J Thorac Cardiovasc Surg 2001;122:1186-1195
© 2001 The American Association for Thoracic Surgery
Evolving Technology (ET) |
From the Departments of Cardiovascular Research and Cardiac Transplantation, Texas Heart Institute,a Houston, Tex; the Division of Cardiothoracic Surgery, Columbia Presbyterian Medical Center,b New York, NY; Sharp Memorial Hospital,c San Diego, Calif; The Cleveland Clinic Foundation,d Cleveland, Ohio; and Thermo Cardiosystems, Inc,e Woburn, Mass.
Received for publication Dec 11, 2000. Revisions requested Feb 1, 2001; revisions received April 24, 2001. Accepted for publication June 18, 2001. Address for reprints: O. H. Frazier, MD, Texas Heart Institute, PO Box 20345, Houston, TX 77225-0345 (E-mail: mmallia@ heart.thi.tmc.edu).
Abstract
Background: Despite advances in heart transplantation and mechanical circulatory support, mortality among transplant candidates remains high. Better ways are needed to ensure the survival of transplant candidates both inside and outside the hospital.
Methods: In a prospective, multicenter clinical trial conducted at 24 centers in the United States, 280 transplant candidates (232 men, 48 women; median age, 55 years; range, 11-72 years) unresponsive to inotropic drugs, intra-aortic balloon counterpulsation, or both, were treated with the HeartMate Vented Electric Left Ventricular Assist System (VE LVAS). A cohort of 48 patients (40 men, 8 women; median age, 50 years; range, 21-67 years) not supported with an LVAS served as a historical control group. Outcomes were measured in terms of laboratory data (hemodynamic, hematologic, and biochemical), adverse events, New York Heart Association functional class, and survival.
Results: The VE LVAStreated and nonVE LVAStreated (control) groups were similar in terms of age, sex, and distribution of patients by diagnosis (ischemic cardiomyopathy, idiopathic cardiomyopathy, and subacute myocardial infarction). VE LVAS support lasted an average of 112 days (range, < 1-691 days), with 54 patients supported for > 180 days. Mean VE LVAS flow (expressed as pump index) throughout support was 2.8 L · min1 · m2. Median total bilirubin values decreased from 1.2 mg/dL at baseline to 0.7 mg/dL (P = .0001); median creatinine values decreased from 1.5 mg/dL at baseline to 1.1 mg/dL (P = .0001). VE LVASrelated adverse events included bleeding in 31 patients (11%), infection in 113 (40%), neurologic dysfunction in 14 (5%), and thromboembolic events in 17 (6%). A total of 160 (58%) patients were enrolled in a hospital release program. Twenty-nine percent of the VE LVAS-treated patients (82/280) died before receiving a transplant, compared with 67% of controls (32/48) (P < .001). Conversely, 71% of the VE LVAStreated patients (198/280) survived: 67% (188/280) ultimately received a heart transplant, and 4% (10/280) had the device removed electively. One-year post-transplant survival of VE LVAStreated patients was significantly better than that of controls (84% [158/188] vs 63% [10/16]; log rank analysis P = .0197).
Conclusion: The HeartMate VE LVAS provides adequate hemodynamic support, has an acceptably low incidence of adverse effects, and improves survival in heart transplant candidates both inside and outside the hospital. The studies of the HeartMate LVAS (both pneumatic and electric) for Food and Drug Administration approval are the only studies with a valid control group to show a survival benefit for cardiac transplantation.
The use of ventricular assist systems has dramatically improved the functional status and quality of life of patients with end-stage heart disease awaiting heart transplantation while at the same time reducing their mortality.
1-6 In fact, many clinical developments in the area of mechanical circulatory support have stemmed from progress in its use in heart transplantation candidates.
7,8 However, because of the constant shortage of donor hearts, heart transplantation has little direct statistical impact on the growing population of patients with end-stage heart disease. Research has continued to support not only the testing of mechanical devices as bridges to transplantation but also the development of long-term devices as alternatives to heart transplantation. One thrust of this research was to develop systems that would permit transplant candidates to be discharged from the hospital, a long-term goal in the development of implantable circulatory assist systems that may ultimately provide "destination therapy" for the many patients for whom heart transplantation is not an option.
Since 1986, more than 2700 patients worldwide have been treated with HeartMate Left Ventricular Assist Systems (LVAS). During this period, two multicenter clinical trials have evaluated the safety and effectiveness of the HeartMate technology as a bridge to transplantation. The first of these trials evaluated the first-generation HeartMate Implantable Pneumatic LVAS (IP LVAS) (Thermo Cardiosystems, Inc, Woburn, Mass), and until now it had been the only trial to compare an LVAS-treated group with a control group.
2,3
The second of the two multicenter clinical trials evaluated the safety and efficacy of the second-generation HeartMate Vented Electric LVAS (VE LVAS) (Thermo Cardiosystems, Inc) in transplant candidates both inside and outside the hospital. After extensive clinical evaluation, the HeartMate VE LVAS finally received approval from the Food and Drug Administration in September 1998 for use as a mechanical circulatory support bridge to cardiac transplantation, both inside the hospital and at home, in patients at imminent risk of death from heart failure. Here, we report the results of the VE LVAS multicenter trial.
Materials and methods
Device
The HeartMate VE LVAS consists of a pusher-plate blood pump driven by an integral electric motor with a percutaneous power and control circuit. The titanium alloy housing contains a flexible diaphragm that divides the interior into two chambers: one contains a textured blood path; the other houses a motor that is vented to the atmosphere via the percutaneous drive line. All bloodcontacting surfaces, except the valves, are textured. The textured polyurethane diaphragm and the sintered titanium surfaces are engineered to trap and firmly anchor blood components, thus creating a stable biologic neointima similar to the lining of natural blood vessels. Once formed, the neointima prevents blood from contacting artificial materials. In addition, the assembled pump contains 25-mm porcine xenograft valves at both the inflow and outflow conduits. The carefully designed blood path and internal flow characteristics of the VE LVAS reduce activation of coagulation and minimize the need for anticoagulants.
Like the HeartMate IP LVAS, the VE LVAS is implanted through a median sternotomy incision extending to just above the umbilicus; this is done with the use of standard cardiopulmonary bypass. An inflow cannula is placed in the left ventricular apex, and an outflow graft is passed over the diaphragm to the ascending aorta, where it is anastomosed end to side at the aortic root. The pump body resides below the diaphragm in either an intraperitoneal position or preperitoneal pocket. A detailed description of the device's orientation and implantation protocol has been published.
1
The HeartMate VE LVAS operates in one of two modes: (1) an asynchronous, fixed-rate mode programmable to a pump rate of 50 to 120 beats/min or (2) an automatic rate mode with a pump rate responsive to the pump fill rate. Once the patient's recovery from the implantation procedure is complete, wearable external components including batteries and a system controller allow the patient to be fully ambulatory and to participate in an exercise rehabilitation program (Figure 1).
|
Patient population
Between February 1996 and September 1998, 280 nonrandomized heart transplant candidates were enrolled at 24 medical centers within the United States for treatment with the HeartMate VE LVAS as a bridge to transplantation. The group consisted of 232 men (83%) and 48 women (17%) and had a median age of 55 years (range, 11-72 years).
For comparison, a historical, nonrandomized control group of 48 patients from the IP HeartMate study was used.
2 These 48 patients were enrolled at 14 of the 24 clinical centers mentioned above between April 1988 and April 1994. This group consisted of 40 men (83%) and 8 women (17%) (median age, 50 years; range, 21-67 years) who met all the criteria for device implantation but did not undergo implantation of a left ventricular assist device at the institutions participating in this study because the device was simply unavailable (before Food and Drug Administration approval, the maker's production capabilities were limited) or because the family refused treatment.
All patients in both groups had to meet specific selection criteria (Table 1). Informed consent and data from all patients were obtained in compliance with protocols approved by each institution's institutional review board. In addition, all patients in the treatment group signed an informed consent form that specifically addressed the implantation and hospital discharge (release program) phases of the trial.
|
Hemodynamic data
Hemodynamic data included mean pulmonary capillary wedge pressure (mm Hg), mean systolic blood pressure (mm Hg), and cardiac index (L · min1 · m2). These data were collected from patients at baseline (study entry), when all were already receiving maximal medical and more than half (166/328) were receiving intra-aortic balloon pump support therapy. After that, systolic blood pressure was recorded weekly throughout the study. Device performance data (ie, pump rate, pump stroke volume, pump flow, and operating mode) were recorded daily for 14 days after VE LVAS implantation and then weekly thereafter.
Hematologic data
Hematocrit, hemoglobin, platelet count, leukocyte count, plasma free hemoglobin, prothrombin time, and activated partial thromboplastin time were measured at baseline and weekly thereafter.
Biochemical data
Blood urea nitrogen, serum creatinine, total bilirubin, aspartate aminotransferase, and alanine aminotransferase were measured at baseline and weekly thereafter.
Adverse events
Adverse events (as defined below) were monitored throughout the study and were recorded as they occurred. Adverse events included bleeding, hemolysis, infection, right heart failure, renal dysfunction, hepatic dysfunction, neurologic dysfunction, pulmonary dysfunction, thromboembolic events, device malfunction, mechanical failure, and death. According to our definitions of adverse events, each adverse event was judged by the principal investigator to be either device- or patient-related. An adverse event was considered device-related if it was related or attributable to the malfunction or failure of the device or if its new onset could not be attributed to another source.
Bleeding
Bleeding was defined as blood loss serious enough to necessitate returning the patients to the operating room or to cause death (eg, cardiac tamponade).
Hemolysis
Hemolysis was defined as 2 consecutive measurements of hemoglobin values greater than 40 mg/dL.
Infection
Systemic infection was defined as infection detected by positive blood, urine, sputum, or tissue cultures in association with an elevated white blood cell count (
12,500/mL), fever (
38.1°C), and treatment with antimicrobial agents. Additionally, data were collected on drive-line infection, which was defined as any infection that (1) was detected by the positive cultures obtained from the drive-line exit site and (2) required treatment with antimicrobial agents.
Right heart failure
Right heart failure was defined as the inability of the heart to provide sufficient flow from the right ventricle to the left ventricle, thus necessitating the use of a right ventricular assist device.
Renal dysfunction
Renal dysfunction was defined as a serum creatinine value of 2.2 mg/dL or more or a blood urea nitrogen value of 50 mg/dL or more.
Hepatic dysfunction
Hepatic dysfunction was defined as a total bilirubin value of 1.4 mg/dL or more or as either an aspartate aminotransferase or alanine aminotransferase value of 50 U/L or more in cases in which the elevation in bilirubin level was not due to hemolysis.
Neurologic dysfunction
Neurologic dysfunction was defined as any central nervous system or gross neuromuscular disorder identified by standard neurologic examination of reflexes, speech, vision, and so on.
Pulmonary dysfunction
Pulmonary dysfunction was defined as a forced expiratory volume in 1 second of less than 45%. It was measured after VE LVAS implantation only if the patient had symptoms.
Thromboembolic event
A thromboembolic event was defined by clinical symptoms of stroke or by sudden neurologic, pulmonary, renal, hepatic, or peripheral vascular changes.
Device malfunction
A device malfunction was defined as any instance in which any component of the system failed to perform in the intended manner.
Mechanical failure
Mechanical failure was defined as the inability of the VE LVAS, including its backup components, to provide circulatory support.
Functional status
New York Heart Association (NYHA) functional status was evaluated in all patients at baseline, on enrollment in the release program, weekly during release, and after transplantation.
Evaluation of patients outside the hospital (release program)
The performance of the VE LVAS outside the hospital was also evaluated. Release from the hospital (the "release program") was limited to those patients who qualified to await transplantation outside of the hospital (see Table 2 for inclusion criteria).
|
Under the release program, patients were released from the hospital in a controlled, stepwise fashion. In general, they were initially allowed to leave the hospital for short day trips; they could then progress over time to full outpatient status (ie, they were allowed to live at home or outside the hospital while awaiting a transplant). All patients were required to return to the hospital weekly for evaluation (ie, NYHA functional status and hemodynamic, hematologic, and biochemical tests) and to report on their activities while away from the hospital. All adverse events (anticipated and unanticipated) that occurred away from the hospital were recorded.
Survival outcomes
Survival to transplantation and survival after transplantation of treated and control patients were recorded.
Statistical methods
Comparisons between VE LVAStreated patients and controls were performed by the Fisher exact test or the Student unpaired t tests as appropriate. Changes in hemodynamic and biochemical data from baseline to final measurement before transplantation or death were analyzed by the Student paired t test. (Baseline and final values were expressed as either mean ± SD or median and range.) Probability was 2-tailed. Survival to transplantation and survival after transplantation were analyzed by the Kaplan-Meier product limit method. Differences in survival distributions between VE LVAStreated and control groups were analyzed by means of a log-rank test. Variables for predicting survival to transplantation were identified by the Cox proportional hazards model and entered stepwise into the model. All variables selected for the hazards model were visually assessed by log-log curves to ensure compliance with the proportional hazards assumption.
Results
Patient demographics
The VE LVAStreated and control groups were similar in terms of age, sex, and distribution of patients by diagnosis (ischemic cardiomyopathy, idiopathic cardiomyopathy, and subacute myocardial infarction) (Table 3).
|
Average length of VE LVAS support
VE LVAS support lasted an average of 112 days (range, <1-691 days), with 54 patients supported for more than 180 days.
Baseline hemodynamic data
The hemodynamic indices of treated and control patients at baseline and after pharmacologic and intra-aortic balloon pump therapy were evaluated and compared (Table 4). Whereas no difference was found in the pulmonary capillary wedge pressure, the mean systolic blood pressure and mean cardiac index were significantly lower in the VE LVAStreated group than in the control group (P < .0001).
|
Hematologic data
No significant changes were found in levels of plasma free hemoglobin before (8.5 ± 12.8 mg/dL) or during LVAS support (7.4 ± 16.0 mg/dL) or in other hematologic variables (Table 5).
|
|
|
|
Infections occurred in 125 (45%) patients, most frequently at the drive-line exit site (90/125; 72%).
After device implantation, renal and hepatic dysfunction (detected by at least 1 abnormal laboratory test) was present in 158 (56%) patients and 263 (94%) patients, respectively. Of these patients, 94 (59%) had entered the trial with renal dysfunction and 174 (66%) had entered with hepatic dysfunction.
Right heart failure occurred in 31 (11%) patients, 10 of whom subsequently underwent transplantation and 21 of whom died before transplantation.
Neurologic complications occurred in 75 (27%) patients but were deemed device-related in only 14 (5%). Other neurologic complications included metabolic encephalopathy, seizures, confusion, and syncope that were attributed to a cause other than the device. Thromboembolic events occurred in 34 (12%) patients. In 17 (6%) of these, the event could not be attributed to another cause and was deemed device-related.
Mechanical failure occurred in 3 (1%) patients and in all 3 instances resulted in death. The mechanical failures were due to 2 disconnections of the outflow assembly from the pump body and 1 pump diaphragm failure. Both of the failure modes have been analyzed and pump improvements implemented. In addition, 435 confirmed device malfunctions occurred during the study period. Most of these (375/435; 86%) were malfunctions of external accessories, including the controller, batteries, battery clips, power base unit cable, and display module (external component malfunction rate = 1.192 per 100 patient-days). The controller accounted for more than half (195/375; 52%) of the external component malfunctions.
During the present study, 25 (9%) of the 280 treated patients needed to use the backup components (hand pump or IP console) because of controller or cable malfunction (11/25 patients) or pump stoppage (14/25 patients). The pump stoppages were largely due to either electrostatic discharge or component malfunctions within the pump commutator. Fixes to both of these malfunctions have been implemented. Of the 25 patients who used the backup components, 18 (72%) survived to transplantation.
Experience outside the hospital (release program)
Patient outcomes
Of 228 VE LVAStreated patients eligible to participate in the release program phase of this study, 160 (70%) actually did participate. Ultimately, only 115 of the 160 released patients achieved full outpatient status. The remaining 45 patients only left the hospital for day trips, overnight trips, and 3-day releases from the hospital and did not reach full outpatient status for a variety of reasons (eg, they were receiving transplants, the study investigators were cautious about allowing them outside the hospital, or the patients chose not to leave). The 160 patients in the release program accumulated 33.9 patient-years away from the hospital. The median length of outpatient stay was 82 days (range, 3660 days). Of the 160 patients, 138 (86%) ultimately received a transplant, 10 (6%) elected to have the device removed without a subsequent transplant, and 12 (8%) died while awaiting a transplant.
Improvement in NYHA functional class
Of the 160 VE LVAStreated patients enrolled in the release program, 153 (96%) had NYHA class IV function at baseline and 7 (4%) belonged to classes I-III. By the time these patients qualified for outpatient treatment, 91 (57%) belonged to NYHA class II and 69 (43%) belonged to NYHA class I. This change in NYHA functional class was significant (P < .001) and indicated that the improvement in functional class was due to mechanical circulatory support (Table 9).
|
|
|
|
|
|
|
Despite improvements in medical therapy for heart failure, such therapy is still of limited efficacy in patients with advanced disease.
10 This assessment is corroborated by the dismal survival (33%) in our group of control patients with end-stage disease who were not supported with an LVAS. The only survivors in the control group were those who received a heart transplant. The median waiting time for the patients who received a transplant was only 4 days. This short waiting time would be impossible today, even after aggressive medical therapy. Mechanical circulatory support with subsequent transplantation thus appears to be the only alternative for many patients with end-stage heart disease. When the long-term limitations of heart transplantation (eg, 50% mortality at 10 years), limited donor availability (2000 donors yearly), and drug-related morbidity (eg, hypertension, renal dysfunction) are considered, the role of long-term assist devices appears even more important.
11-13
In the present study, the HeartMate VE LVAS provided adequate hemodynamic support, had an acceptably low incidence of adverse effects, and improved NYHA functional class and survival in heart transplant candidates both inside and outside the hospital. The best indicator of its safety and efficacy in transplant candidates was the survival benefit it conferredboth before and after transplantation. (A similar post-transplantation survival benefit was also shown for the HeartMate IP LVAS in its earlier clinical trial.) This benefit can be attributed to the normalization of end-organ function and return to NYHA class I function in the majority of cases.
14 Consequently, concerns over the negative immunologic impact of using the VE LVAS as a bridge to transplantation (eg, an increase in reactive antibodies) are not warranted, because the survival in this patient group was excellent with no evidence of increased infection or rejection episodes.
15
Another indicator of the VE LVAS device's safety and efficacy was the rate of adverse events in this study, which mirrored the rate established in the earlier IP LVAS trial.
2 Although to a certain extent our assignment of adverse events to the device or to the patient in this study was subjective, it represented an effort to distinguish between the 2 types of causes and to assess the relative contribution of the device to adverse events while still reporting their total incidence.
In the present study, organ dysfunction, seen in most patients before implantation or immediately after implantation, was substantially corrected during LVAS support. Although in most cases aspirin and dipyridamole were the only antiplatelet agents used, the incidence of thromboembolic events in the present trial was low and constituted one of the major advantages of using the VE LVAS. On the other hand, the incidence of bleeding was relatively high but not surprisingly so in light of the severity of the disease and expected development of coagulopathies.
16,17 In present clinical practice, this problem is apparently being reduced by the more routine use of antifibrinolytic agents.
17,18 Similarly, the incidence of infections, and especially drive-line infections, was high; however, design modifications in the VE LVAS, along with better timing of device placement, might decrease infection rates in the future.
19,20 Our pulmonary dysfunction data were of limited value because forced expiratory volume in 1 second was not measured in all patients. The incidence of right heart failure and subsequent high mortality justifies the more routine use of nitric oxide, prostaglandins, and phosphodiesterase inhibitors to prevent this complication.
21 Unfortunately, no data have been collected on the incidence of right-sided dysfunction and the subsequent need for prolonged use of intravenous inotropic agents.
Another interesting finding of the present multicenter trial, which followed the experiences of several single centers in the use of LVAS as a bridge to recovery,
4,22 was the outcome of patients who underwent elective device explantation. Most previous findings have been limited to that subset of patients being bridged to transplantation and therefore not selected as candidates for recovery. Thus, as our understanding of end-stage myocardial disease and the processes involved in recovery improves, so might the outcomes of patients treated with the VE LVAS. In fact, experience gained from the use of left ventricular assist devices as a bridge to transplantation could be useful in selecting patients with end-stage cardiac disease who would be good candidates for implantation (ie, those at low risk for perioperative complications and thus having a better chance of recovery).
The ultimate goal in developing ventricular assist systems is to create technology that will allow the transplant candidate to lead a normal, active life while receiving mechanical circulatory support. The present multicenter trial suggests that the VE LVAS improves the outcome of patients with end-stage heart disease who are awaiting a transplant, and it has defined a subset of patients who cannot be supported by medical therapy. It also provides important insights into any future application of this technology for permanent use as heart replacement therapy (eg, the REMATCH trial).
23 Moreover, in the field of heart transplantation, it is the only study with a control cohort (ie, patients who did not have an LVAS implanted but otherwise met all inclusion criteria) to show a survival benefit for transplantation. This observation is especially important since data on comparable groups of patients (eg, patients with end-stage heart disease who fulfill criteria for left ventricular assist device implantation) are hard to obtain now owing to the ethical issues involved. We believe that the control group we used is the best comparison group for patients with terminal disease that we are likely to develop.
With better medical therapy, it is more difficult to determine the survival benefit of heart transplantation. In a recent study of German heart transplant programs, Deng and colleagues
24 showed that a survival benefit was achieved only in the sickest patients. However, that study lacked a control cohort. In the present study, the patients who were on the waiting list for a transplant and reached the level of cardiac deterioration that would qualify them for insertion of a HeartMate LVAS (either pneumatic or electric) achieved a survival benefit only after LVAS implantation and subsequent cardiac transplantation. Furthermore, the survival benefit was much greater for those patients who received the pump than for those who did not.
Appendix
Institutions/investigators
Brigham and Women's Hospital, Boston, Massachusetts
Gregory Couper, MD
Nancy Cummings, RN
The Cleveland Clinic Foundation, Cleveland, Ohio
Patrick McCarthy, MD
Rita Vargo, RN
Ann Foray-Kaplon, RN
Cathie Vacha, RN
Sheryl Kamis, RN
Michael Yeager, RN
Columbia Presbyterian Medical Center, New York, New York
Eric A. Rose, MD
Mehmet Oz, MD
Katharine Catanese, RN, MSN
Fairfax Hospital, Falls Church, Virginia
Nelson Burton, MD
Aaron Hill, CCP
Karen Muncy
Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania
Michael Acker, MD
Mary Lou O'Hara, RN, MSN
LDS Hospital, Salt Lake City, Utah
James Long, MD
Jennifer Pitt, RN
John Marks
Ann Bagley
Loyola University Medical Center, Joliet, Illinois
Bryan Foy, MD
Kimberly Martin, RN
Marianne Laff, RN
Lutheran Hospital of Indiana, Fort Wayne, Indiana
Alan Peterson, MD
Margaret Scatena, RN
Linda Beatty, RN
Medical College of Virginia, Richmond, Virginia
George Maier, MD
Kathy Baker, RN
Maureen Flattery, RN
Columbia Hospital at Medical City, Dallas, Texas
Michael Mack, MD
Tollea McMellon, RN
Newark Beth Israel Medical Center, Newark, New Jersey
Laszlo Fuzesi, MD
Louise Klapp, RN, CCP
Elizabeth Cortez, RN, CCP
Ochsner Foundation Hospital, New Orleans, Louisiana
Clifford Van Meter, MD
Sonja Hollenbach, RN
Patricia Uber, PharmD
Lindsay Kersker, RN
Rush-PresbyterianSt Luke's Medical Center, Chicago, Illinois
William Piccione, MD
Annette Mattea, RN, MSN
Kathleen Grady, RN, PhD
Sacred Heart Medical Center, Spokane, Washington
Timothy Icenogle, MD
David Sato
Steve Himley
Debbie Butters, RN, MSN
Mark Puhlman, RN, MSN
Sharp Memorial Hospital, San Diego, California
Walter Dembitsky, MD
Suzanne Chillcott, RN
St Joseph's Hospital, Atlanta, Georgia
Douglas Murphy, MD
Paige Davis, RN
Susan Leighty
St Luke's Medical Center, Milwaukee, Wisconsin
Alfred Tector, MD
Diane Dressler, RN, MSN
Paulette Schauer, RN
Lynne Mathiak, RN
Temple University Hospital, Philadelphia, Pennsylvania
Valluvan Jeevanandam, MD
Satoshi Furukawa, MD
Barbara Todd, RN, MSN
Joelle Hargraves, RN, MSN
Texas Heart Institute, Houston, Texas
O. H. Frazier, MD
Timothy Myers
Dan Tamez
Kathy Miller
Washington Hospital Center, Washington, DC
Steven Boyce, MD
Leslie Sweet, RN
Sheila Walsh, RN
University of Alabama Medical Center, Birmingham, Alabama
William Holman, MD
Linda Brantley, PA
Candace Wainscott, RN
University of Iowa, Iowa City, Iowa
Wayne Richenbacher, MD
Sara Vance, RN
Carolyn Laxson, RN
University of Michigan Medical Center, Ann Arbor, Michigan
Francis Pagani, MD
Hilary Monaghan, RN
University of Wisconsin Hospital and Clinics, Madison, Wisconsin
Robert Mentzer, MD
Charles Canver, MD
Linda Jacobs, RN
Janice Yakey, RN
Acknowledgments
We thank Laura Damme, RN, and Allen Alexander, MD, for their assistance in preparing the manuscript; Mr Jerry Heatley for performing the statistical analyses required for this article; and the Section of Scientific Publications at the Texas Heart Institute for editorial assistance.
References
This article has been cited by other articles:
![]() |
H. Kamiya, A. Koch, F.-U. Sack, P. Akhyari, A. Remppis, T. J. Dengler, M. Karck, and A. Lichtenberg Who needs 'bridge' to transplantation in the presence of the Eurotransplant high-urgency heart transplantation program? Eur. J. Cardiothorac. Surg., December 1, 2008; 34(6): 1129 - 1133. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. John, F. Kamdar, K. Liao, M. Colvin-Adams, L. Miller, L. Joyce, and A. Boyle Low thromboembolic risk for patients with the Heartmate II left ventricular assist device. J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1318 - 1323. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. John, F. Kamdar, K. Liao, M. Colvin-Adams, A. Boyle, and L. Joyce Improved Survival and Decreasing Incidence of Adverse Events With the HeartMate II Left Ventricular Assist Device as Bridge-to-Transplant Therapy. Ann. Thorac. Surg., October 1, 2008; 86(4): 1227 - 1235. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Greenberg, B. Czerska, R. M. Delgado, R. Bourge, M. R. Zile, M. Silver, M. Klapholz, E. Haeusslein, M. R. Mehra, P. Mather, et al. Effects of Continuous Aortic Flow Augmentation in Patients With Exacerbation of Heart Failure Inadequately Responsive to Medical Therapy: Results of the Multicenter Trial of the Orqis Medical Cancion System for the Enhanced Treatment of Heart Failure Unresponsive to Medical Therapy (MOMENTUM) Circulation, September 16, 2008; 118(12): 1241 - 1249. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Patel, E. S. Weiss, J. Schaffer, S. L. Ullrich, D. C. Rivard, A. S. Shah, S. D. Russell, and J. V. Conte Right Heart Dysfunction After Left Ventricular Assist Device Implantation: A Comparison of the Pulsatile HeartMate I and Axial-Flow HeartMate II Devices Ann. Thorac. Surg., September 1, 2008; 86(3): 832 - 840. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Osaki, N. M. Edwards, M. Velez, M. R. Johnson, M. A. Murray, J. A. Hoffmann, and T. Kohmoto Improved survival in patients with ventricular assist device therapy: the University of Wisconsin experience. Eur. J. Cardiothorac. Surg., August 1, 2008; 34(2): 281 - 288. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Aggarwal, F. Cheema, M. C. Oz, and Y. Naka Long-Term Mechanical Circulatory Support Card. Surg. Adult, January 1, 2008; 3(2008): 1609 - 1628. [Full Text] |
||||
![]() |
O.H. Frazier, T. J. Myers, and I. Gregoric Total Artificial Heart Card. Surg. Adult, January 1, 2008; 3(2008): 1629 - 1638. [Full Text] |
||||
![]() |
J. Haft, W. Armstrong, D. B. Dyke, K. D. Aaronson, T. M. Koelling, D. J. Farrar, and F. D. Pagani Hemodynamic and Exercise Performance With Pulsatile and Continuous-Flow Left Ventricular Assist Devices Circulation, September 11, 2007; 116(11_suppl): I-8 - I-15. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Stone Current Status of Mechanical Circulatory Assistance Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2007; 11(3): 185 - 204. [Abstract] [PDF] |
||||
![]() |
S. Chumnanvej, M. J. Wood, T. E. MacGillivray, and M. F. V. Melo Perioperative Echocardiographic Examination for Ventricular Assist Device Implantation Anesth. Analg., September 1, 2007; 105(3): 583 - 601. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. W. Miller, F. D. Pagani, S. D. Russell, R. John, A. J. Boyle, K. D. Aaronson, J. V. Conte, Y. Naka, D. Mancini, R. M. Delgado, et al. Use of a Continuous-Flow Device in Patients Awaiting Heart Transplantation N. Engl. J. Med., August 30, 2007; 357(9): 885 - 896. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Rogers, J. Butler, S. L. Lansman, A. Gass, P. M. Portner, M. K. Pasque, R. N. Pierson III, and for the INTrEPID Investigators Chronic Mechanical Circulatory Support for Inotrope-Dependent Heart Failure Patients Who Are Not Transplant Candidates: Results of the INTrEPID Trial J. Am. Coll. Cardiol., August 21, 2007; 50(8): 741 - 747. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Slaughter, S. S. Tsui, A. El-Banayosy, B. C. Sun, R. L. Kormos, D. K. Mueller, H. T. Massey, T. B. Icenogle, D. J. Farrar, J. D. Hill, et al. Results of a multicenter clinical trial with the Thoratec Implantable Ventricular Assist Device J. Thorac. Cardiovasc. Surg., June 1, 2007; 133(6): 1573 - 1580. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. D. Feller, E. N. Sorensen, M. Haddad, R. N. Pierson, F. L. Johnson, J. M. Brown, and B. P. Griffith Clinical Outcomes Are Similar in Pulsatile and Nonpulsatile Left Ventricular Assist Device Recipients Ann. Thorac. Surg., March 1, 2007; 83(3): 1082 - 1088. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schenk, A. El-Banayosy, W. Prohaska, L. Arusoglu, M. Morshuis, W. Koester-Eiserfunke, L. Kizner, E. Murray, P. Eichler, R. Koerfer, et al. Heparin-induced thrombocytopenia in patients receiving mechanical circulatory support J. Thorac. Cardiovasc. Surg., June 1, 2006; 131(6): 1373 - 1381. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Padalino, R. G. Ohye, A. C. Chang, R. J. Gajarski, E. L. Bove, and E. J. Devaney Bridge to Transplant Using the MicroMed DeBakey Ventricular Assist Device in a Child with Idiopathic Dilated Cardiomyopathy. Ann. Thorac. Surg., March 1, 2006; 81(3): 1118 - 1121. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schenk, P. M. McCarthy, E. H. Blackstone, J. Feng, R. C. Starling, J. L. Navia, L. Zhou, K. J. Hoercher, N. G. Smedira, and K. Fukamachi Duration of inotropic support after left ventricular assist device implantation: Risk factors and impact on outcome J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 447 - 454. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Konstam, B. Czerska, M. Bohm, R. M. Oren, J. Sadowski, S. Khanal, W. T. Abraham, A. Wasler, J. B. Dahm, A. Gavazzi, et al. Continuous Aortic Flow Augmentation: A Pilot Study of Hemodynamic and Renal Responses to a Novel Percutaneous Intervention in Decompensated Heart Failure Circulation, November 15, 2005; 112(20): 3107 - 3114. [Abstract] [Full Text] [PDF] |
||||