates
22 and represent elements of concern for a device that is anchored to the aortic root by means of a single suture line. Even though replacement of the xenograft may not always be needed, reintervention must be undertaken whenever a perivalvular regurgitation is observed to verify stability of the prosthesis. Among causes of nonstructural valve deterioration, overgrowth of fibrous tissue was also noted early after AVR with stentless xenografts,
10 as previously described with stented porcine bioprostheses.
25 The most common cause of nonstructural failure of O'Brien-Angell valves leading to reintervention was, however, transprosthetic obstruction. By virtue of its design, this xenograft needs to be implanted in a subcoronary, but supraannular position, to avoid protrusion of the porcine aortic wall support into the left ventricular outflow tract. Despite the apparent simplicity of this concept, technical realization may not always be as immediate. A variety of anatomic and pathologic conditions involving the aortic root will adversely influence the ability to correctly position the bioprosthesis. Most important, a xenograft that is implanted with a single suture line is more flexible than a twosuture line stentless valve and will be more easily distorted when anchored to an asymmetric aortic root. This may cause obstruction because of the protrusion of the aortic wall into the aortic orifice, regurgitation because of the crowding of the leaflets, or both. Our experience with over 200 implantations of freehand stentless xenografts suggests that technical mishaps are not infrequent with the O'Brien-Angell valve, although these are negligible with either the Biocor PSB and the Toronto SPV valve. The present findings are at variance with data reported by O'Brien and associates
19 but faithfully reproduce the series by Hvass and associates,
22 who described unsatisfactory postoperative gradients (i.e., nonstructural dysfunction in the present report) in most of their first 50 implantations, thereby confirming the existence of a steep learning curve.
On the basis of the current experience, reoperation on a failed stentless xenograft does not represent an exceedingly challenging technical task and can be performed with excellent survival (no casualties among the seven replacements of the bioprosthesis).
When the functional properties of stentless aortic xenografts are evaluated in terms of valve stenosis or regurgitation, comparison with commonly used stented bioprostheses appears extremely favorable, as evident from our clinical series, as well as others.
6,16 A wealth of data is already available that describes the optimal functional results after AVR with the Toronto SPV valve, including the documented tendency of the transprosthetic pressure gradients to decrease over time and the effective orifice area to increase over time.
23,26 Resolution of the inflammatory changes of the aortic wall "excluded" by the xenograft and regression of the left ventricular hypertrophy have been implicated to account for these observations. The present experience with the Toronto SPV valve confirms that the reported results
16,17,23,26 are readily reproducible. Information on hemodynamic performance of the Biocor PSB valve is instead limited to one other institution.
21 Comparison with our series is unsound because of the difference in the age (mean age 36 vs 70 years) and indication to AVR (rheumatic valve disease in 64% vs senile calcific degeneration in 70%) of the patient population, which possibly explains the poorer freedom from adverse events in the experience of others.
21 On the basis of our midterm results on a mostly senescent group of patients, the Biocor PSB valve proved as dependable as the Toronto SPV xenograft both in terms of freedom from valve-related morbidity and of functional profile. On the contrary, even when cases labeled as nonstructural deterioration of the prosthesis were excluded from observation, peak pressure gradients across the O'Brien-Angell xenograft tended to be higher in each valve size (Table VI). Because this graft is implanted with a single suture line, however, decrease of the degree of obstruction cannot be expected in the future. Interpretation of these results does not differ from the explanation of cases of nonstructural deterioration. When compared with the more lengthy and complicated technique of grafting of the Biocor PSB and Toronto SPV valve, implantation of the O'Brien-Angell valve appears simple. Our work suggests that greater simplicity may translate into a less predictable early hemodynamic result.
It must be emphasized that the exact relationship between postoperative transprosthetic pressure gradients and symptoms is still undefined. In fact, all patients showed clinical improvement at follow-up that was not significantly different among the three groups.
Limitations.
There are some notable limitations to our study. The present work is a retrospective analysis on a nonrandomized group of consecutive patients. Thus the issue of comparability of patients among the three groups becomes crucial. Another potential limitation may be the decision to select only elderly patients for this operation. Because of the predicted life expectancy of individuals in the seventh or eighth decade of life, the patients at risk 10 years or longer after xenograft valve replacement could decrease independently from the morbidity of the prosthesis. The use of statistical analysis to predict actual rather than actuarial events will then become necessary.
On the other hand, this experience presents some unique aspects. It is the only work that compares three different stentless aortic xenografts at a single institution. This would allow disclosure of different learning curves with different devices, if it were the case. It is a quantitatively sizable series, in that previously published reports rarely exceed 200 patients.
16,23 Most significantly, it offers insight into factors that will predict occurrence of adverse events during a midterm follow-up.
