|
|
||||||||
J Thorac Cardiovasc Surg 1998;116:628-632
© 1998 Mosby, Inc.
CARDIOPULMONARY SUPPORT AND PHYSIOLOGY |
Ann Arbor, Mich
Funded in part by National Institutes of Health RO1 grant HD15434.
Presented in part at the Extracorporeal Life Support Organization Annual Meeting, Detroit, Mich, September 1997.
Received for publication April 6, 1998. Revisions requested June 17, 1998; revisions received June 30, 1998. Accepted for publication July 1, 1998. Address for reprints: Robert J. Schreiner, MD, University of Michigan Medical Center, 1500 E Medical Center Dr, 2920 Taubman Center, Box 0331, Ann Arbor, MI 48109-0331.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Because of the high extracorporeal flows frequently required during venovenous ECLS (50-100 mL/kg per minute), currently cannulation of the venous system involves conduits of large caliber. Such adequate sites include the right atrium and the common femoral/common iliac system. These areas can be readily accessed via the right internal jugular and right common femoral veins.
Oxygen supplementation during ECLS depends on extracorporeal flow rate and the amount of oxygen available for systemic distribution. In contrast to venoarterial bypass, venovenous ECLS requires both drainage and infusion from prepulmonary capacitance veins. Because of this, a fraction of highly oxygenated ECLS blood is returned to the circuit and is not available for systemic use. Several factors govern extracorporeal flow and recirculation, including cannula position, venous diameter, thoracoabdominal pressure differential, and flow direction.
The cumulative effects of these variables may significantly differ when atrial drainage and femoral reinfusion (AF ECLS) is compared with femoral drainage with atrial reinfusion (FA ECLS) and may result in distinctly different levels of support achieved during venovenous ECLS. Despite the routine use of opposite flow directions in the United States
3 and Europe (using FA access and a long atrial reinfusion catheter
4,5), no direct comparison between these 2 ECLS modes has been performed. This study was undertaken to prospectively compare the amount of extracorporeal support provided by FA and AF flow during venovenous ECLS using our standard access.
| Methods |
|---|
|
|
|---|
|
After initiation of ECLS, standard pressurecontrolled rest ventilator settings were instituted.
3 These included pressure-controlled inverse-ratio ventilation at a respiratory rate of 6 breaths/min with peak inspiratory pressures of 30 cm H2O at an end-expiratory pressure of 10 cm H2O (
P = 20 cm H2O). Patients were pharmacologically paralyzed. Fiberoptic pulmonary artery (PA) catheters (Oximetrix, Abbott Critical Care Systems, North Chicago, Ill) were placed in all patients to allow continuous SvO2 monitoring during the study. An ultrasonic flow probe (Transonic Systems Inc, Ithaca, NY) was placed on the infusion limb of the ECLS circuit to obtain continuous flow measurements.Immediately after cannulation, bypass was conducted in an AF direction. Flow direction was briefly reversed to FA by changing clamp position. Patients were initially maintained on the form of bypass that provided the highest PA saturation during this preliminary comparison. Regardless of initial flow direction, formal measurements were not performed until stable venovenous bypass was attained; this was defined as no change in PA SvO2 for 15 minutes. The following three measurements were collected on the direction of bypass originally selected.
Highest ECLS flow
Extracorporeal flow was gradually advanced by increasing the speed of the roller pump until the compliance chamber on the drainage limb of the circuit intermittently collapsed. Speed was then reduced to allow continuous uninterrupted flow. Flow, measured ultrasonically from the infusion limb of the circuit, was recorded in liters per minute as the maximal flow and corrected for body weight (milliliters per kilogram per minute).
Highest PA SvO2
ECLS flow was increased until the highest obtainable PA SvO2 was reached. This point occurred invariably at the maximum ECLS flow.
Amount of extracorporeal flow required to maintain equivalent PA saturation
ECLS flow was adjusted to attain a patient-specific PA oxygen saturation (based on the level of PA SvO2 achieved with initial bypass) obtainable with both modes of bypass.
Circuit flow was then reversed by changing clamp position on the modified bridge (Fig 1
). After stable bypass was achieved in the opposite direction, a similar set of data was collected.
The patient continued to receive the form of bypass that provided the higher PA SvO2. This procedure was performed once in each patient, within the first 24 hours of ECLS. Two-tailed paired t tests compared data from the 2 bypass directions. Data are expressed as mean ± standard deviation.
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Historically, our practice has been to direct ECLS flow in an AF direction, draining deoxygenated blood from the atrium and reinfusing it into the femoral vein. This is predicated on the notion that the large caliber of the atrial chamber may afford greater drainage and therefore translate into higher flows when compared with drainage from the smaller femoral vein. When AF ECLS was initially chosen as the primary direction to conduct flow, ventilator rest settings included low-volume ventilation with pressure limits and short inspiratory times.
Current ECLS rest settings routinely use pressure-controlled inverse-ratio ventilation in an effort to maintain pulmonary recruitment. The increased intrathoracic pressures that accompany this technique may significantly reduce atrial drainage and, by reciprocally shunting blood toward the lower extremities, thereby increase femoral drainage. These pressure relationships may account for our finding FA ECLS to provide significantly higher flows than AF ECLS.
We used changes in PA SvO2 as an approximation of ECLS support. PA SvO2 reflects the mixing of oxygen saturations that occurs from returning systemic blood combining with that of the ECLS circuit. Additionally, PA SvO2 is not directly affected by contributions from the native lungs, as is arterial saturation. Although this value does not represent true native SvO2 because of extracorporeal augmentation, we believe it does accurately reflect relative changes in ECLS support. PA SvO2 is routinely relied on at our institution to assess adequacy of support during venovenous ECLS.
During venovenous ECLS, both drainage and infusion occur via the same prepulmonary vasculature, and therefore a certain fraction of ECLS affluent unavoidably consists of recently delivered extracorporeally oxygenated blood. The magnitude of this recirculation directly affects peripheral oxygen delivery. During traditional AF ECLS, flow is directed from the infusion cannula directly toward the drainage cannula, and the systemic distribution of ECLS effluent necessitates passage of oxygenated blood past the orifice of the drainage cannula. Femoral drainage with atrial infusion (FA bypass) may reduce recirculation by directing oxygenated blood through the tricuspid valve for subsequent systemic distribution. Although not directly measured, reduced recirculation likely accounted for the significantly higher PA SvO2 observed with FA bypass, as well as the provision of similar levels of support at lower circuit flow rates.
Although for a given paired comparison FA bypass usually provided higher PA SvO2, higher maximal flow, and required less flow to maintain similar support, this was not universal. As governed by Poiseuille's law, flow is greatly influenced by cannula diameter and length. All patients with equivalent cannulas in both access sites demonstrated higher flows in the FA direction. The 1 patient dropped from analysis because of cannula size discrepancy (atrial 23F, femoral 21F) had predictably higher flows with AF bypass. However, even with lower flows, PA SvO2s in this patient were significantly higher in the FA direction. Another patient (included in data analysis) demonstrated higher PA SvO2 and required less flow to maintain SvO2 with AF bypass, while having lower maximal flow in this direction. Radiography revealed appropriate cannula placement in this patient. Despite this isolated exception, FA bypass provided significantly higher SvO2, higher flow, and required less flow to maintain an equivalent SvO2 in the group as a whole.
Although ECLS may be required for weeks in patients with respiratory failure, we compared differences in flow direction only on the first day of bypass. It is unclear how flow and recirculation relationships will change over time. Although not tested in this study, it is possible that FA flow may be compromised to a greater degree than AF flow by intravascular volume depletion, thereby limiting the application of the current findings.
To our knowledge, no complication resulted from the conduct of FA bypass. Routine circuit examinations did reveal a higher incidence of thrombus formation adjacent to Y connectors. Inasmuch as the modified bridge contains 2 more of these connectors than the standard bridge, it may increase the potential for embolic events. In routine ECLS any clot that dislodged during stable flow in either direction would be directed at the oxygenators. However, flow reversal under these circumstances could have resulted in free intravascular clot being directed toward the native pulmonary circulation. It is for this reason that flow reversals were not routinely performed after the first day of bypass. The modified bridge is more complex than the standard one and requires the manipulation of 2 tubing clamps rather than 1. Because of the added tubing, the bridge is larger and can be more difficult to transport without kinking. Regardless, after adequate staff training we found the device to be manageable and to require little additional care.
In summary, FA support during adult venovenous ECLS provided higher maximum flow, higher PA SvO2, and required less flow to maintain a given SvO2 than did traditional AF bypass. Although we no longer routinely use the modified bridge, it permitted safe conversion between the 2 modes of support and allowed for direct comparisons within individual patients. On the basis of this study, it has become our practice to initiate flow in the FA direction in adult venovenous ECLS.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. C. McGee Jr., P. M. McCarthy, and N. Moazami Temporary Mechanical Circulatory Support Card. Surg. Adult, January 1, 2008; 3(2008): 507 - 534. [Full Text] |
||||
![]() |
S. R. Carr, J. P. Cantor, A. S. Rao, T. V. Lakshman, J. E. Collins, and J. S. Friedberg Peritoneal perfusion with oxygenated perfluorocarbon augments systemic oxygenation. Chest, August 1, 2006; 130(2): 402 - 411. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Lequier Extracorporeal Life Support in Pediatric and Neonatal Critical Care: A Review J Intensive Care Med, September 1, 2004; 19(5): 243 - 258. [Abstract] [PDF] |
||||
![]() |
N. Moazami and P. M. McCarthy Temporary Circulatory Support Card. Surg. Adult, January 1, 2003; 2(2003): 495 - 520. [Full Text] |
||||
![]() |
J. B Zwischenberger and S. K Alpard Artificial lungs: a new inspiration Perfusion, July 1, 2002; 17(4): 253 - 268. [Abstract] [PDF] |
||||
![]() |
T. Pranikoff, R. B. Hirschl, R. Remenapp, F. Swaniker, and R. H. Bartlett Venovenous Extracorporeal Life Support Via Percutaneous Cannulation in 94 Patients Chest, March 1, 1999; 115(3): 818 - 822. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |