J Thorac Cardiovasc Surg 2003;125:378-384
© 2003 The American Association for Thoracic Surgery
Evolving Technology (ET) |
From the Departments of Anesthesiology and Divisions of Cardiothoracic Surgery, Mayo Clinic and Foundation, Rochester, Minn; Texas Heart Institute at St Luke's Episcopal Hospital, Houston, Tex; Wake Forest University School of Medicine, Winston-Salem, NC; and the Harvard Clinical Research Institute, Boston, Mass.
Received for publication March 15, 2002. Revisions requested May 24, 2002; revisions received June 26, 2002. Accepted for publication Aug 6, 2002. Address for reprints: David J. Cook, MD, Mayo Clinic, 200 First St SW, Rochester, MN 55905 (E-mail: cook.david{at}mayo.edu).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The ideal temperature for cardiopulmonary bypass (CPB) is a subject of debate. A body of evidence indicates that warm bypass can improve cardiac, pulmonary, and bleeding outcomes,
6-9 but concerns about warm bypass persist
10,11 because hypothermia is the most effective intervention for brain protection when ischemia occurs.
12,13
An optimal temperature-management strategy might combine cerebral hypothermia simultaneously with a warm body during CPB. A variety of interventions, including topical cooling and selective retrograde or antegrade perfusion, have been tried to provide more selective brain cooling during bypass.
14,15 However, these techniques have not been embraced because they can be technically cumbersome. More recently, dual-temperature control through a single aortic catheter has been described.
16,17
In large-animal bypass studies the Cobra catheter (Cardeon Corp, Cupertino, Calif) has been shown to allow independent control of arch and descending aortic temperature.
17 Additionally, it has been demonstrated to profoundly reduce brain and ocular embolization.
18 These 2 features, intended to improve patient outcomes by regionalizing perfusion, have been termed targeted circulatory management. The aim of this investigation was to determine whether the Cobra catheter delivers adequate perfusion while providing cerebral hypothermia and relatively normothermic systemic temperature in adults undergoing CPB.
| Methods |
|---|
|
|
|---|
After institutional review board approval and written informed consent, final eligibility was determined with the intraoperative determination of aortic dimensions and severity of atherosclerosis. Grade IV atherosclerosis
19,20 (severe; atheroma >5 mm in addition to
1 of the following: mobile or ulcerating lesions, extensive calcification, large protruding atheromatous debris or thrombus, or porcelain aorta) or an ascending aortic diameter of less than 19 mm or greater than 35 mm excluded patients intraoperatively before catheter placement.
The anesthetic management was according to the discretion of the attending anesthesiologist at each of the participating institutions. However, protocol required radial and femoral arterial lines, as well as a central venous pressure catheter or pulmonary artery catheter and transesophageal or epiaortic echocardiography for all patients. All patients had a No. 16 catheter placed retrograde into the jugular bulb through the internal jugular vein for intermittent sampling of cerebral venous oxygen saturation (SjVO2). Patients had body temperature measured at both the bladder (descending aortic circulation) and the nasopharynx (n = 18), jugular bulb (n = 19), or both (arch circulation).
The CPB circuit selection was determined by the participating institutions; however, the circuit modification required for dual-temperature control was standardized. The primary heat exchanger was, as usual, incorporated into the hollow-fiber oxygenator. Distal to the oxygenator, a bifurcation in the tubing allowed the warm oxygenated blood to be delivered either to the catheter port supplying the distal lumen (descending aorta) of the Cobra catheter (Figure 1) or to a second heat exchanger (BIOtherm; Medtronic, Minneapolis, Minn) used to cool the aortic arch perfusate. The hypothermic aortic arch perfusate was delivered to the proximal (arch) ports of the aortic catheter. An arterial filter (Affinity 351, Medtronic) was placed in both limbs of the aortic inflow line.
|
After sternotomy and anticoagulation, the Cobra catheter was placed through a standard purse-string suture in the ascending aorta. Positioning was confirmed echocardiographically, and the 2 ports on the catheter were connected to the respective
-inch (arch) and
-inch (descending aortic) inflow lines. Venous cannulation was then completed, and CPB was initiated.
When CPB was established, the Cobra baffle or wing was inflated with saline, and cooling of the arch perfusate was initiated. An efficacy end point was to achieve 3°C of arch cooling measured at the nasopharynx or jugular bulb before placement of the aortic crossclamp. After crossclamp placement, the target temperature for the arch circulation was 28°C to 31°C. For the descending aortic circulation, the target temperature was 35°C to 37°C. The goal was to establish a minimum of a 5°C temperature gradient between the 2 circulations.
Temperatures, flows, and mean arterial pressures in arch (radial arterial line) and descending aortic circulations (femoral arterial line) were measured continuously, while venous blood samples from the jugular bulb and bypass venous return line were drawn intermittently. The Cobra device used was rated for maximum flow rates of 2.75 L/min for the arch lumen and 3.25 L/min for the corporeal lumen, with a maximum pressure decrease of 200 mm Hg.
Patients were statistically characterized by using tables, figures, and descriptive statistics for demographic and baseline clinical variables. All statistical summaries and analyses were provided for the intent-to-treat population, which was defined as all consenting subjects who met the selection criteria. The one patient enrolled who did not undergo targeted circulatory management is included in the analysis of all data.
The analysis of variance technique was used to analyze the continuous variables, such as temperature, over time. A
2 test was used to analyze discrete variables, such as sex. Paired t tests were used to compare physiologic variables, such as venous oxygen saturation and temperature, in arch and descending aortic circulations.
Unless otherwise indicated, all comparisons of the treatment effect were performed with a 1-sided test at a less than .05 level of significance.
| Results |
|---|
|
|
|---|
In 19 of the 20 patients, the Cobra catheter was successfully placed, requiring no repositioning. In one instance, after initial placement, initiation of cooling resulted in decreasing bladder temperature and a stable nasopharyngeal temperature. This temperature pattern did not change after rotation of the catheter by the attending surgeon, and therefore bypass proceeded with the use of the Cobra device as a standard aortic cannula (a single temperature was delivered through both perfusion lumens). In the remaining patients the Cobra catheter allowed for the rapid establishment of cerebral hypothermia relative to systemic normothermia.
Before bypass, mean arch (n = 20) and bladder (n = 19) temperatures were 35.9°C ± 0.5°C and 36.1°C ± 0.3°C, respectively. With the onset of wing inflation and cooling, the arch perfusate temperature was initially set at a mean of 24.9°C, with the descending aortic perfusate temperature set at a mean of 35.3°C. The crossclamp was applied a mean of 3.4 minutes later, by which time a mean temperature differential between bladder and arch circulation of 4.3°C was established. The mean temperatures at this time were 30.5°C ± 2.1°C at the nasopharynx or jugular bulb and 34.8°C ± 1.7°C at the bladder.
In the 19 patients undergoing differential temperature management, the targeted maximum differential was 6.7°C ± 0.8°C. A mean maximum temperature gradient of 7°C ± 1.2°C was established at 22.0 ± 18 minutes of cooling time (Figure 2). At that time, mean arch and bladder temperatures were 28.2°C ± 2.2°C and 34.6°C ± 1.7°C, respectively. Gradual rewarming of the arch circulation was initiated at the discretion of the surgeon, such that a modest temperature differential and mild cerebral hypothermia remained at the time of crossclamp removal. At crossclamp removal, temperatures in the arch and descending aortic circulations were 32.9°C ± 1.7°C and 35.3°C ± 0.8°C, respectively. At that time, active rewarming of the arch circulation was stopped. During bypass with the Cobra device, all measured arch temperatures were lower than bladder temperatures, with a P value of less than .000 by using the paired t test. After weaning from bypass, the mean arch and bladder temperatures were 36.1°C ± 1.2°C and 36.5°C ± 0.8°C (Figure 2
).
|
The adequacy of cerebral and systemic perfusion was assessed on the basis of measurements of mixed systemic venous oxygen saturation (SVO2) and SjVO2. Figure 3 shows the values for SVO2 and SjVO2 in the prebypass period during stable targeted circulatory management with the Cobra device and in the postbypass period. During stable CPB, the SVO2 and SjVO2 ranged from 71% to 87% and 68% to 98%, respectively. Before and after bypass, the SjVO2 was lower than the SVO2 (P = .004 and .003, respectively, paired t test), whereas during bypass, the 2 saturations did not differ.
|
| Discussion |
|---|
|
|
|---|
The Cobra device is unique in several ways. It is a dual-lumen aortic catheter intended as a substitute for a traditional aortic cannula and is capable of establishing and maintaining cerebral hypothermia with simultaneous warm bypass to the rest of the body.
17 By means of a second heat exchanger, we show that it is possible to establish greater than 3°C of cerebral hypothermia before crossclamp application and maintain cerebral hypothermia (32.9°C ± 1.7°C) late in bypass at the time of crossclamp removal. This is important because aortic manipulation with the crossclamp is probably the most important cause of brain embolization in cardiac surgery.
21-25
In addition to providing brain cooling at the time of crossclamp placement and removal, a greater degrees of cerebral hypothermia can be achieved and maintained through the bulk of the surgical procedure when using the Cobra catheter. In this study of 20 patients at 3 institutions, we achieved a maximum mean gradient of 7°C with a mean brain temperature of 28.2°C when the mean bladder temperature was 34.6°C. Greater degrees of cerebral hypothermia could be established in cases of longer duration without the requirement to cool the body to the same degree. Although not the design of this study, a brain temperature of 17°C to 20°C and a body temperature of 25°C to 27°C might be practical for such cases, reducing cooling, rewarming, and bypass time and potentially reducing the bleeding associated with profound whole-body hypothermia.
The other unique feature related to temperature management with the Cobra device is elimination of most of the need for rewarming. When this catheter is used, the bulk of body mass stays relatively normothermic. The requirement to replace the thermal debt is minimal, and therefore little energy needs to be applied to reach target temperature before weaning. This eliminates the overheating of the brain and cerebral oxygenation stress that can frequently occur when rewarming the body from hypothermic bypass.
26,27 With the Cobra device, a low level of selective cerebral cooling can be maintained through the last part of bypass, or the temperature of the arch circulation can be gradually equilibrated with body temperature. Because high brain temperatures and low oxygen saturations during late bypass are associated with worsened neurologic outcome,
28 managing temperature in this fashion probably has cerebral advantages.
Although dual-temperature control expands what is possible during bypass and would be predicted to offer both neurologic and cardiac advantages, the Cobra catheter has a more unique feature that might prove more valuable than dual-temperature bypass.
A novel design feature of the Cobra catheter is a deployable wing that facilitates aortic arch flow segmentation and dual-temperature management. However, a more unique feature of this wing design is its ability to protect the cerebral circulation from embolization originating in the heart or aortic root.
18 The catheter has a perfusion lumen on the cephalad surface of the cannula, as well as a distal port that perfuses the descending aorta. The catheter is nonocclusive proximally, and the intrinsic resistance of the device results in a flow to the arch lumen that is high relative to the body surface area perfused. This results in continuous overflow around the proximal edge of the wing. This proximal overflow continually irrigates the aortic root and directs the blood in the root around the lesser curvature of the arch and down the descending aorta. In large-animal bypass studies the Cobra wing reduced embolization to the brain by approximately 90%.
18 This dramatic result was achieved in the absence of any cerebral hypothermia. If cerebral cooling was added, we would predict that the reduction in brain embolization with the Cobra device would be even greater.
The Cobra device is easy to use and is equivalent in function to traditional arterial cannula with or without the use of the wing. The wing is prepared and deployed by using standard techniques (ie, fluid inflation with syringe). The catheter is placed through a standard ascending aortic aortotomy and is sized and shaped so that it sits in the middle of the flow stream in the transverse aortic arch. Its positioning is easily confirmed by imaging the distal portion of the cannula with transesophageal echocardiography. Perfusion with the device at conventional flow rates (2-2.4 L · min-1 · -2) results in mean arterial pressures above the autoregulatory threshold. The adequacy of cerebral perfusion and systemic perfusion is evidenced in this clinical trial by SjVO2 and SVO2 values equal to or greater than those seen under nonbypass conditions (Figure 3
).
In one patient in this trial, use of the catheter resulted in temperature changes opposite those expected. Body temperature cooled, while the nasopharyngeal temperature remained warm. Rotation of the device did not change this situation, and therefore the surgeon elected to use the catheter as a conventional device. In retrospect, it is not possible to definitively determine the cause of the difficulty in this early case experience. It is possible that the catheter was rotated on insertion, such that the arch ports faced the lesser curvature of the aorta; however, rotation did not change the temperature control. It is more likely, however, that either the water lines to the arch and body heat exchangers or the connections for the 2 temperature probes were reversed. It should also be noted that this case was the first use of the Cobra device at that hospital. In greater than 130 subsequent clinical placements of the Cobra catheter at 15 institutions, this experience has not been repeated.
An experimental large-animal study has indicated that the Cobra device can dramatically reduce brain embolization by shunting those emboli down the descending aorta.
18 It must therefore be asked whether use of this device would adversely affect organs supplied by the descending aorta. Although this trial is not designed to answer that question, we would speculate that this effect should be quite small. Both experimental
18 and clinical
22 studies found that the mean percentage of the aortic embolic load entering the cerebral circulation was approximately 7% to 8%.
18,22 As such, redirecting 90% of that 8% would have a small effect on the total embolization down the descending aorta. Furthermore, that additional fraction of emboli would be distributed to a much larger mass of body, and therefore the concentration effect of the redirected emboli should be very small. Additionally, those redirected emboli would be delivered to organs with much lower metabolic demands and better collateralization than the brain. Nevertheless, the randomized clinical outcome trial for the Cobra catheter will need to assess renal and visceral organ outcomes, as well as its neurologic effect.
In this investigation we demonstrate that the Cobra device meets the standards of adequacy of perfusion during bypass, as well as its ability to provide targeted dual-temperature control. Although this catheter is designed to provide neuroprotection by means of cerebral hypothermia and emboli reduction, it was not the purpose of this investigation to determine whether this device could improve neurologic outcome. Although 19 of the 20 patients in this trial underwent National Institutes of Health Stroke Scale assessment within 36 hours of operation and no patient demonstrated stroke or focal neurologic event, the study is not powered to provide neurologic outcomes. In contrast to this small prospective study, a neuroprotection trial must be randomized and appropriately powered for low-incidence events. A randomized, prospective, multicenter trial examining neurologic, cognitive, cardiac, pulmonary, and renal outcomes with the Cobra catheter has been initiated.
| Appendix 1 |
|---|
|
|
|---|
50 U/mL, bilirubin level > 1.5 mg/dL, or increased prothrombin time of > 5 seconds above control values), bleeding disorder, or coagulopathy | Appendix 2 |
|---|
|
|
|---|
|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. J. Cook, J. Huston III, M. R. Trenerry, R. D. Brown Jr, K. J. Zehr, and T. M. Sundt III Postcardiac Surgical Cognitive Impairment in the Aged Using Diffusion-Weighted Magnetic Resonance Imaging Ann. Thorac. Surg., April 1, 2007; 83(4): 1389 - 1395. [Abstract] [Full Text] [PDF] |
||||
![]() |
T.M. Sundt, K.J. Zehr, J.A. Dearani, R.C. Daly, C.J. Mullany, C.G.A. McGregor, F.J. Puga, T.A. Orszulak, and H.V. Schaff Is early anticoagulation with warfarin necessary after bioprosthetic aortic valve replacement? J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 1024 - 1031. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kaukuntla, A. Walker, D. Harrington, T. Jones, and R. S. Bonser Differential brain and body temperature during cardiopulmonary bypass--a randomised clinical study Eur. J. Cardiothorac. Surg., September 1, 2004; 26(3): 571 - 579. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||