Conclusions
Analysis of over 230 patients undergoing AVR with three types of freehand stentless xenografts showed that this more complex surgical procedure can be performed at a very low operative risk, comparable among the three groups of patients. Actuarial survival is also satisfactory as is freedom from lethal valve-related events. In the present experience, midterm outcome with the O'Brien-Angell valve proved less satisfactory in terms of valve-related morbidity, including structural deterioration of the graft and reoperation. On the contrary, prevalence of adverse events after implantation of either the Biocor PSB and the Toronto SPV was negligible, and hemodynamic performance is optimal. Whether the difference in design of the O'Brien-Angell valve, in and of itself, can justify these results remains to be proved.
Appendix: Discussion
Dr. Edward D. Verrier (Seattle, Wash.). The surgical options in the treatment of aortic valve and aortic root disease have dramatically increased in the last 10 years with the introduction of a variety of new mechanical valves, stented porcine and bovine pericardial bioprostheses, aortic homografts, pulmonary autografts, and now with the introduction of stentless aortic valves. Biomechanical analysis of the composite left ventricular outflow tract, anulus, leaflets, sinuses, and aortic wall, as well as complex computer modeling such as finite element analysis, have suggested that bioprostheses that mimic the structure, anatomy, and physiology of the normal aortic root and valve should decrease the stress, strain, and compliance changes that therefore would equate with a more durable prosthesis without requiring anticoagulation. The short-term tradeoff for the theoretic long-term benefit has to do with the technical ease of insertion of the stentless valves. These valves must be sewn in at the annular level but must also have the posts constructed usually with a second suture line.
Dr. Luciani and colleagues at Verona have implanted in a nonrandomized protocol three different stentless prostheses and have followup data of just over 3 years. Their conclusions are that the Biocor and the Toronto SP valve are similar and that both are superior to the O'Brien-Angell valve in terms of valve-related events, valve degeneration, and reoperation. Operative death and projected actuarial survival from nonlethal valve-related events are similar in all three groups. This is the only study to date comparing three stentless valves in a single institution.
I have a number of concerns, however, about the study. First, the study is nonrandomized, and there certainly is a discrepancy of the size of the three groups. How were the valves selected for individual patients? Was the series concomitant or sequential, and if it was sequential, what order did you trial each valve? How many surgeons were involved, and did all the surgeons insert all the valves?
Dr. Luciani. The series is not sequential, but rather concomitant. The reason why we started with three different valves, which is a very curious experience, was that the Biocor PSB valve was the very first one to be introduced into the market in Europe and Italy, in particular. When the Toronto SPV and the O'Brien-Angell valves became available for the Italian market, we decided to test a valve that was similar to the Biocor PSB and a valve that was completely different in design. Thus most of the experience is simultaneous and not sequential. There were no specific, prospective criteria by which different models were used by the surgeons; this was left to the surgeon. Obviously this introduces an element of bias, and the fact that the study is retrospective also represents an element of bias.
Your third question was relative to the different surgeons; there were three senior surgeons who supervised or directly performed all the procedures. This factor was taken into the multivariable analysis, and it did not turn out to be associated with a different outcome in terms of functional results or valve deterioration.
Dr. Verrier. Since the follow-up is so short, it is impossible for me to believe that the conclusions are anything other than technical considerations rather than being intrinsically wrong with the O'Brien-Angell valve. In fact, in the manuscript, you note that the cause of dysfunction was "possibly technical in at least three cases, resulting in either inappropriate sizing or incorrect supraannular positioning of the graft." At the recent Seventh International Symposium on Cardiac Bioprosthesis in Barcelona where your group reported most of these data, Hvass presented 5 years of follow-up of 366 patients with the O'Brien-Angell valve. They reported two valve-related cardiac deaths, both from endocarditis, and only five reoperations, or less than 1% for nonstructural dysfunction, and none for structural dysfunction. This is in sharp contrast to the present series in which eight patients (16%) were reoperated on for nonstructural dysfunction and two patients (4%) for structural dysfunction. Hvass reported overall 5-year actual survival of 85% and 95% for operative survivors. Would you please comment on the discrepancy between your data and the data from Angell and his group?
Dr. Luciani. Dr. Hvass published a series of papers beginning with 1995, one of which, published in the European Journal of Cardiothoracic Surgery, disclosed a learning curve with the implantation of the O'Brien-Angell valve device. Curiously, the same experience was not reported by Dr. O'Brien, but the person who has the largest experience with the valve is in fact Dr. Hvass in Paris, and what he found was that, in the first 50 valves implanted, the prevalence of "nonstructural" dysfunction possibly because of technical failure was quite high. The naming of the phenomenon was quite different; in fact, he had higher transprosthetic gradients and did not call these nonstructural valve failures as we did, but possibly the phenomenon was similar. The second 50 patients he analyzed had a much lower prevalence of high gradient; in the last 50 patients in whom he performed the operation (this was a series of 150 consecutive patients) he actually could achieve excellent results with minimal prevalence of this phenomenon. This event may relate to the structure of the valve that makes it more prone to technical failure. I do agree with you that our failures are possibly related to technical cause in many of the nonstructural failure groups, which apparently occurred at the moment of implantation and then revealed themselves early on after follow-up.
Dr. Verrier. My final question relates to echocardiographic follow-up in these patients. Because there are no stentless valve series in either Europe, Canada, or the United States that can answer the question of durability, which will be the key to wider acceptance of this type valve, most investigators have focused on the small increase in valve area noted over the first 3 months of implantation and the accelerated decrease in left ventricular wall mass over the first 6 months. This is a particular advantage noted by most investigators with the stentless prosthesis over the stented bioprosthesis. Can you give any data supporting the superiority of any of these three valves in terms of these very definable, objective endpoints?
Dr. Luciani. We do not have any data to support the increase in effective orifice size 3 months after implantation. We have only been following the flow across the valve in terms of peak and mean transprosthetic gradients and the degree of mass regression.
Dr. Thomas A. Pfeffer (Los Angeles, Calif.). During the past 4 years, our group in Los Angeles has implanted 93 Freestyle valves manufactured by Medtronic which, as far as I am aware, is the only stentless valve with imminent release for general use in the United States. At 4 years, approximately 78% of our patients have no aortic insufficiency and about 22% have grade 1 or more aortic insufficiency, none of whom have undergone reoperation. This is essentially unchanged from our initial postoperative echocardiographic findings. My question relates somewhat to the discussion of the Ross procedure earlier this morning that had to do with the importance of valve sizing for this procedure. Didn't you comment on the technique of implantation, whether all of them were subcoronary or modified subcoronary or done as root replacements, and whether this would have any impact on the results with these valves?
Dr. Luciani. I am sorry that, because of time constraints, I could not go into the details of technique of implantations. The three valves that we used cannot be implanted in the different manners that you explained; they can only be implanted as subcoronary, freehand grafts, because they do not retain the native roots; you cannot scallop them and, alternatively, implant them as miniroot, total root, or complete freehand. Both the Biocor PSB and Toronto SPV valve need to be implanted as freehand with inflow and outflow sutures much like freehand homografts.
The O'Brien-Angell valve has a different technique of implantation, which is single suture line; in fact, it is three distinct running suture lines that anchor the valve not at annular level but at supraannular level. We believe that this specific feature of the valve, which is at one time very attractive, at the same time may be somewhat less readily apparent for the surgeon and less ready to be learned by the surgeon in terms of implantation technique.
Dr. Philip W. Wright (Honolulu, Hawaii). Are there disqualifying characteristics of systolic left ventricular function preoperatively in the selection of patients that influence the outcome or, in your practice, would influence the type of patients that you operate on?
Dr. Luciani. This is a series of nonselected patients in the way they were not prospectively randomized, so that left ventricular function was taken into consideration, and the low number of cardiac deaths compared with series that used stented procedures may explain the difference of death connected with device implantation.
Appendix
Varibles entered in the multivariable analysis with the Cox proportional hazards method as covariates included: age, gender, indication to operation, endocarditis as a cause of aortic insufficiency, presence of a bicuspid aortic valve, associated coronary artery disease, name of surgeon, need for associated procedures, size of xenograft valve, type of prosthetic device, duration of aortic crossclamp time, use of blood cardioplegia, and immediate postoperative transaortic peak pressure gradient.
Acknowledgments
We thank Luisa Zanolla, MD, for the invaluable help in the statistical analysis of the data.
Footnotes
References
This article has been cited by other articles:
![]() |
D. Pavoni, L. P. Badano, F. Ius, E. Mazzaro, R. Frassani, S. Gelsomino, and U. Livi Limited Long-Term Durability of the Cryolife O'Brien Stentless Porcine Xenograft Valve Circulation, September 11, 2007; 116(11_suppl): I-307 - I-313. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Martinovic, I. Farah, M. Everlien, S. Lindemann, I. Knez, T. Wittlinger, H. Greve, and P. Vogt Eight-year results after aortic valve replacement with the CryoLife-O'Brien Stentless Aortic Porcine Bioprosthesis J. Thorac. Cardiovasc. Surg., September 1, 2005; 130(3): 777 - 777. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Polvani, F. Barili, L. Dainese, M. Muratori, M. Porqueddu, A. Sala, and P. Biglioli Long-Term Results After Aortic Valve Replacement With the Bravo 400 Stentless Xenograft Ann. Thorac. Surg., August 1, 2005; 80(2): 495 - 501. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Martinovic, M. Everlien, I. Farah, T. Wittlinger, I. Knez, H. Greve, and P. Vogt Midterm Results After Aortic Valve Replacement With a Stentless Bioprosthesis Aortic Valve Ann. Thorac. Surg., July 1, 2005; 80(1): 198 - 203. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. B. Doty and J. R. Doty Stentless Aortic Valve Replacement: Bioprostheses Card. Surg. Adult, January 1, 2003; 2(2003): 889 - 898. [Full Text] |
||||
![]() |
G. B. Luciani, F. Santini, S. Auriemma, L. Barozzi, P. Bertolini, and A. Mazzucco Long-term results after aortic valve replacement with the Biocor PSB stentless xenograft in the elderly Ann. Thorac. Surg., May 1, 2001; 71 (2007): S306 - S310. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |