J Thorac Cardiovasc Surg 2001;121:1090-1100
© 2001 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
From the Department of Cardiothoracic Surgery, University Hospital of Wales, Cardiff, Wales,a and the Medical Data Research Centre, Providence Health System, Portland, Ore.b
Funding for this study was provided by Medtronic, Inc, Minneapolis, Minn.
Received for publication Sept 5, 2000. Revisions requested Oct 24, 2000; revisions received Dec 18, 2000. Accepted for publication Dec 20, 2000. Address for reprints: Eric G. Butchart, FESC, FRCS, Department of Cardiothoracic Surgery, University Hospital of Wales, Heath Park, Cardiff CF14 4XN, United Kingdom.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Prosthetic valve assessment has been hindered by failure to take patient-related and management-related factors into account. In addition, cardiac surgeons' views have been colored by certain preconceptions about valve design that have been incorporated by valve manufacturers into their marketing strategy. There is a tendency to categorize all types of mechanical valves into 3 design categories (caged ball, tilting disc, and bileaflet) and to make all-embracing generalizations about the characteristics, sophistication, and modernity of each (ie, first generation, second generation). Worse still, and potentially more dangerous, recommendations on anticoagulation management are often based on broad design categories rather than individual prostheses.
4,5
Because some of the early tilting disc valve designs were susceptible to valve thrombosis, many assume that this problem afflicts all tilting disc designs, whereas analysis of available data shows that this is clearly not the case.
1 Similarly, bileaflet designs are perceived by many as more "modern," "third generation," and less susceptible to the problems of thrombosis and thromboembolism; however, some types of bileaflet valve have been shown to be more prone to valve thrombosis than tilting disc valves, especially in developing countries,
6 and one recently introduced bileaflet valve had to be withdrawn from clinical use before trials were completed because of its high susceptibility to valve thrombosis.
7
The Medtronic Hall tilting disc valve (Medtronic, Inc, Minneapolis, Minn) was introduced to clinical practice in 1977, the same year as the St Jude Medical bileaflet valve (St Jude Medical, Inc, St Paul, Minn). Although both are widely regarded as safe, reliable devices, it seems illogical that the latter is regarded by some as more "modern." A great deal of research and development went into the refinement of the tilting disc concept that eventually resulted in the production of the Medtronic Hall valve.
8 Unique design features were incorporated to introduce very low susceptibility to thrombosis. In comparison with previous tilting disc designs, the relative size of the minor orifice was increased and the disc was made to lift out of the housing and rotate on opening, all features designed to improve washing of vulnerable points and eliminate areas of low velocity flow. In addition, housing cross members were placed in the center of blood flow and shaped to avoid causing flow disturbance.
Dissatisfaction with earlier tilting disc valve designs led to the introduction of the Medtronic Hall valve in Cardiff in 1979, and its use continues today.
| Methods |
|---|
|
|
|---|
|
|
In addition to warfarin, some patients received dipyridamole at the discretion of their own physicians, but there was no prescribing consistency. Many patients discontinued the drug because of side effects and those who took it did so for only part of their follow-up. Overall, 13% of AVR patients, 40% of MVR patients, and 32% of DVR patients took dipyridamole for part of their follow-up. The percentage of follow-up days taking dipyridamole was 7% for AVR, 28% for MVR, and 23% for DVR.
Follow-up
In 1979, the Medtronic Hall valve was still a relatively new prosthesis, and the first author (E.G.B.) implanted the first Medtronic Hall valve in the United Kingdom. This series was therefore planned as a prospective observational study from the outset with follow-up data recorded and entered into a computer at every outpatient visit for each patient. Most patients attended follow-up clinics at the University Hospital at intervals of 6 to 12 months. In the case of patients attending follow-up clinics at other hospitals, data were regularly retrieved from these sources also. At the time of this analysis, the small number of patients who had ceased to attend clinics regularly were contacted by telephone to update their follow-up.
Patients were also "flagged" in the National Health Service (NHS) Central Register, which maintains a death register for the United Kingdom. The NHS Register then notified the study investigators of all deaths of patients in the study and provided a copy of the death certificate. The hospital records of all patients who died in any hospital were scrutinized for further details. A detailed analysis of deaths earlier in this series has already been published,
11,12 and the method of using NHS Register notification has previously been validated.
13 Autopsies were performed in 48% of all deaths and 70% of sudden or unwitnessed deaths.
11
Definitions and statistical analysis
For definitions of mortality and morbidity, standard guidelines were used.
14 Early events refers to events within the first 30 postoperative days. Late events refers to occurrences at all subsequent times.
Early events (<30 days) were expressed as percentages and late events (>30 days) were expressed as percent per year. All events, including those in the first 30 days, were included in the event-free curves for mortality and morbidity. Right censored data, including patient survival, valve-related reoperation, obstructive valve thrombosis, prosthetic endocarditis leading to reoperation or death, and paravalvular leak were constructed by means of the actuarial method of Kaplan and Meier.
15 Interval-censored data such as thromboembolism and bleeding complications that were known to have occurred during a particular interval of follow-up were assessed by means of the Turnbull generalization of the Kaplan-Meier method and by the conventional Kaplan-Meier method taking the midpoint of follow-up intervals as the time of failure. Because the 2 methods yielded very similar results, probably because most intervals are small,
16 we used the latter method. The significance of differences between groups was tested by means of the log-rank method. The Weibull distribution model showed good agreement with the Kaplan-Meier method to estimate the freedom from thromboembolism. Instantaneous hazard functions of thromboembolism were derived from the Weibull fits by taking the derivative of the cumulative hazard function. The Cox proportional hazard model was used to assess the relationship between various risk factors and patient survival.
17 All analyses were performed using the SPLUS software package, version 2000 (MathSoft, Inc, Seattle, Wash).
| Results |
|---|
|
|
|---|
|
|
|
|
years (data from English Life Tables, published by the United Kingdom Office for National Statistics).
|
|
|
|
|
|
Thrombosis
Four patients had obstructive valve thrombosis, 2 with AVR and 2 with MVR. Presentation was insidious over several weeks in each case. Two patients (1 AVR, 1 MVR) were treated by thrombectomy and 2 by re-replacement with another Medtronic Hall valve. All 4 were postmenopausal women in their 70s. Three patients survived reoperation. The patient who died had severely impaired left ventricular function caused by coronary disease and subendocardial fibrosis. Valve thrombosis may have been a terminal event associated with low cardiac output in this case. The 3 remaining patients had all had anticoagulation interruption: One diabetic patient had not received anticoagulants for 4 months after an intracerebral hemorrhage; one grossly obese patient with hypercholesterolemia had her warfarin stopped 3 times in the course of a month in preparation for a minor orthopedic operation that was canceled twice; and an asthmatic smoker receiving steroids had been given vitamin K in addition to having the anticoagulation regimen interrupted. All these patients also had pulmonary infections in the period preceding the onset of valve thrombosis, raising the possibility that valve thrombosis was due to a combination of a pre-existing prothrombotic state, the prothrombotic effects of pulmonary infection, and anticoagulation interruption. The detailed case histories of all 4 patients and their risk factors for valve thrombosis have already been described.
18 No valve thromboses occurred among the DVR patients. Among the 70% of patients with sudden, rapidly progressing, or unwitnessed death submitted to autopsy, there were no cases of valve thrombosis.
11 The linearized rates of valve thrombosis were 0.04%/y for AVR and 0.03%/y for MVR.
Thromboembolism
A total of 443 thromboembolic events occurred in this series, 340 of which were minor, almost all transient or reversible ischemic attacks (transient ischemic attacks or reversible ischemic neurologic deficits); 103 were major (almost all strokes) that led to permanent damage or death. The linearized rates of minor events were 1.7%/y for AVR, 3.2%/y for MVR, and 2.8%/y for DVR. The linearized rates of stroke were 0.6%/y for AVR, 0.8%/y for MVR, and 0.6%/y for DVR. The linearized rates of overall thromboembolism were 2.3%/y for AVR, 4.0%/y for MVR, and 3.4%/y for DVR(Table VII). Overall, 10% of patients had 1 thromboembolic event and 6% had more than 1 event. Only 4% of AVR patients and 5% of MVR patients had a stroke at any time. In patients who had multiple events at 1 follow-up interval, the linearized rate was computed in such a way that applied the worst severity of the event to all events. The overall actuarial estimates of freedom from thromboembolism at 10 and 15 years were 84% and 79% for AVR, 77% and 65% for MVR, and 76% and 72% for DVR(Fig 7). Thromboembolism risk was highest during the first 30 days: computed as a "1-month" linearized rate, it was 8%/y and 33%/y for AVR and MVR, respectively. In contrast, the linearized rate for all thromboembolism was 2%/y for AVR and 4%/y for MVR after the first month, declining further during long-term follow-up(Fig 8). Significant risk factors for thromboembolism were MVR, diabetes, hypertension, and a history of previous embolism(Table VIII).
|
|
|
|
Prosthetic valve endocarditis
Prosthetic valve endocarditis developed in 59 patients, 33 of whom died (1 at reoperation). The remainder had successful reoperations. The linearized rates of endocarditis were 0.4%/y for AVR, 0.4%/y for MVR, and 0.7%/y for DVR. The actuarial estimates of freedom from endocarditis at 10 and 15 years were 96% and 95% for AVR, 95% and 94% for MVR, and 94% and 94% for DVR.
Paravalvular leaks
Paravalvular leak in the absence of endocarditis occurred in 13 patients: 2 AVR, 9 MVR, and 3 DVR, the latter all in the mitral position. The linearized rates for AVR, MVR, and DVR were 0.04%/y, 0.1%/y, and 0.2%/y, respectively. The thin sewing ring of the size 20 and 22 aortic prostheses was not associated with any paravalvular leaks. The actuarial freedom from reoperation for paravalvular leak at 15 years was 99.7% for AVR, 99% for MVR, and 98% for DVR.
Reoperation and explantation
Any operation that repaired, altered, or replaced a previously inserted valve was called a reoperation. There were 58 reoperations (22 AVR, 25 MVR, 11 DVR), 48 of which (19 AVR, 21 MVR, 8 DVR) resulted in explantation. The linearized rates of reoperation were 0.4%/y for AVR, 0.5%/y for MVR, and 0.7%/y for DVR. The linearized rates of explantation were 0.4%/y for AVR, 0.4%/y for MVR, and 0.5%/y for DVR. The actuarial estimates of freedom from reoperation at 15 years were 96% for AVR, 95% for MVR, and 93% for DVR. The actuarial estimates of freedom from explantation at 15 years were 96% for AVR, 96.5% for MVR, and 94% for DVR. The reasons for reoperation are listed inTable IX.
|
| Discussion |
|---|
|
|
|---|
The goal of the Medtronic Hall design was to improve on previous tilting disc designs with the aim of improving durability and hemodynamic performance and reducing thrombogenicity. The results of this 20-year experience show that these aims have been achieved. There were no mechanical failures. Hemodynamic measurements were not performed on all patients, but echocardiographic assessment of mitral prostheses was carried out early in the study, with peak rate of increase of left ventricular dimension during diastole being used for the assessment. These evaluations showed that the hemodynamic performance of the Medtronic Hall valve was significantly better than that of Björk-Shiley (Shiley, Inc, Irvine, Calif) and Starr-Edwards prostheses (Baxter Healthcare Corp, Edwards Division, Santa Ana, Calif) of the same size, and left ventricular filling in 29- or 31-mm prostheses was not significantly different from that of subjects with normal mitral valves. In patients with 27-mm and 29- to 31-mm prostheses, 91% and 89%, respectively, improved by one class or more.
23 In the current analysis, the NYHA performance status of most patients improved by at least one class, an improvement that was maintained at 10 years. In patients whose dominant lesion was aortic stenosis and who did not have coronary disease, arguably the "purest" test of prosthesis performance, survival was almost identical to that of the age- and sex-matched general population for the first 7
years and not significantly different from the general population out to 10 years(Fig 5
). This provides a combined indirect assessment of hemodynamic effectiveness and freedom from any excess of life-threatening events in comparison with the general population. Both characteristics of valve performance are probably attributable to optimum blood flow through the prosthesis.
Optimum flow characteristics not only provide the best relief of transvalvular gradients, allowing left ventricular mass regression after AVR, for example, but also, by providing good "washing" of all components and minimizing flow disturbance, reduce the risk of thrombus deposition and platelet activation.
21 The Medtronic Hall valve has been shown to have an equivalent hemodynamic performance to the St Jude Medical valve in the larger sizes and to have superior hemodynamic performance to it in the smaller aortic sizes,
24 particularly with increasing cardiac output.
25 It is also possible to orientate the Medtronic Hall aortic prosthesis so that the larger orifice faces the greater curvature of the ascending aorta, thereby creating near-physiologic flow.
26,27 The same can never be achieved with a bileaflet valve, because it has 3 orifices
26; consequently, valve gradients across the Medtronic Hall valve in its optimum orientation are lower than can be achieved with the St Jude Medical valve in any orientation.
28 This improved and less turbulent flow in comparison with the bileaflet design may contribute to the trend toward greater left ventricular mass regression that has been reported with the Medtronic Hall valve in comparison with the St Jude Medical valve.
29 It may also partly explain the finding of less hemolysis with the Medtronic Hall valve when normally functioning Medtronic Hall and St Jude Medical valves are compared.
30 Anderson and associates
31 have shown normalization of left ventricular mass index after AVR with the Medtronic Hall valve, using ultrafast computed tomographic scan, even in the smallest sizes 20 and 21. Furthermore, a recent study has shown more rapid left ventricular mass regression with the Medtronic Hall valve in comparison with the St Jude Medical valve after AVR.
32
In the series reported here, the aortic valve was implanted preferentially with the larger orifice toward the greater curvature of the aorta, rotating the valve only if this position caused the downward deflection of the disc to impinge on the septum. The majority of aortic prostheses (89.1%) were thus in the optimum orientation. The good long-term survival after AVR in comparison with the general population in this series may relate to the very low turbulence and low gradients produced by the prosthesis in this orientation. In an earlier analysis, we have shown that the gradually increasing incidence of stroke with advancing age in this series did not differ significantly from the age-related background incidence of stroke in the general population.
33 This finding may also reflect the beneficial effect of optimum flow in combination with low thrombogenicity. Transcranial Doppler detection of high-intensity transient signals has recently shown significantly fewer high-intensity transient signals with the Medtronic Hall valve in its optimum orientation both in comparison with its worst orientation and in comparison with the St Jude Medical valve in any orientation, in keeping with this hypothesis.
34 Using data from earlier analysis of this series, we have shown that cerebrovascular events after AVR were closely related to well known stroke risk factors, that is, hypertension, cigarette smoking, diabetes, and arterial disease. Patients who were in sinus rhythm, normotensive, nonsmokers, not diabetic, and who had no coronary artery disease had no cerebrovascular events during a 13-year follow-up period.
22
The mitral prosthesis in this series was implanted with the larger orifice posteriorly in almost all patients. This orientation has been shown to permit flow patterns within the left ventricle similar to those seen physiologically
35 and probably contributed to the almost normal rate of ventricular filling seen with the larger sized mitral prostheses referred to above.
23
Thromboembolic events after mitral valve replacement are determined largely by conditions that promote relative stagnation in the left atrium (atrial fibrillation, increased left atrial size, residual mitral gradient, and impaired left ventricular function) and the extent to which these adverse conditions are controlled by effective anticoagulation.
21 In earlier analyses, we have shown that the optimum INR of 3.0 prevents most serious thromboembolic events in patients with the Medtronic Hall valve in the mitral position
10 and that the incidence of events has an association with known stroke risk factors and a seasonal fluctuation with winter predominance, in keeping with pulmonary infection as a trigger factor.
36
Much earlier in this series, we reported a zero rate of valve thrombosis with the Medtronic Hall valve. At that stage, in 1988, in just over 1000 patients, there had been no cases of valve thrombosis.
37 Since then 4 patients have had obstructive valve thrombosis, but the overall incidence remains extremely low (0.04%/y). A previously published analysis of these patients identified the combination of the prothrombotic effects of pulmonary infection and anticoagulation interruption as the probable interaction that triggered thrombus deposition on the prosthesis in 3 of 4 patients.
18
The valve thrombosis rate reported here is lower than that reported for other large series of mechanical valves and some series of bioprostheses with comparable lengths of follow-up.
3 Eleven of 12 St Jude Medical mitral valve series with more than 300 valve-years reported since 1989 had higher thrombosis rates. For these 11 series with a total of 28,000 years, the mean thrombosis rate was 0.13%/y (range 0.07%/y-0.45%/y).
3
The very good thromboembolic performance of the Medtronic Hall valve in this series has been achieved with low-intensity anticoagulation, giving the additional benefit of a low incidence of serious bleeding events. To place it in context, this needs to be compared with the background incidence of major hemorrhage in the general population. Data from the placebo arms of several anticoagulation trials show that major hemorrhage occurs at a linearized rate between 0.7%/y and 1.6%/y (mean 1.2%/y) in a population of approximately the same age group as this valve series.
38-40 The linearized rates for AVR, MVR, and DVR (1.2%/y, 1.4%/y, and 1.6%/y, respectively) fall within the range in the general population, indicating that the patients in this series were not exposed to an excess risk as the result of their anticoagulation management.
Overall, our results at 20 years demonstrate that the Medtronic Hall valve has excellent durability, very good hemodynamic performance, and exceptionally low thrombogenicity, leading to good long-term survival with low complication rates and a very low requirement for reoperation. Contrary to the perceptions of many surgeons about the supposed design advantages of bileaflet valves, our long-term results support the large number of recent reports showing that more physiologic flow conditions can be created by the Medtronic Hall valve. In our view this prosthesis offers superior hemodynamic performance and lower thrombogenicity than currently available bileaflet designs and should now be considered the gold standard against which other prostheses should be judged.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. W. Emery, A. M. Emery, A. Knutsen, and G. V. Raikar Aortic Valve Replacement with a Mechanical Cardiac Valve Prosthesis Card. Surg. Adult, January 1, 2008; 3(2008): 841 - 856. [Full Text] |
||||
![]() |
T. Gudbjartsson, T. Absi, and S. Aranki Mitral Valve Replacement Card. Surg. Adult, January 1, 2008; 3(2008): 1031 - 1068. [Full Text] |
||||
![]() |
J. L. Svennevig, M. Abdelnoor, and S. Nitter-Hauge Twenty-Five Year Experience With the Medtronic-Hall Valve Prosthesis in the Aortic Position: A Follow-Up Cohort Study of 816 Consecutive Patients Circulation, October 16, 2007; 116(16): 1795 - 1800. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lund and M. Bland Risk-corrected impact of mechanical versus bioprosthetic valves on long-term mortality after aortic valve replacement J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 20 - 26. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Bakhtiary, M. Schiemann, O. Dzemali, T. Wittlinger, M. Doss, H. Ackermann, A. Moritz, and P. Kleine Stentless bioprostheses improve postoperative coronary flow more than stented prostheses after valve replacement for aortic stenosis J. Thorac. Cardiovasc. Surg., April 1, 2006; 131(4): 883 - 888. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Garcia-Rinaldi Clopidogrel-Aspirin Inhibit the Thrombogenicity of St. Jude Mechanical Aortic Prostheses Ann. Thorac. Surg., February 1, 2006; 81(2): 787 - 787. [Full Text] [PDF] |
||||
![]() |
R. Tominaga, K. Kurisu, Y. Ochiai, Y. Tomita, M. Masuda, S. Morita, and H. Yasui A 10-Year Experience With the Carbomedics Cardiac Prosthesis Ann. Thorac. Surg., March 1, 2005; 79(3): 784 - 789. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ruel, R. G. Masters, F. D. Rubens, P. J. Bedard, A. L. Pipe, W. G. Goldstein, P. J. Hendry, and T. G. Mesana Late incidence and determinants of stroke after aortic and mitral valve replacement Ann. Thorac. Surg., July 1, 2004; 78(1): 77 - 83. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. B Kale, D. S Saksena, Y. C Agnihotri, and K. H Ayyer Thrombectomy With Disc Rotation of Medtronic Valves Asian Cardiovasc Thorac Ann, December 1, 2003; 11(4): 309 - 313. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Ikonomidis, J. M. Kratz, A. J. Crumbley III, M. R. Stroud, S. M. Bradley, R. M. Sade, and F. A. Crawford Jr Twenty-year experience with the St Jude Medical mechanical valve prosthesis J. Thorac. Cardiovasc. Surg., December 1, 2003; 126(6): 2022 - 2031. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola Choice of prosthetic heart valve for adult patients J. Am. Coll. Cardiol., March 19, 2003; 41(6): 893 - 904. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. L. Grunkemeier and Y. Wu "Our complication rates are lower than theirs": Statistical critique of heart valve comparisons J. Thorac. Cardiovasc. Surg., February 1, 2003; 125(2): 290 - 300. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Gudbjartsson, S. Aranki, and L. H. Cohn Mechanical/Bioprosthetic Mitral Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 951 - 986. [Full Text] |
||||
![]() |
P. Kleine, M. Scherer, U. Abdel-Rahman, A. A. Klesius, H. Ackermann, and A. Moritz Effect of mechanical aortic valve orientation on coronary artery flow: Comparison of tilting disc versus bileaflet prostheses in pigs J. Thorac. Cardiovasc. Surg., November 1, 2002; 124(5): 925 - 932. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Butchart, N. Payne, H.-H. Li, K. Buchan, K. Mandana, and G. L. Grunkemeier Better anticoagulation control improves survival after valve replacement J. Thorac. Cardiovasc. Surg., April 1, 2002; 123(4): 715 - 723. [Abstract] [Full Text] |
||||
| |||