J Thorac Cardiovasc Surg 1999;117:302-309
© 1999 Mosby, Inc.
SURGERY FOR ADULT CARDIOVASCULAR DISEASE |
From the Department of Cardiovascular and Thoracic Surgery,a the Division of Cardiovascular Medicine,c and the Department of Anesthesia,e Stanford University School of Medicine, Stanford, Calif, Cardiac Surgeryf and Cardiology Sections,d Department of Veterans Affairs Medical Center, Palo Alto, Calif, and the Department of Cardiovascular Physiology and Biophysics,b Research Institute of the Palo Alto Medical Foundation, Palo Alto, Calif.
Presented in part at the American Heart Association Seventieth Scientific Sessions, Orlando, Fla, Nov 9-12, 1997.Supported by grants HL-29589 and HL-48837 from the National Heart, Lung and Blood Institute and the Veterans Administration Medical Research Service. Drs Glasson, Green, Dagum, and Komeda are Carl and Leah McConnell Cardiovascular Surgical Research Fellows. Dr Glasson is also a Katharine McCormick Scholar and recipient of The Thoracic Surgery Foundation Research Fellowship Award.
Received for publication Jan 23, 1998. Revisions requested April 2, 1998.; revisions received July 27, 1998. Accepted for publication Sept 29, 1998. Address for reprints: D. Craig Miller, MD, Department of Cardiovascular and Thoracic Surgery, Falk Cardiovascular Research Center, Stanford University School of Medicine, Stanford, CA 94305-5247.
| Abstract |
|---|
|
|
|---|
.001). Mitral annular area was fixed in the Physio-Ring group (4.6 ± 0.1 cm2) and, surprisingly, also static in the Duran ring group (4.8 ± 0.1 cm2; P = .26 vs Physio-Ring). Furthermore, mitral annular 3-dimensional shape changed in the no-ring group during the cardiac cycle, but not in the Physio-Ring or Duran groups.| Introduction |
|---|
|
|
|---|
The functional importance of annuloplasty ring flexibility, however, remains unclear. Experimental data from our laboratory in normal sheep suggest that preservation of the natural dynamic changes in mitral annular size and shape during the cardiac cycle would be the most physiologic approach,
4 implying that flexible annuloplasty rings may be beneficial in terms of preservation of natural mitral annular dynamic motion. Recent clinical investigations from Japan have suggested that mitral annular configuration and dynamics are more physiologic in patients with a flexible Duran ring than in those who received a rigid Carpentier ring.
5-7 Mitral annular area in these studies, however, was estimated from planimetric measurements of 2-dimensional (2D) circumferential annular projections reconstructed from echocardiographic long-axis image planes, without the benefit of measuring discrete or consistent mitral annular sites or three-dimensional (3D) data.
To investigate the putative benefit of maintaining "physiologic" annular flexibility after annuloplasty in terms of annular 3D dynamics, we used the myocardial marker technology
8 in a conscious, closed-chest ovine preparation to analyze and compare continuous 3D mitral annular motion in 3 groups of animals: normal sheep without an annuloplasty ring (group N), animals with a flexible Duran ring (group D), and those with the semi-rigid CE Physio-Ring (group P).
| Methods |
|---|
|
|
|---|
During cardiopulmonary bypass with the heart arrested, 8 miniature tantalum radiopaque markers were sutured equidistantly around the circumference of the mitral anulus, 1 near each commissural area (markers 1 and 5) and 3 along the perimeters of the anterior (markers 6-8) and posterior (markers 2-4) leaflets, as shown in Fig. 1. The animals were then randomized to 1 of 3 groups: no ring, group N; Physio-Ring, group P; or Duran ring, group D. Ring sizes were selected according to the size of the anterior leaflet and the intertrigonal distance using clinically accepted standard methods. In group P, all 7 animals received a 28 mm Physio-Ring; in group D, 5 animals received a 31 mm Duran ring, and 2 received a 29 mm Duran ring. A micromanometer pressure transducer (P4.5-X6, Konigsberg Instruments, Inc, Pasadena, Calif) was placed in the LV chamber via the apex. After marker and ring placement, mitral valve competence was demonstrated in all animals by infusing saline solution into the ventricle and by epicardial color Doppler echocardiography.
|
For all data acquisition runs, the hearts were in normal sinus rhythm and ventilation was arrested briefly at end-expiration. All hemodynamic and biplane videofluoroscopic data recordings were obtained during steady-state conditions and over a physiologic range of peak LV systolic pressures induced by vena caval occlusion. Data sequences containing premature ventricular contractions were discarded and subsequently repeated.
All animals received humane care in compliance with the "Principles of Laboratory Animal Care" formulated by the National Society for Medical Research and the "Guide for Care and Use of Laboratory Animals" prepared by the National Academy of Sciences and published by the National Institutes of Health (DHEW [NIH] Publication 85-23, revised 1985). This study was approved by the Stanford Medical Center Laboratory Research Animal Review Committee and conducted according to Stanford University policy.
Data acquisition
Images were acquired with the animal in the right lateral decubitus position with a Philips Optimus 2000 biplane lateral ARC 2/Poly DIAGNOST C2 system (Philips Medical Systems, North America Company, Pleasanton, Calif) with the image intensifiers in 9-inch fluoroscopic mode. Data from the 2 x-ray views were digitized and processed to yield the 3D coordinates of each marker every 16.7 ms. Specific details of the data acquisition and analysis steps have been described previously.
9 Using myocardial markers for the 3D assessment of cardiac dynamics has some minor limitations; this technique allows accurate and reproducible determination of marker position with a mean overall error of only 0.1 ± 0.6 mm.
11 All annular markers were sutured in similar locations regardless of whether a ring was placed or not.
Data analysis
Data from 2 consecutive steady-state beats for all animals were averaged and analyzed.
Hemodynamics
For each cardiac cycle, the time of end-diastole was defined as the videofluoroscopic frame containing the beginning of positive deflection of the electrocardiographic (ECG) voltage (R wave), and end-systole was defined as the videofluoroscopic frame immediately (16.7 ms) before the time of peak negative LV dP/dt.
Systolic LV function
Instantaneous LV volume was calculated with a multiple elliptical cylindrical/cone model as described previously.
9 LV end-systolic pressure (Pes) and volume (Ves) during preload reduction were analyzed with the use of an iterative computer algorithm to define the end-systolic pressure volume relationship to assess global LV systolic performance.
12 Least-squares linear regression was used to fit a line of the form
Pes = Ees (Ves - V0)
to these end-systolic points, where Ees and V0 are the slope and volume axis intercept of the end-systolic pressure volume relationship, respectively.
Mitral annular geometry
Mitral annular area, perimeters, and dimensions were computed from the 3D marker coordinates, without assuming circular or planar geometry. To compute mitral annular area, the center of mass of all 8 annular markers was determined as the mitral annular centroid. The 8 mitral annular markers and centroid were then used to divide the mitral anulus into 8 triangular "pie-pieces" whose individual areas were added together to compute overall mitral annular area. Mitral annular area was also calculated as the sum of 6 triangular areas derived from all possible triplets of annular marker coordinates to determine whether method of area calculation would alter the area calculations. Because this latter calculation correlated well with the former calculation using the mitral annular centroid (r = 0.97 [group N], 0.97 [group P], and 0.83 [group D]), only areas calculated with the use of the mitral annular centroid are presented in this report. The mitral annular muscular region length was calculated as the sum of the 6 contiguous mitral annular segment (mostly posterior) lengths between annular markers 8 and 6 (Fig. 1
). The septal-lateral mitral annular dimension was calculated as the linear distance in 3D space between annular markers 3 and 7 (middle of the anterior mitral anulus to a corresponding point on the posterior anulus), and the commissure-commissure mitral annular dimension was computed as the linear distance in 3D space between annular markers 1 and 5 (Fig. 1
).
The shape of the entire mitral valve and subvalvular apparatus was also analyzed with the use of a 3D coordinate system defined with the origin at the midpoint of the commissural markers (markers 1 and 5) with the negative z-axis directed through the LV apical marker (ie, along the LV long axis). This reference system allowed description of the distance above or below the annular origin of any marker in 3D coordinates; positive z-coordinates were toward the atrium and negative z-coordinates were closer to the LV apex. To quantitate further the mitral annular shape changes occurring during the cardiac cycle, the z-coordinates of each mitral annular marker at end-diastole were compared with those at end-systole to determine whether ring annuloplasty altered the presence and behavior of the previously described mitral annular "saddlehorn" configuration.
4,9
Statistical analysis
All data are reported as mean ± 1 standard error of the mean (±1 SEM) unless otherwise stated. Hemodynamic and marker-derived data from 2 consecutive steady-state beats in all 7 hearts for each group were time-aligned at the upstroke of the R wave of the ECG (end-diastole). The mean and SEM for each variable were computed for the 14 beats in each group at end-diastole and at 16 time samples (267 ms) before and after end-diastole, such that a curve could be generated depicting the behavior of any variable throughout 1 cardiac cycle (approximately 534 ms in this experiment). Comparison of hemodynamic and geometric variables between groups was performed by means of repeated-measures analysis of variance; when indicated by a significant F statistic (P
.05), specific differences were isolated using the Bonferroni correction for multiple comparisons. Mitral annular dimensions at specified time points in the cardiac cycle were also compared within each group using the Student t test for paired observations.
| Results |
|---|
|
|
|---|
Hemodynamics
Data were obtained for all 3 groups. No mitral regurgitation was present in any animal, as evidenced by transthoracic color Doppler echocardiography at the time of data acquisition. Average heart rate for all sheep did not differ between groups, with an overall mean of 101 ± 8 min1 (mean ± 1 SD). End-diastolic LV pressure, LV ejection fraction, LV stroke work, and LV Ees were not statistically different between groups (Table I), indicating that all animals had similar global LV systolic function despite the presence or type of annuloplasty ring.
|
.001) in very early systole. As expected, mitral annular area did not change during the cardiac cycle in the Physio-Ring group (fixed at 4.6 ± 0.1 cm2). Surprisingly, mitral annular area did not change during the cardiac cycle in the Duran ring group either, also being fixed at the same size as that in the Physio-Ring group (4.8 ± 0.1 cm2 vs 4.6 ± 0.1 cm2, P = .26).
|
.001), as previously reported.
|
.001 vs maximum), and the commissure-commissure dimension decreased during both presystole and systole to a minimum of 33.9 ± 0.5 mm (P
.001 vs maximum), as described previously.
|
|
| Discussion |
|---|
|
|
|---|
The experimental model in this report used a system that allowed accurate beat-by-beat calculation of mitral annular area and shape in true 3D space, independent of any floating internal reference system, for example, those used for echocardiographic or angiographic measurements. It is important to note that this method of data acquisition may explain the differences between our current observations and the findings reported previously by Okada and Yamaura and their colleagues.
5-7
All animals studied were healthy adult sheep, and none had evidence of hemodynamically significant mitral regurgitation by transthoracic Doppler echocardiography at the time of data collection. All data were obtained with the animals in normal sinus rhythm, with equivalent amounts of autonomic blockade in all groups. Hemodynamics and LV systolic mechanics, including heart rate, LV end-diastolic pressure, LV ejection fraction, LV stroke work, and LV elastance, were not significantly different between groups, indicating similar global LV function. Thus any observed differences in mitral annular dynamics between the groups cannot be explained on the basis of differences in LV systolic performance.
In group N (the no ring group), mitral annular area decreased during the cardiac cycle as reported previously.
4 This area reduction was absent in the Physio-Ring group, as expected; although the Physio-Ring is called a "semi-rigid" ring, its more flexible posterior elements are constructed of Elgiloy bands separated by polyester film strips, making it relatively rigid compared to the totally flexible Duran ring. Consequently, the lack of changes in annular size and shape noted in the Physio-Ring group were not surprising. On the other hand, the total flexibility of the Duran ring has been thought to be advantageous because it could theoretically change its size and shape during the cardiac cycle as would the natural anulus. Surprisingly, the results from this study do not confirm this concept. In the Duran ring group, there was also no change in annular area during the cardiac cycle. Moreover, both ring groups showed absence of posterior (or muscular) annular shortening, as well as the physiologic mitral annular shape change detected by the septal-lateral and commissure-commissure annular dimensions. Thus these observations indicate that ring annuloplasty, whether totally flexible or semi-rigid, freezes the anulus in a fixed size and shape; neither ring preserved mitral annular natural flexibility.
The previously described anterior annular "saddlehorn" shape, albeit a subtle one, was maintained in both ring groups. More important, the increase in height of this anterior mitral annular "saddlehorn" during LV systole was sustained to an equivalent extent in both ring groups. This systolic excursion may be important in minimizing LV outflow tract impedance during ejection,
9 which others have suggested could be impeded by rigidity of the anterior mitral anulus.
14,20 These current data do not support this speculation, because the normal upward displacement of the anterior anulus was not altered by the presence of either type of ring.
It is important to note that fixation of the anulus with either type of ring reduced its size significantly, which is one of the clinical goals when an annuloplasty ring is being used. The minimum mitral area observed at any time during the cardiac cycle in group N (no ring) was nearly 50% larger than the fixed annular area in sheep with either type of ring. Additionally, this annular area reduction was nearly equivalent between ring groups, indicating that sizing techniques used for both types of rings were comparable. This substantial annular size reduction after annuloplasty may actually explain the lack of area and shape change during the cardiac cycle we observed in the ring groups. Most mitral annular size reduction in sheep actually occurs before ventricular contraction
4 when the anulus physiologically becomes smaller during atrial systole. Consequently, any additional reduction in mitral annular size, once it is fixed and smaller than the physiologic minimum, would likely require a force far greater than that generated by ventricular contraction. Furthermore, because of the length-tension relationship of cardiac muscle, the shorter muscle fiber lengths in a smaller anulus means that it is likely to generate less annular force. Finally, the lack of differences in mitral annular dynamics seen between the 2 ring groups could also possibly be attributed to the sutures used to implant the rings. The interrupted horizontal mattress sutures placed closely together around the entire circumference of the ring must be associated with some degree of "rigidizing" effect on the mitral anulus.
This study of both flexible and semi-rigid annuloplasty rings showed identical effects on mitral annular dynamics, that is, abolishment of annular flexibility independent of ring type. As such, maintenance of annular flexibility probably does not have any major clinical importance in the overall salutary effects of ring annuloplasty. Recent studies from our laboratory have shown that early systolic incomplete mitral leaflet closure resulting from acute posterolateral wall LV ischemia is associated primarily with annular enlargement without changes in LV geometry, leaflet prolapse, or apical displacement of any part of the mitral apparatus.
21 As such, annuloplasty should be useful in the treatment of patients with ischemic mitral regurgitation: Reducing the size of the anulus brings the leaflet edges closer together, thereby allowing more coaptation to occur. Data from the current experiment support this concept; furthermore, these observations indicate that mitral annular size reduction, independent of maintenance of annular flexibility, is the primary corrective mechanism. Ongoing studies in our laboratory have been designed to answer the question of whether or not annular area reduction alone (regardless of type of ring used) is sufficient to prevent the development of ischemic mitral regurgitation.
Limitations
The use of myocardial markers for 3D assessment of cardiac dynamics has some minor inherent limitations. This technology allows accurate and reproducible determination of marker position, with a mean overall error of only 0.1 ± 0.6 mm every 16.7 ms.
11 Additionally, marker technology allows tracking of discrete sites on the mitral apparatus in 3D space with respect to a fixed external laboratory coordinate system. Such a fixed external coordinate system is not possible with sonomicrometry, which usually is based on an internal reference system. Marker technology does require suturing of small tantalum coil markers to the area of interest, in this case the mitral anulus. The possibility of annular markers interfering with normal mitral annular motion should be minimal, however, because the markers are very small, weighing less than 20 mg each. The small size of the tantalum coils is another advantage over sonomicrometry, which uses larger and heavier crystals requiring wire attachments.
The main limitation in this study is the use of an animal model. Because the fiber architecture of the ovine mitral anulus may not be identical to that of human beings, these results may possibly not be directly applicable to man, especially patients with impaired LV systolic function. One would expect less dynamic annular motion if depressed LV systolic performance is present; thus a similar experiment in impaired hearts might show less pronounced differences between the no-ring and the ring groups. Further studies of mitral annular motion, leaflet dynamics, and mitral regurgitation during myocardial ischemia are currently underway in our laboratory to determine the effects of ring annuloplasty on overall mitral apparatus function in the setting of a clinically relevant ischemic insult.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. P. Ryan, B. M. Jackson, H. Hamamoto, T. J. Eperjesi, T. J. Plappert, M. St. John-Sutton, R. C. Gorman, and J. H. Gorman III The Influence of Annuloplasty Ring Geometry on Mitral Leaflet Curvature Ann. Thorac. Surg., September 1, 2008; 86(3): 749 - 760. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Nguyen, A. Cheng, F. A. Tibayan, D. Liang, G. T. Daughters, N. B. Ingels Jr., and D. C. Miller Septal-lateral annnular cinching perturbs basal left ventricular transmural strains Eur. J. Cardiothorac. Surg., March 1, 2007; 31(3): 423 - 429. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. T. Chan, C. S. Chiu, L.-C. Cheng, and T. W. Au Mitral Valve Annuloplasty With Carpentier-Edwards Physio Ring: Mid-term Results Asian Cardiovasc Thorac Ann, October 1, 2006; 14(5): 382 - 386. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Cheng, T. C. Nguyen, M. Malinowski, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Effects of Undersized Mitral Annuloplasty on Regional Transmural Left Ventricular Wall Strains and Wall Thickening Mechanisms Circulation, July 4, 2006; 114(1_suppl): I-600 - I-609. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Fukuda, R. Grimm, J.-M. Song, T. Kihara, M. Daimon, D. A. Agler, B. L. Wilkoff, A. Natale, J. D. Thomas, and T. Shiota Electrical Conduction Disturbance Effects on Dynamic Changes of Functional Mitral Regurgitation J. Am. Coll. Cardiol., December 20, 2005; 46(12): 2270 - 2276. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, J. R. Glasson, D. T. Lai, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Annular Height-to-Commissural Width Ratio of Annulolasty Rings In Vivo Circulation, August 30, 2005; 112(9_suppl): I-423 - I-428. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Tibayan, F. Rodriguez, F. Langer, M. K. Zasio, L. Bailey, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Does septal-lateral annular cinching work for chronic ischemic mitral regurgitation? J. Thorac. Cardiovasc. Surg., March 1, 2004; 127(3): 654 - 663. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Dayan, A. Oliker, R. Sharony, F. G. Baumann, A. Galloway, S. B. Colvin, D. C. Miller, and E. A. Grossi Computer-generated three-dimensional animation of the mitral valve J. Thorac. Cardiovasc. Surg., March 1, 2004; 127(3): 763 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Sharony, P. C. Saunders, A. Nayar, E. McAleer, A. C. Galloway, J. Delianides, C. F. Schwartz, R. M. Applebaum, I. Kronzon, S. B. Colvin, et al. Semirigid partial annuloplasty band allows dynamic mitral annular motion and minimizes valvular gradients: an echocardiographic study Ann. Thorac. Surg., February 1, 2004; 77(2): 518 - 522. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Tibayan, F. Rodriguez, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Paneth Suture Annuloplasty Abolishes Acute Ischemic Mitral Regurgitation but Preserves Annular and Leaflet Dynamics Circulation, September 9, 2003; 108(90101): II-128 - 133. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, D. T. Lai, F. Tibayan, D. Liang, F. Rodriguez, G. T. Daughters, P. Dagum, N. B. Ingels Jr, and C. Miller Annular Versus Subvalvular Approaches to Acute Ischemic Mitral Regurgitation Circulation, September 24, 2002; 106(12_suppl_1): I-27 - I-32. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, P. Dagum, D. T. Lai, F. Tibayan, D. Liang, G. T. Daughters, M. Hayase, N. B. Ingels Jr, and D. C. Miller Will a Partial Posterior Annuloplasty Ring Prevent Acute Ischemic Mitral Regurgitation? Circulation, September 24, 2002; 106(12_suppl_1): I-33 - I-39. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, D. T. Lai, F. A. Tibayan, P. Dagum, G. T. Daughters, D. Liang, N. B. Ingels Jr, and D. C. Miller Hemodynamic performance of an unstented xenograft mitral valve substitute J. Thorac. Cardiovasc. Surg., September 1, 2002; 124(3): 541 - 552. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, D. T. Lai, F. Tibayan, D. Liang, G. T. Daughters, P. Dagum, N. B. Ingels Jr, and D. C. Miller Septal-lateral annular cinching abolishes acute ischemic mitral regurgitation J. Thorac. Cardiovasc. Surg., May 1, 2002; 123(5): 881 - 888. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Dagum, T. Timek, G. R. Green, G. T. Daughters, D. Liang, N. B. Ingels Jr, and D. C. Miller Three-dimensional geometric comparison of partial and complete flexible mitral annuloplasty rings J. Thorac. Cardiovasc. Surg., October 1, 2001; 122(4): 665 - 673. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. L. Nielsen, T. A. Timek, D. T. Lai, G. T. Daughters, D. Liang, J. M. Hasenkam, N. B. Ingels, and D. C. Miller Edge-to-Edge Mitral Repair: Tension on the Approximating Suture and Leaflet Deformation During Acute Ischemic Mitral Regurgitation in the Ovine Heart Circulation, September 18, 2001; 104(90001): I-29 - 35. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek and D. C. Miller Experimental and clinical assessment of mitral annular area and dynamics: what are we actually measuring? Ann. Thorac. Surg., September 1, 2001; 72(3): 966 - 974. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, S. L. Nielsen, D. Liang, D. T. Lai, P. Dagum, G. T. Daughters, N. B. Ingels Jr., and D. C. Miller Edge-to-edge mitral repair: gradients and three-dimensional annular dynamics in vivo during inotropic stimulation Eur. J. Cardiothorac. Surg., April 1, 2001; 19(4): 431 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Borghetti, M. Campana, C. Scotti, D. Domenighini, P. Totaro, G. Coletti, M. Pagani, and R. Lorusso Biological versus prosthetic ring in mitral-valve repair: enhancement of mitral annulus dynamics and left-ventricular function with pericardial annuloplasty at long term Eur. J. Cardiothorac. Surg., April 1, 2000; 17(4): 431 - 439. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov, D. M. Cosgrove III, T. Shiota, J. Qin, H. Tsujino, W. J. Stewart, J. D. Thomas, M. Porqueddu, J. A. White, and E. H. Blackstone Cosgrove-Edwards Annuloplasty System: midterm results Ann. Thorac. Surg., March 1, 2000; 69(3): 717 - 721. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Scrofani Cinefluoroscopic assessment of human mitral anulus after mitral valvuloplasty J. Thorac. Cardiovasc. Surg., December 1, 1999; 118(6): 1154 - 1155. [Full Text] |
||||
![]() |
D. C. Miller Cinefluoroscopic assessment of human mitral anulus after mitral valvuloplasty J. Thorac. Cardiovasc. Surg., December 1, 1999; 118(6): 1155 - 1156. [Full Text] |
||||
![]() |
P. Dagum, G. R. Green, J. R. Glasson, G. T. Daughters, A. F. Bolger, L. E. Foppiano, N. B. Ingels Jr, and D. C. Miller POTENTIAL MECHANISM OF LEFT VENTRICULAR OUTFLOW TRACT OBSTRUCTION AFTER MITRAL RING ANNULOPLASTY J. Thorac. Cardiovasc. Surg., March 1, 1999; 117(3): 472 - 480. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||