Preservation of the aortic valve offers advantages to a composite valve graft (CVG) procedure in certain patients with aortic root aneurysms. Failures have been noted,
1-4 however, and at present, the most durable aortic root-sparing operation appears to be the reimplantation procedure originally described by David and Feindel (the T. David-I operation).
2,3 A tubular polyester graft sized by using David's original formula
4 (ie, Graft diameter = [(hleaflet x 2) 0.67] + [2 hAo wall]) is attached to the ventriculoaortic junction beneath the leaflet insertion level, and the native valve is sutured inside the graft. The coronary ostia are reimplanted as Carrel buttons. Reimplantation operations have met with considerable clinical success
3 since 1988, but durability is not perfect.
3,5-7 We describe successful use of a stentless porcine valve in a patient with severe aortic regurgitation (AR) after David-I-type valve-sparing aortic root replacement.
Clinical summary
A 51-year-old female nurse presented to Stanford University Medical Center with heart failure symptoms caused by severe recurrent AR. Her medical history included type I diabetes mellitus and a family history of an uncharacterized aortic disorder that had resulted in the sudden deaths of her mother, sister, grandfather, and great-grandfather. Because of an increase in aortic root diameter (which was only modestly dilated, 4.2 cm), mild central AR in a trileaflet valve, and family history, she underwent a reimplantation type of aortic valve-sparing aortic root replacement in July 1993 with a 26-mm woven, double-velour, Hemashield Dacron graft (Medi-Tech; Boston Scientific Corp, Natick, Mass). The classic T. David-I procedure was used. She had no complications, a discharge echocardiogram showed no residual AR, and, subsequently, she was clinically well. Three years later, mild AR was noted on an echocardiogram, but she was free of symptoms. Four years postoperatively, she began having mild shortness of breath with exertion; an echocardiogram showed moderate AR. In November 1999 (6
years postoperatively), she began having more shortness of breath and had moderate-to-severe AR on echocardiography (Figure 1). Reoperation was recommended in January 2000. A computed tomographic angiography scan and angiogram showed bovine arch anatomy, no evidence of downstream aortic dilatation or dissection, intact ascending aortic graft and coronary anastomoses, and normal coronary arteries. The exact mechanism of the severe AR could not be identified by means of echocardiography, and moderate left ventricular (LV) dilatation and global hypokinesia were also present. Because of complications of her diabetes (peripheral neuropathy, peripheral vascular disease, and cutaneous feet ulcers), she had an aversion to warfarin anticoagulation and insisted on another tissue valve.
|
The patient's hospital course was unremarkable. A follow-up echocardiogram 12 months later showed mild LV hypertrophy without dilatation and normal LV systolic function. The aortic root again was normal size, and the SPV valve leaflets appeared normal (Figure 2). There was no AR, and the maximal flow velocity across the SPV was 2.2 m/s, corresponding to a peak transvalvular gradient of approximately 19 mm Hg (mean gradient not calculated). The patient is clinically doing well 2 years after reoperation.
|
Both the T. David reimplantation
3 and the Yacoub remodeling
8 valve-sparing aortic root replacement procedures provide a reasonable alternative in selected patients with aortic root aneurysms or dissections who wish to avoid anticoagulation. Durability is excellent out to 5 to 10 years,
3 but not enough patients remain at risk beyond 10 years to base sound inferences on these data.
3,8,9 On the other hand, the long-term durability of the Yacoub-type procedure in patients with Marfan syndrome is not as satisfactory.
1 Should these operations fail, surgical treatment has consisted of either orthotopic aortic valve replacement (usually with a mechanical valve) inside the graft or total aortic root replacement with a CVG. Because of the constraints imposed by working inside the aortic graft, visualization and exposure are not ideal, even if one elects the simple aortic valve replacement option, and the valve size is usually small; conversely, redoing the entire aortic root replacement as a CVG is a bigger operation. In this case we describe a patient in whom it was possible to implant a stentless porcine valve inside the aortic graft after failure of a previous T. David-I procedure to avoid the need for anticoagulation.
5,6 Although a stented pericardial or porcine valve could also have been used, only a very small bioprosthesis would have fit. Redo CVG with either a Freestyle stentless porcine aortic root (Medtronic, Inc, Minneapolis, Minn) or a homemade CVG with a stented bioprosthesis would also have been reasonable options, but this would be a much more complicated operation. The simplified approach we used was not difficult, but the field was somewhat crowded; of course, the size of the valve is limited by the diameter of the graft used at the initial operation. Parenthetically, either a scalloped subcoronary Medtronic Freestyle or a St Jude Medical SPV stentless valve could have been used here. The short-term (2 years) clinical result in this patient argues for wider consideration of this particular tissue valve alternative in what is an admittedly rare circumstance in which the patient refuses warfarin or is medically noncompliant or anticoagulation is contraindicated.
The reason for structural valve deterioration (SVD) of the aortic valve is not clear. Previous reports documenting SVD after the T. David-I operation have noted, as in our case, that the aortic leaflets were thickened and retracted; alternatively, geometric distortion caused by technical error can cause late AR,
9 or occasionally, an adolescent patient can outgrow the fixed ventriculoaortic junction with resulting stenosis.
3-7 It is possible that these changes were caused by abnormal coaptation of the aortic valve leaflets caused by incorrect sizing of the Dacron graft
6 or that the original T. David-I technique does not create pseudosinuses in the graft (as does the Yacoub remodeling method), such that the leaflets might hit the graft during opening and cusp abrasion might occur. The major potential drawback of the Yacoub technique, in our opinion, is that the ventriculoaortic junction (or aortic anulus), being not firmly fixed, can dilate postoperatively, causing late recurrent AR, especially in patients with Marfan syndrome.
1 Kunzelman's engineering group,
10 using elegant computer-modeling studies, recently postulated that the absence of pseudosinuses increases diastolic closing stresses on the abnormal stretched leaflets, which might predispose to premature cusp SVD. To get around this potential problem, the Cochran modification
11 of the original T. David-I procedure, DePaulis' modified aortic root replacement method,
12 or David's current T. David-V technique (with graft pleats placed at the anulus and at the tops of the commissures or sinotubular junction) all create billowing pseudosinuses while still retaining the advantages of valve reimplantation, including solid fixation of the ventriculoaortic junction.
References
This article has been cited by other articles:
![]() |
J. M. Albes, U. A. Stock, and M. Hartrumpf Restitution of the Aortic Valve: What is New, What is Proven, and What is Obsolete? Ann. Thorac. Surg., October 1, 2005; 80(4): 1540 - 1549. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. G. Bakaeen, J.-C. M. Walkes, and M. J Reardon Stentless Bioprosthetic Aortic Valve Replacement After a Homograft Root Replacement: Toronto SPV Implantation After a Homograft Root Ann. Thorac. Surg., April 1, 2005; 79(4): 1397 - 1399. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |