JTCS Speed Up Your Browser
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Hiroji Imamura
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Saito, Y.
Right arrow Articles by Okamura, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Saito, Y.
Right arrow Articles by Okamura, A.
Related Collections
Right arrow Trachea and bronchi

J Thorac Cardiovasc Surg 2002;123:161-167
© 2002 The American Association for Thoracic Surgery


General Thoracic Surgery

New tubular bioabsorbable knitted airway stent: Biocompatibility and mechanical strength

Yukihito Saito, MDa, Kenichirou Minami, MDa, Masashi Kobayashi, MDa, Yoshihisa Nakao, MDa, Hideyasu Omiya, MDa, Hiroji Imamura, MDa, Noriko Sakaida, MDb, Akiharu Okamura, MDc

From the Department of Thoracic and Cardiovascular Surgery,a the Division of Surgical Pathology,b Kansai Medical University, Moriguchi, and the Division of Surgical Pathology,c International Center for Diagnostic Pathology, Kobe, Japan.

Received for publication April 3, 2001. Revisions requested May 14, 2001; revisions received June 4, 2001. Accepted for publication July 3, 2001. Address for reprints: Yukihito Saito, MD, Associate Professor, Department of Thoracic and Cardiovascular Surgery, Kansai Medical University, 10-15 Fumizonocho, Moriguchi, Osaka 570-8507, Japan (E-mail: saitoy@ takii.kmu.ac.jp).


    Abstract
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Objective: This study examines the biocompatibility and suitability of a new tubular bioabsorbable knitted stent made of poly-L-lactic acid in normal rabbit airways and examines the mechanical strength of this stent in vitro.
Methods: A tubular knitted airway stent (group B, n = 15) made of poly-L-lactic acid wire was implanted operatively in New Zealand White rabbits intratracheally; silicone stents served as controls (group A, n = 8). The cervical trachea was exposed, and the stent was implanted. Up to 40 weeks after stent implantation, the rabbits were killed, at which time bronchoscopy, histologic examination, and scanning electron microscopic study was done. We tested poly-L-lactic acid stents and silicone stents for their mechanical strength in vitro. We subjected stents to area loads and measured their mechanical strengths.
Results: In group A, which received silicone stents, 3 (37.5%) rabbits died within 4 weeks of stent implantation as a result of airway obstruction by secretions inside the stent lumen. In group B, poly-L-lactic acid stents, 1 (6.7%) rabbit died 3 weeks after implantation because of weakness caused by anorexia. In the remaining animals, except for 1 animal with stent trouble, the bronchial lumen was fully open until the 40th week after implantation. After 40 weeks of follow-up, the stents disappeared, except for nonabsorbable suture in the bronchial wall. None of the animals in group B died of airway complication. Histologic examination and scanning electron microscopic examination of the group A silicone stents showed marked regression of ciliated cells under the stent. In group B the ciliated epithelium was preserved, and there were numerous capillary blood vessels in the submucosa. In scanning electron microscopy of the group B poly-L-lactic acid stents, the ciliated cells were preserved between the mesh holes of the stent. For diameters between 4 and 6 mm, the mechanical strength of silicone stents was greater than that of poly-L-lactic acid stents. However, the mechanical strength of poly-L-lactic acid stents increased as a function of their diameter.
Conclusion: A new tubular bioabsorbable stent made of poly-L-lactic acid is biocompatible in normal rabbit airways, indicating that poly-L-lactic acid is a promising material for airway stents for clinical use.


    Introduction
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Tracheobronchial stenoses can be caused by tracheobronchomalacia, extrinsic compression, postintubation tracheal injuries, sequelae after tracheostomy, or malignant or benign tumors. An ideal stent has yet to be developed. The most common types of stents in clinical use are made of silicone or metallic wire.

The Dumon stent, a silicone tube, has been well established, especially for treatment of stenoses with intraluminal tumor growth or granulomatous tissue. A metallic stent, the nitinol self-expandable airway stent, is usually used in patients with airway stenosis because of extrinsic tumoral compression.

Metallic stents are sometimes used in patients with benign bronchial stenosis after lung transplantation caused by granulation of the bronchial anastomosis site or in pediatric patients with congenital tracheal stenosis and tracheobronchomalacia. Metallic stents have the following advantages compared with silicone stents: (1) simplicity of insertion and fixation; (2) better clearance secretions; (3) accommodation to varying tracheal dimensions; and (4) a low internal/external diameter ratio. However, metallic stents are not without problems. Once covered by epithelium, they are not easy to remove with a conventional bronchoscopic procedure alone, and an operation may be required.

Silicone stents can be removed, but they have several problems, such as disturbance of the physiologic mucociliary function of the tracheobronchial epithelium under the stent. Silicone stents are relatively thick and cause mucociliary function to be lost in the stented area. Secretions tend to accumulate in the lumen, which can cause obstruction.

In the pediatric patient who is growing, it may be necessary to exchange existing stents for larger ones. However, this is not easily done if the stent is metallic. Bioabsorbable airway stents, on the other hand, offer benefits: extraction of the device is unnecessary, and normal airway is preserved after stent resorption.

The aim of this study was to examine the biocompatibility and mechanical strength of a new tubular bioabsorbable stent made of poly-L-lactic acid (PLLA).


    Materials and methods
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Evaluation of biocompatibility in 2 types of airway stents
The PLLA knitted tubular tracheal stents were made of 170-µm wire with an outer diameter of 6 mm and a length of 12 mm. The silicone stents were tubes 0.8 mm thick, with an outer diameter of 6 mm and a length of 12 mm (Figure 1).



View larger version (87K):
[in this window]
[in a new window]
 
Fig. 1. Silicone stent (left) and tubular knitted PLLA stent (right).

 
The rabbits (New Zealand White rabbits) were allocated to one of 2 groups. We used silicone stents for control group A, which comprised 8 rabbits with a mean weight of 3.10 kg (range, 2.70-3.50 kg). We used PLLA stents in experimental group B, which comprised 15 rabbits with a mean weight of 3.10 kg (range, 2.65-3.60 kg).

The rabbits were anesthetized with subcutaneous injection of atropine (0.75 mg) and inhalation of 1.5% halothane. The animals maintained spontaneous breathing without intubation. In the midline the cervical trachea was prepared under direct visualization for implantation of the stent. Unnecessary dissection was avoided so as not to interfere with the blood circulation of the trachea. The trachea was opened transversely between cartilage rings for two thirds of its circumference. The stent was implanted intratracheally and fixed with the same continuous 5-0 polypropylene suture used to close the tracheostomy.

In group A the rabbits were killed at 8 weeks (5 rabbits) after stent implantation. In group B the animals were killed at 8 weeks (5 rabbits), 20 weeks (5 rabbits), and 40 weeks (4 rabbits) after stent implantation, at which time tracheal endoscopy was done. The cervical trachea was excised and divided longitudinally into two pieces, one fixed in formalin solution for histologic examination and the other in buffered glutaraldehyde solution for scanning electron microscopic study. All of the surgical and euthanasia procedures were performed in accordance with the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources, National Research Council, and published by the National Academy Press, revised 1996.

In vitro evaluation of mechanical strength in 2 types of airway stents
We tested PLLA stents and silicone stents in vitro for their mechanical strength. Using a digital force gauge (DFG-0.2K; Shimpo Corp, Kyoto, Japan) and a noncontact profile projector (PJ 300; Mitsutoyo Corp, Kyoto, Japan), we subjected stents to area loads at room temperature (22°-24°C) and measured their deformation rates. Area load was applied over the full length of the stent with a 50-mm2 plate mounted on a digital force gauge with a measuring range of ±1.960 N and an accuracy of ±1 mN. The stent diameter before and after loading was measured 6 times each on the screen of a noncontact profile projector to an accuracy of 1/1000 mm (Figure 2). We calculated the deformation rate as follows:



View larger version (15K):
[in this window]
[in a new window]
 
Fig. 2. Schema of measurement of diameter of stents with a noncontact profile projector.

 

    Results
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
Evaluation of biocompatibility in 2 types of airway stents
All rabbits tolerated the surgical operation and the implantation of the stent. Perioperative complications included airway obstruction and, in the neck of 1 rabbit, a subcutaneous abscess.

In group A, which received silicone stents, 3 rabbits died within 4 weeks of stent implantation as a result of airway obstruction by secretions inside the stent lumen. The 4-week survival of the rabbits in group A was 62.5%. In the remaining rabbits stridor developed within 8 weeks after stent implantation. Therefore, all rabbits surviving up to 8 weeks after stenting were killed in the eighth postoperative week. Obstructing material was visible inside the stents of all rabbits (Figure 3).



View larger version (38K):
[in this window]
[in a new window]
 
Fig. 3. Bronchoscopic findings in group A (silicone): A, Immediately after stent implantation; B, severe airway stenosis 8 weeks after implantation; and C, after removal of silicone stent, obstructive material was visible inside the stent.

 
In group B, which received PLLA stents, only 1 animal died 3 weeks after implantation because of weakness caused by anorexia. At autopsy, this animal was found to have a subcutaneous abscess of the neck, but the bronchial lumen was not stenosed. The 4-week survival of the rabbits in group A was 93.3%. In the remaining animals, except for 1 animal with stent trouble (incomplete expansion of the knitted wire), the bronchial lumen was fully open until the 40th week after stent implantation (Table 1). After 40 weeks' follow-up, the stents disappeared, except for nonabsorbable sutures in the bronchial wall (Figure 4). None of rabbits in group B died of airway complications.


View this table:
[in this window]
[in a new window]
 
Table 1. Postimplantation course of animals
 


View larger version (109K):
[in this window]
[in a new window]
 
Fig. 4. Bronchoscopic findings in group B (PLLA). A, Immediately after stent implantation; B, 8 weeks after implantation, airway was fully open; C, 20 weeks after implantation, a part of the PLLA stent was not visible because it was covered with tracheal epithelium; and D, 40 weeks after implantation, the PLLA stent was absorbed, and nonabsorbable suture in the bronchial wall was observed.

 
Histologic examination of the group A silicone stents at 8 weeks after implantation showed marked regression of ciliated cells under the stent. In some areas the epithelium had eroded, and there was chronic lymphocytic inflamation in the submucosa (Figure 5, A). There was mucosa swelling at both ends of the stents. In the group B PLLA stents the ciliated epithelium was preserved, and there were numerous capillary blood vessels in the submucosa at 8 weeks after implantation(Figure 5Go, B). Scanning electron microscopy of the group A silicone stents at 8 weeks after implantation showed marked regression of ciliated cells under the stent (Figure 6, A and B).



View larger version (58K):
[in this window]
[in a new window]
 
Fig. 5. Histologic findings in both groups. A, Epithelial erosion was observed under the stent 8 weeks after implantation (group A). B, Epithelium was preserved, and numerous capillary blood vessels in the submucosa were observed 8 weeks after implantation (group B).

 


View larger version (173K):
[in this window]
[in a new window]
 
Fig. 6. Scanning electron microscopic findings in both groups. A and B Marked regression of ciliated cells under the stent 8 weeks after implantation (group A). (A, original magnification 500 x; B, original magnification 2000x.) C and D, Ciliated cells were preserved between the mesh holes of the stent 8 weeks after implantation (group B). (C, original magnification 100x; D, original magnification 1000x.) *PLLA wire.

 
In the group B PLLA stents the ciliated cells were preserved between the mesh holes of the stent 8 weeks after implantation(Figure 6Go, C and D).

In vitro evaluation of mechanical strength in 2 types of airway stents
For diameters between 4 and 6 mm, the deformation rate of the silicone stent was lower than that of the PLLA stent. However, the mechanical strength of PLLA stents increased as a function of their diameter. At a diameter of 8 mm, the deformation rate of PLLA stents was similar to that of silicone stents. At diameters of 10 and 13 mm, the deformation rate of PLLA stents was significantly lower than that of silicone stents (Figure 7).



View larger version (18K):
[in this window]
[in a new window]
 
Fig. 7. Comparison of deformation rate in 2 types of airway stents. Between 4 and 6 mm, the deformation rate of silicone stents was significantly lower than that of PLLA stents. At diameters of 10 and 13 mm, the deformation rate of PLLA stents was significantly lower than that of silicone stents.

 

    Discussion
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 
After the introduction of the MontgomeryGo 1 T-tube in 1965, one of the most widely used types of silicone stents was designed by Dumon.Go 2 Because silicone stents are relatively thick, the internal diameter of the stent is thus smaller than the normal airway lumen. The stents interfere with normal airway mucociliary function, which can result in accumulation of secretions inside the stent and obstruction of its lumen.

The ideal stent should be simple to insert, fix, and remove; be biocompatible; not obstruct the airway; allow clearance of secretions; and accommodate to varying tracheal dimensions and shapes.Go Go 3,4

Metallic expandable wire stents, such as the Wallstent,Go 5 have a low internal/external diameter ratio. They cannot usually migrate after insertion, and the mucociliary clearance function is better maintained. However, once covered by epithelium, it is not easy to remove them with a bronchoscopic procedure. An operation may be required.

Silicone and metallic stents used in the tracheobronchial tree for the treatment of airway problems both have disadvantages, as mentioned above. Long-term results in tracheobronchial airway stenting after implantation has not fully been discussedGo 6 because airway stents are usually only used for short-term maintenance of quality of life in patients who have advanced malignant airway diseases, mainly caused by lung cancer.

In the pediatric patient who is growing, it may be necessary to exchange existing stents for larger ones. However, this is not easily done if the stent is metallic. Although silicone stent can be exchanged, it is unfavorable for stenting the small airway in a child because of its thick wall. Its most disadvantageous feature is its disturbance of the physiologic mucociliary function of the small airway.

The new tubular bioabsorbable knitted airway stent has advantages, such as a low internal/external diameter ratio and good clearance secretions. Bioabsorbable stents are optimal if the need for stenting the airway is only temporary. For benign and post-transplantation stenosis in particular, a bioabsorbable stent can be an alternative means of treatment.

The biocompatibility of bioabsorbable materials, such as PLLA and polyglycolic acid, has been good in other organs. Rods of bioabsorbable PLLA and pledgets of polyglycolic acid are in clinical use for the fixation of bone fractures and reinforcement of suture lines in lung parenchyma, with favorable results.Go Go 7-9 In our study PLLA was chosen for an absorbable bronchial stent because it has the longest biodegradation time of the basic molecules available. Its biodegradation time is about 6 months. The bioabsorbable PLLA stent examined in our study is a different type from the spiral stent previously reported.Go Go 10,11

The new type of PLLA stents are mechanically strong enough to retain their tubular shape against extrinsic mechanical force. The mechanical strength of the PLLA stent increased as a function of diameter. At diameters of 4 and 6 mm, the PLLA stents we tested were mechanically weaker than the silicone stents; however, knitted tubular stents made of PLLA wire thicker than 170 µm can be as strong as silicone stents. At diameters of 10 mm or more, the mechanical strength of PLLA stents was stronger than that of silicone stents. Therefore, for patients with various degrees of airway diseases, knitted tubular PLLA stents can be used like silicone stents.

In conclusion, the biocompatibility of airway stents made of PLLA is good enough to make it a promising material for airway stents in clinical use. This bioabsorbable airway stent can be useful in the treatment of tracheomalacia in a child who is growing.

This bioabsorbable knitted tubular stent and its construction require further investigation before clinical applications begin.


    References
 Top
 Abstract
 Introduction
 Materials and methods
 Results
 Discussion
 References
 

  1. Montgomery WW. T-tube tracheal stent. Arch Otolaryngol. 1965;82:320-1.
  2. Dumon JF. A dedicated tracheobronchial stent. Chest. 1990;97:348-32.
  3. Loeff DS, Filler RM, Gorenstein A, Ein S, Philippart A, Bahoric A, et al. A new intratracheal stent for tracheobronchial reconstruction: experimental and clinical studies. J Pediatr Surg. 1988;23:1173-7.[Medline]
  4. Becker HD. Stenting of central airways. J Bronchology. 1995;2:98-106.
  5. Bolliger CT, Heitz M, Hauser R, Probst A, Perruchoud AP. An airway Wallstent for the treatment of tracheobronchial malignancies. Thorax. 1996;51:1127-9.[Abstract/Free Full Text]
  6. Mehta AC, Dasgupta A. Airway stents. Clin Chest Med. 1999;20:139-51.[Medline]
  7. Majola A, Vainionpaa S, Vihtonen K. Absorption, biocompatibility and fixation properties of polylactic acid in bone tissue: an experimental study in rats. Clin Orthop. 1991;268:260-9.
  8. Hirata T, Fukuse T, Mizuno S, Wada H. Clinical application of biodegradable rib connecting pins in thoracotomy. Thorac Cardiovasc Surg. 1999;47:183-7.[Medline]
  9. Nakamura T, Shimizu Y, Watanabe S, Hitomi S, Kitano M, Matsunobe S. A new bioabsorbable pledgets and non-woven fabrics made from polyglycolide (PGA) for pulmonary surgery. Thorac Cardiovasc Surg. 1990;38:81-5.[Medline]
  10. Korpela A, Aarnio P, Sariola H, Tormala P, Harjula A. Comparison of tissue reactions in the tracheal mucosa surrounding a bioabsorbable and silicone airway stents. Ann Thorac Surg. 1998;66:1772-6.[Abstract/Free Full Text]
  11. Korpela A, Aarnio P, Sariola H, Tormala P, Harjula A. Bioabsorbable self-reinforced poly-L-lactide, metaric, and silicone stents in the management of experimental tracheal stenosis. Chest. 1999;115:490-5.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
RadiologyHome page
R. H. Thornton, R. L. Gordon, R. K. Kerlan, J. M. LaBerge, M. W. Wilson, K. A. Wolanske, M. B. Gotway, G. S. Hastings, and J. A. Golden
Outcomes of tracheobronchial stent placement for benign disease.
Radiology, July 1, 2006; 240(1): 273 - 282.
[Abstract] [Full Text] [PDF]


Home page
ChestHome page
M. Noppen, G. Stratakos, J. D'Haese, M. Meysman, and W. Vinken
Removal of Covered Self-Expandable Metallic Airway Stents in Benign Disorders: Indications, Technique, and Outcomes
Chest, February 1, 2005; 127(2): 482 - 487.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
Y. Saito, K.-i. Minami, H. Kaneda, T. Okada, T. Maniwa, Y. Araki, H. Imamura, H. Yamada, K. Igaki, and H. Tamai
New Tubular Bioabsorbable Knitted Airway Stent: Feasibility Assessment for Delivery and Deployment in a Dog Model
Ann. Thorac. Surg., October 1, 2004; 78(4): 1438 - 1440.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Respir. Crit. Care Med.Home page
A. Ernst, D. Feller-Kopman, H. D. Becker, and A. C. Mehta
Central Airway Obstruction
Am. J. Respir. Crit. Care Med., June 15, 2004; 169(12): 1278 - 1297.
[Abstract] [Full Text] [PDF]


Home page
Asian Cardiovasc. Thorac. Ann.Home page
A. Dodge-Khatami, H. W. Niessen, L. H Koole, M. G Klein, T. M van Gulik, and B. A. de Mol
Tracheal Replacement in Rabbits with a New Composite Silicone-Metallic Prosthesis
Asian Cardiovasc Thorac Ann, September 1, 2003; 11(3): 245 - 249.
[Abstract] [Full Text]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Hiroji Imamura
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Saito, Y.
Right arrow Articles by Okamura, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Saito, Y.
Right arrow Articles by Okamura, A.
Related Collections
Right arrow Trachea and bronchi


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS