Implantation of mechanical assist devices has saved the lives of patients awaiting heart transplantation.
1 The HeartMate left ventricular assist device (Thermo Cardiosystems, Inc, Woburn, Mass) is made to connect the inflow cannula to the apex of the left ventricle.
2 We describe a successful case in which the device was placed back-to-front with the inflow cannula inserted into the diaphragmatic wall of the right ventricle and the outflow graft through the left side of the diaphragm and in the left pleura to the ascending aorta (Fig 1).
|
A 15-year-old boy who was born with transposition of the great arteries underwent several balloon septostomies at birth because of desaturation. Despite this treatment, the desaturation was persistent. He therefore underwent an operation on an emergency basis at 3 months of age with an atrial switch as described by Mustard (pericardial baffle). At 4 years of age he required a reoperation because of residual atrial shunt and stenosis of the superior vena cava. Ten days after this operation a VVI pacemaker system was implanted because of atrioventricular dissociation. He then had an active life for almost 9 years, after which he gradually noted progressive fatigue and dyspnea. He had a cough, nausea, and cyanosis. Cardiac assessment revealed severe biventricular failure, gross atrioventricular valve regurgitation, and pulmonary hypertension (although reversible), and he was listed for heart transplantation. Despite "optimal" inotropic support, his condition continued to deteriorate with signs of multiorgan failure.
Because no donor was available, we decided to implant a HeartMate left ventricular assist system, even though we expected major technical difficulties since the right ventricle, where the inflow cannula had to be placed, was the systemic ventricle.
Surgical procedure.
A median sternotomy was performed, and the heart, which was heavily adherent, was carefully dissected. Because the patient was thin, we decided to place the pump intraperitoneally.
3 The aorta was cannulated. The systemic atrium was cannulated, and a left-sided bypass was instituted. This made the subsequent cannulation of the cavae easier. The aorta was crossclamped and blood cardioplegic solution was administered antegradely. The optimal positioning of the HeartMate vented electric device was back-to-front, with the inflow directed toward the diaphragmatic wall of the right ventricle and the outflow toward the apex region. Two incisions were made in the diaphragm corresponding to the inflow and outflow tracts. To avoid damage to the liver, from the front of the pump, we made a "pillow" from polytetrafluoroethylene (Preclude pericardial membrane
) stuffed with Dacron felt. When the device was in place, the inflow was on the right side and the outflow on the left side (Fig 2). An expected pulmonary hypertension was prevented with inhaled aerosolized epoprostenol (prostacyclin).
4
|
Comment.
The need for assist devices in patients with complex congenital heart conditions, as after atrial correction for transposition of the great arteries, could be a surgical technical challenge. Because the right ventricle is the systemic ventricle in transposition, several points must be considered. To place the device on the right side in the abdomen could be troublesome, as it might compress the liver and create unfavorable angles for the inflow and outflow cannulas, which could result in impaired flows. The trabeculae in the right ventricle could disturb the flow to the inflow cannula, leading to inadequate function of the device. A way to minimize this is to place the cannula on the diaphragmatic wall toward the septum. That worked well in this patient. However, placing the device back-to-front intraperitoneally raised some doubts, although problems did not appear. The device functioned very well. This patient had no major diet problems and no reflux from the stomach. The outflow graft was used in its full length, and as it made its way through the left pleura a major concern was the possibility of kinking. The patient had transient pain from the left hemithorax, probably due to movements of the graft. As yet there have been no signs of major infections.
In conclusion, we were successful in implanting a HeartMate vented electric left ventricular assist system back-to-front in a patient with severe heart failure after a failing Mustard procedure.
Footnotes
*W. L. Gore & Associates, Inc, Flagstaff, Ariz. ![]()
References
This article has been cited by other articles:
![]() |
R. S. George, E. J. Birks, R. C. Radley-Smith, A. Khaghani, and M. Yacoub Bridge to Transplantation With a Left Ventricular Assist Device for Systemic Ventricular Failure After Mustard Procedure Ann. Thorac. Surg., January 1, 2007; 83(1): 306 - 308. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |