J Thorac Cardiovasc Surg 2000;119:77-082
© 2000 Mosby, Inc.
SURGERY FOR ACQUIRED CARDIOVASCULAR DISEASE |
From the Division of Cardiothoracic Surgery,a The Milton S. Hershey Medical Center, Hershey, Pa; Cardiothoracic Surgeon, Sarasota Memorial Hospital,b Sarasota, Fla; and Department of Anesthesiac and Division of Cardiology,d The Milton S. Hershey Medical Center, Hershey, Pa.
Address for reprints: Ralph J. Damiano, Jr, MD, Chief, Division of Cardiothoracic Surgery, MC H165, The Milton S. Hershey Medical Center, PO Box 850, 500 University Dr, Hershey, PA 17033 (E-mail: rdamiano{at}psghs.edu).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The Zeus Robotic Surgical System (Computer Motion, Inc, Goleta, Calif) consists of 3 major components: a surgeon interface device consisting of a video monitor and 2 instrument handles, a computer controller, and 2 robotic arms(Fig 1). The surgeon manipulates the instrument handles, which are identical to traditional microsurgical instruments. The computer controller digitizes the surgeons motions and relays the information in real time to the 2 robotic arms. These arms are attached to the operating room table and precisely control the instrument tips that are within the surgical field. A third voice-controlled robotic arm is used to hold and move the endoscope.
|
| Methods |
|---|
|
|
|---|
Five enrolled patients were excluded from participation. The reasons for exclusion from the study were ejection fraction less than 40% (1 patient), the presence of a diffusely diseased or intramyocardial LAD (2 patients), myocardial infarction within 7 days of the operation (1 patient), and intraoperative hemodynamic instability before CPB (1 patient).
Operating room setup and monitoring.
The patients were placed supine on the operating room table, and arterial and venous access was obtained. After induction of general anesthesia, the 3 robotic arms of the Zeus system were attached to the operating table(Fig 1
, B). The patient and the robotic arms were then prepared and draped in a sterile fashion.
LITA harvest and CPB.
For the first 9 patients, a standard median sternotomy was performed and the LITA was taken down in a traditional manner. The pericardium was opened and the heart was exposed. The patient was heparinized and the activated clotting time was maintained above 450 seconds. The ascending aorta and right atrial appendage were cannulated and CPB was established. The patient was cooled to 32°C. A cannula was placed in the aortic root for antegrade administration of cardioplegic solution and a coronary sinus catheter was placed for retrograde cardioplegia administration.
In the tenth patient, the LITA was harvested thoracoscopically through 3 endoscopic ports. The camera port was placed in the 4th intercostal space at the anterior axillary line and the 2 instrument ports were in the 2nd and 6th intercostal spaces in the mid axillary line. A small (4 cm) anterior thoracotomy was made in the 3rd intercostal space. This patient was placed on CPB with the Heartport system (Heartport, Inc, Redwood City, Calif).
12
Setup of robotic system.
The Zeus Robotic Microsurgical System consisted of 2 robotic arms that were used to manipulate modified endoscopic instruments (Karl Storz, Culver City, Calif). The arms were attached to the operating table. The surgeon controlled the instruments by manipulating specially designed handles at a console(Fig 2). The handles allowed for 4 full ranges of motion (pan, roll, tilt, and in/out), as well as grasping. The surgeons motions were directly and precisely translated from the console to the robotic arms by a computer controller. Custom-designed software allowed for tremor elimination, as well as motion scaling over a range of 2:1 to 10:1.
|
|
In the tenth patient, the Heartport Endoclamp device was used and cardioplegic solution was administered antegradely into the aortic root. Subsequently, cardioplegic solution was administered in a retrograde fashion through a percutaneously placed coronary sinus cannula.
Robotically assisted coronary anastomosis.
The heart was left in situ and was not suspended in a pericardial cradle. The chest retractor was relaxed to allow a more normal anatomic position of the chest wall. After cardioplegic arrest, an arteriotomy was made in the distal LAD. In the last 5 patients, this was done with specialized scissors and the Zeus system. A continuous end-to-side anastomosis was performed endoscopically with the robotic instruments. The heel was sewn first and parachuted down, followed by completion of the toe of the anastomosis. No manipulation of the heart was necessary to perform the anastomosis, and visualization of each stitch through the subxiphoid camera port was excellent. An assistant held the LITA pedicle through the chest incision with standard instruments. The running anastomosis was performed with a specially designed 7-cm double-armed 7-0 polytetrafluoroethylene suture.*
Motion scaling (2.5:1) was used for each anastomosis. After completion of the anastomosis, the suture was tied endoscopically. No repair stitches were required. The clamp on the LITA and the aortic crossclamp were removed, and the patient was rewarmed. The patient was then weaned from CPB. Protamine sulfate was administered to reverse the heparin, and the chest was closed in a standard fashion. No patient required inotropic support.
Short-term graft flow.
Blood flow through the LITA graft was measured with a 2.0- to 2.5-mm ultrasonic flow probe and flowmeter (HT311, Transonic Systems, Inc, Ithaca, NY) and was recorded with FlowTrace 32 software (Transonic Systems, Inc). If flow was not judged to be adequate, selective coronary angiography was performed in the operating suite. The LITA graft was selectively catheterized via the left radial artery. An OEC model 9600 digital imaging system was used to visualize the anastomosis (Omega Electronics Corporation, Salt Lake City, Utah).
Patient follow-up.
During the postoperative period, the following data were collected: hemodynamic parameters, chest tube drainage, need for inotropic medications, length of intensive care unit stay, and length of hospital stay. Eight weeks after the operation, all patients underwent a second coronary angiogram of the LITA-LAD anastomosis to assess graft patency. Follow-up was 100%.
| Results |
|---|
|
|
|---|
Intraoperative results.
The robotic system was quick and easy to assemble. The average time required for setup was 14.9 ± 2.1 minutes. No intraoperative complications related to port placement were encountered, and there were no mechanical failures of the robotic system.
The total number of grafts performed was 2.3 ± 0.3 per patient with a mean crossclamp time of 51.1 ± 1.6 minutes. The time required to perform the endoscopic LITA-LAD anastomosis with the robotic system was 23.6 ± 1.4 minutes (range 18-30 minutes). All anastomoses were successfully performed and no repair stitches were required.
The patient who had a limited thoracotomy and CPB with a Heartport device had a crossclamp time of 71 minutes and an anastomosis time of 28 minutes.
All patients underwent intraoperative ultrasonic determination of flow through the LITA graft. Eighty percent (8/10) of grafts measured were patent and had excellent diastolic flow. Average graft flow was 31 ± 7 mL/min. Two of the 10 grafts had inadequate flow and were taken down and reconstructed manually. In 1 case, the patient who had a Heartport procedure, an intraoperative coronary angiogram was performed. The graft was not visualized. However, the distal anastomosis was probe patent. A tentative diagnosis of graft spasm was made. In the second case, angiography was unavailable. Subsequent review of the videotape revealed a technical error. This case represented a failure of surgical technique, not of the robotic system. Conversion to the manual technique was easily accomplished in both patients, and all patients left the operating room with a patent LITA-LAD anastomosis.
Postoperative course.
There was only one postoperative complication. This patient required return to the operating room the evening of surgery for excessive mediastinal hemorrhage. The source of hemorrhage was identified as the lateral trocar insertion site. The bleeding was stopped successfully with electrocautery.
No other complications were encountered in the postoperative period. The average length of stay in the intensive care unit was 1.1 ± 1 days. The average length of stay in the hospital was 4.1 ± 0.2 days.
Late follow-up.
A 100% late follow-up was achieved. The mean period of follow-up was 142.1 ± 10.7 days. No late complications occurred. All patients are presently in New York Heart Association class I.
Eight weeks after the operation, graft patency was assessed by a second coronary angiogram. These studies were all performed at our institution and revealed that 100% (8/8) of the LITA-LAD grafts were patent(Fig 4). One graft had a 50% stenosis at its insertion site.
|
| Discussion |
|---|
|
|
|---|
Recently, robotically assisted microsurgery systems have been introduced to increase the precision of endoscopic surgery. These computer-guided systems can control both surgical instruments
8,9 and endoscopic cameras.
17 These systems may help to overcome the obstacles to endoscopic CABG. Specifically, robotic instrumentation may provide access to multiple target vessels and offer the surgeon the ability to operate precisely in confined spaces. The most significant limitation to the use of traditional endoscopic instruments in CABG is the lack of precision that results from the additive effects of instrument length and operator tremor. Robotic microsurgical systems address this limitation by filtering high-frequency motions, thus eliminating tremor. Dexterity is further enhanced through computer motion-scaling. This allows the surgeon to make large, easy-to-perform macroscopic movements at the console and have these scaled down by the computer to microscopic movements of the instrument tip inside the patient. Endoscopic magnification of the operative site also improves the accuracy of suture placement by providing enhanced visualization. The magnification is 3 to 4 times higher than that with traditional surgical loupes.
Finally, the Zeus system incorporates a robotic voice-controlled arm (Aesop) to control the endoscope. This eliminates the need for an assistant and has been shown to improve the stability of the image, reduce camera clearings, and decrease the operative time during surgical procedures.
17,18 It has been our early experience that this technology greatly facilitates microsurgery. Under high magnification, a stable image is absolutely critical. Moreover, the ability to control the movement of the Aesop robotic arm by voice effectively gives the surgeon a third arm.
In previous animal studies, endoscopic CABG has been shown to be feasible with acceptable short-term and long-term graft patency.
11,13 The present study analyzed the short-term outcome of robotically assisted endoscopic CABG in human subjects. Short-term graft patency was 80% (8/10). Conversion to manual anastomosis was easily completed in the 2 patients with inadequate flow, and all patients left the operating room with patent LITA-LAD grafts. Graft patency 2 months after the operation was 100% (8/8) as demonstrated by angiography. These results demonstrate the feasibility of endoscopic CABG. More important, no device-related complications were observed. Anastomoses can be performed with robotic assistance with acceptable short-term results. The operating room time required for equipment setup was minimal.
The principal shortcoming of this study was that it does not represent a completely endoscopic approach. Other CABGs were performed through traditional incisions. Clearly, a completely endoscopic procedure will be a greater challenge, but one that appears to be achievable.
The other parts of endoscopic CABG have been performed without a surgical incision. Heartport technology allows for CPB without the need for a thoracic incision.
19 Endoscopic LITA harvest is a widely described procedure
20,21 and can be easily performed manually. The use of an expensive robotic system does not seem to be warranted for this part of the procedure. Therefore we did not perform the LITA harvest with the robotic system. We did perform a manual endoscopic harvest of the LITA in 1 patient.
On the other hand, prior attempts to perform manual endoscopic coronary anastomoses have been fraught with difficulty and have not been clinically feasible. Robotic assistance does represent an enabling technology for this critical part of the procedure. This report indicates that the most difficult part of the procedure, the coronary anastomosis, can be performed endoscopically with robotic assistance. Early experience in Europe by Mohrs group
22 at the University of Leipzig and Reichenspurners group
23 in Munich with endoscopic CABG also has been positive.
These early results with robotic systems suggest that a completely endoscopic approach to CABG is feasible in carefully selected patients. It is our belief that patient morbidity will be significantly reduced with this approach. However, further data are needed to support this hypothesis. It should be possible in human beings to perform multiple CABGs with an endoscopic approach and peripheral cannulation. Moreover, preliminary work in animal models and in the clinical setting suggest that beating heart approaches are also feasible.
24
Our initial clinical experience indicates that robotic assistance is an enabling technology that allows for the performance of endoscopic CABG. Only time will tell whether the extra costs of this system will be justified by its potential advantages. Further clinical trials are warranted to explore the potential of this exciting new technology.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
V. Falk and F. W. Mohr Minimally Invasive Myocardial Revascularization Card. Surg. Adult, January 1, 2008; 3(2008): 697 - 710. [Full Text] |
||||
![]() |
R. J. Damiano Jr Robotics in cardiac surgery: The emperor's new clothes J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 559 - 561. [Full Text] [PDF] |
||||
![]() |
D. de Canniere The seven stages of an idea J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 562 - 564. [Full Text] [PDF] |
||||
![]() |
S. Mierdl, C. Byhahn, V. Lischke, T. Aybek, G. Wimmer-Greinecker, S. Dogan, S. Viehmeyer, P. Kessler, and K. Westphal Segmental Myocardial Wall Motion During Minimally Invasive Coronary Artery Bypass Grafting Using Open and Endoscopic Surgical Techniques Anesth. Analg., February 1, 2005; 100(2): 306 - 314. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.W. R. Bolton and J. E. Connally Results of a phase one study on robotically assisted myocardial revascularization on the beating heart Ann. Thorac. Surg., July 1, 2004; 78(1): 154 - 158. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Jacobs, D. Holzhey, B. B. Kiaii, J. F. Onnasch, T. Walther, F. W. Mohr, and V. Falk Limitations for manual and telemanipulator-assisted motion tracking--implications for endoscopic beating-heart surgery Ann. Thorac. Surg., December 1, 2003; 76(6): 2029 - 2035. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Morgan, M. E. Ginsburg, J. R. Sonett, D. L.S. Morales, T. Kohmoto, L. A. Gorenstein, C. R. Smith, and M. Argenziano Advanced thoracoscopic procedures are facilitated by computer-aided robotic technology Eur. J. Cardiothorac. Surg., June 1, 2003; 23(6): 883 - 887. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Damiano Jr Endoscopic coronary artery bypass grafting--the first steps on a long journey J. Thorac. Cardiovasc. Surg., March 1, 2003; 125(90030): S64 - 65. [Full Text] [PDF] |
||||
![]() |
S. Mierdl, C. Byhahn, S. Dogan, T. Aybek, G. Wimmer-Greinecker, P. Kessler, D. Meininger, and K. Westphal Segmental Wall Motion Abnormalities During Telerobotic Totally Endoscopic Coronary Artery Bypass Grafting Anesth. Analg., April 1, 2002; 94(4): 774 - 780. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. W Susilo and A. P Schulz Totally Robotic Technique in Multivessel Coronary Disease -- Is it Possible? Asian Cardiovasc Thorac Ann, March 1, 2002; 10(1): 92 - 94. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Murtra The adventure of cardiac surgery Eur. J. Cardiothorac. Surg., February 1, 2002; 21(2): 167 - 180. [Full Text] [PDF] |
||||
![]() |
U. Kappert, R. Cichon, J. Schneider, V. Gulielmos, T. Ahmadzade, J. Nicolai, S.-M. Tugtekin, and S. Schueler Technique of closed chest coronary artery surgery on the beating heart Eur. J. Cardiothorac. Surg., October 1, 2001; 20(4): 765 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Damiano Jr, H. A. Tabaie, M. J. Mack, J. R. Edgerton, C. Mullangi, W. P. Graper, and S. M. Prasad Initial prospective multicenter clinical trial of robotically-assisted coronary artery bypass grafting Ann. Thorac. Surg., October 1, 2001; 72(4): 1263 - 1269. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Lomanto, W.-K. Cheah, J. B. So, and P. M. Goh Robotically Assisted Laparoscopic Cholecystectomy: A Pilot Study Arch Surg, October 1, 2001; 136(10): 1106 - 1108. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Kappert, J. Schneider, R. Cichon, V. Gulielmos, S.-M. Tugtekin, J. Nicolai, K. Matschke, and S. Schueler Development of Robotic Enhanced Endoscopic Surgery for the Treatment of Coronary Artery Disease Circulation, September 18, 2001; 104(90001): I-102 - 107. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Dogan, T. Aybek, K. Westphal, S. Mierdl, A. Moritz, and G. Wimmer-Greinecker Computer-enhanced totally endoscopic sequential arterial coronary artery bypass Ann. Thorac. Surg., August 1, 2001; 72(2): 610 - 611. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. W. Mohr, V. Falk, A. Diegeler, T. Walther, J. F. Gummert, J. Bucerius, S. Jacobs, and R. Autschbach Computer-enhanced ""robotic"" cardiac surgery: Experience in 148 patients J. Thorac. Cardiovasc. Surg., May 1, 2001; 121(5): 842 - 853. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Damiano Jr Editorial: Endoscopic coronary artery bypass grafting--The first steps on a long journey J. Thorac. Cardiovasc. Surg., October 1, 2000; 120(4): 806 - 807. [Full Text] [PDF] |
||||
![]() |
W. D. Boyd, R. Rayman, N. D. Desai, A. H. Menkis, W. Dobkowski, S. Ganapathy, B. Kiaii, G. Jablonsky, F. N. McKenzie, and R. J. Novick Closed-chest coronary artery bypass grafting on the beating heart with the use of a computer-enhanced surgical robotic system J. Thorac. Cardiovasc. Surg., October 1, 2000; 120(4): 807 - 809. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||