|
|
||||||||
J Thorac Cardiovasc Surg 2003;125:1043-1049
© 2003 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
From the Department Cardiothoracic Surgery and Anesthesiology, Huddinge University Hospital,a and the Division of Medical Engineering, Department of Medical Laboratory Science and Technology, Karolinska Institute,b Stockholm, Sweden.
Supported by Karolinska Institute and Cardia Innovation AB, Stockholm, Sweden.
Received for publication June 14, 2002. Accepted for publication Aug 6, 2002. Address for reprints: Jan van der Linden, MD, Department of Cardiothoracic Surgery and Anesthesiology, M85, Karolinska Institute, Huddinge University Hospital, SE-141 86 Stockholm, Sweden (E-mail: jan.vanderlinden{at}thsurg.hs.sll.se).
| Abstract |
|---|
|
|
|---|
1.0%) at both carbon dioxide flows. With suction of 10 L/min the median air content was still low (
0.50%) at a simultaneous carbon dioxide flow of 10 L/min. Conversely, suction of 25 L/min caused a marked increase in air content at carbon dioxide flows of both 5 and 10 L/min (P < .001).
1% remaining air) in a cardiothoracic wound model was provided by a gas diffuser at a carbon dioxide flow of 10 L/min. A conventional 0.25-inch tube failed to do so (19.5%-51.7% remaining air). Additional suction deteriorated air displacement with the gas diffuser when suction exceeded carbon dioxide inflow. | Introduction |
|---|
|
|
|---|
|
Several important factors influence carbon dioxide deairing during cardiac surgery, such as type of insufflation device, carbon dioxide flow, and coronary and rough suction. These factors were controlled in this study.
The first aim of this study was to assess a new carbon dioxide insufflation device at different carbon dioxide flows for air displacement in a cardiothoracic wound cavity model, where a conventional open-ended tube served as a control. A second aim was to study the influence of suction on air displacement in the same model.
| Materials and methods |
|---|
|
|
|---|
Instrumentation
The carbon dioxide flow was measured with a back pressure-compensated oxygen flowmeter, because a flowmeter for medical carbon dioxide was unavailable at the time of the study. The oxygen reading scale was adjusted for carbon dioxide with a universal flowmeter (ABB/Fisher & Porter, Göttingen, Germany) because of the higher density of carbon dioxide gas. The universal flowmeter consisted of a measuring tube (FP -16 G-5/81) with a spherical stainless steel float (SS-14). The universal flowmeter was not used for measurements in the study because of its lack of back-pressure compensation. This problem was avoided during the calibration by measuring the carbon dioxide outflow distal to the end of the insufflation device. The reading scale of the universal flowmeter was calculated for the used gas (medical carbon dioxide; AGA Gas AB, Stockholm, Sweden) at 20°C and at 1013 mbar with a computer program (FlowSelect version 2.0; ABB/Fisher & Porter).
Coronary suction, which is usually set at an effect of 1 to 1.5 L/min, was set at 1.5 L/min and was provided by a standard roller pump and calibrated according to the manual of the manufacturer (CAPS; Stöckert Instrumente GmbH, Munich, Germany). The rough suction was set at 10 and 25 L/min (maximum) and was controlled by two flowmeters with regulators coupled in parallel. These flowmeters were also calibrated with the universal flowmeter.
The degree of air displacement in the wound cavity model was assessed by analyzing the remaining air content (%Air), which is given by the following equation:
where %O2 is the measured oxygen concentration and %O2(ref) is the oxygen concentration in atmospheric air near sea level (20.95%).
15 The oxygen content was measured with an oxygen sensor (CheckMate 9900; PBI-Dansensor A/S, Ringsted, Denmark), which has a gas sampling volume of less than 2 mL, a response time of <2 seconds (>20.95% change in oxygen concentration in both directions), a range of measurement of 0.0001% to 100% oxygen, and an accuracy of 1% of the measured value. The sampling probe was a 1.5 mm thick polytetrafluoroethylene tube. The oxygen instrument was connected to a personal computer for recording of data.
Setup
The degree of air displacement was studied in an anatomic torso model with an open cardiothoracic wound containing a silicone replica of the heart and great vessels. The shape of the model was based on the maximal measurements of the open chest wounds of five adults undergoing cardiac surgery (standard sternotomy and during cardiopulmonary bypass with empty heart). We presupposed that a wound cavity with a large opening would be more difficult to deair because of increased diffusion. The torso was placed on the operating table of an operating theater normally ventilated for cardiac surgery (downward laminar airflow from the ceiling above the operating table, approximately 2500 m3/h). The wound opening was 20 cm long (midline) and 12 cm wide. The volume of the wound cavity without the artificial heart was 2.5 L. The external volume of the artificial heart, including the great vessels, was 1.0 L giving a residual cavity volume of 1.5 L. The orifices of the insufflation devices were positioned 5 cm below the wound opening adjacent to the diaphragm. The tube was pointed toward the center of the wound cavity and not toward the site of measurements. Carbon dioxide was insufflated into the wound cavity at flows of 5 and 10 L/min. The remaining air content was measured at the highest part of the right atrium, 5 cm below the wound opening, and at the highest part of the ascending aorta, 3 cm below the wound opening. These positions are close to the sites of the atrial and aortic incisions in valve surgery.
Measurements
First, the air displacement efficiencies of the two insufflation devices were assessed without suction. A stable oxygen concentration was considered to be present when values were fluctuating around a constant value through a period of 30 seconds. Thereafter the oxygen concentration was recorded 10 times in succession, once every 5 seconds (n = 10). The efficiency of the gas diffuser was further studied with the addition of the varying degrees of continuous suction (1.5, 10, and 25 L/min) at the site of the artificial left atrial appendage. When a stable oxygen concentration, as defined previously, was present, suction was applied and the oxygen concentration was recorded once every 5 seconds during 60 seconds. Each recording procedure with suction was repeated 10 times (n = 10). At all measurements the remaining carbon dioxide in the model was removed with the rough sucker before every change of carbon dioxide flow or insufflation device. Thereafter air movements around the model were left to settle for 1 minute.
Statistics
This was an analysis of variance design, but because of unsuitable distribution characteristics a more simple and conservative nonparametric analysis was chosen. Mann-Whitney U tests and Wilcoxon tests were used whenever appropriate. Data in the diagrams are presented as median and range.
| Results |
|---|
|
|
|---|
|
0.24% at 10 L/min carbon dioxide and
1.0% at 5 L/min carbon dioxide) both at the right atrium and at the ascending aorta when a suction of 1.5 L/min was applied. With a carbon dioxide flow of 10 L/min the air content was lower than in the case when no suction was used both at the right atrium (P = .001) and at the ascending aorta (P < .001). With a suction of 10 L/min the median air content was still low both at the right atrium (0.37%) and at the ascending aorta (0.50%) with a simultaneous carbon dioxide flow of 10 L/min but much higher (>8.9%, P < .001) at both sites with a carbon dioxide flow of 5 L/min. A suction of 25 L/min increased the median air content at the right atrium to 9.9% (P < .001) at a carbon dioxide flow of 10 L/min and to 31.4% (P < .001) at a flow of 5 L/min. The corresponding values at the ascending aorta were 41.1% (P < .001) and 75.1% (P < .001). With suctions of 10 and 25 L/min the air content at the right atrium was lower than at the ascending aorta at both carbon dioxide flows (P = .001).
|
0.46%) both at the right atrium and at the ascending aorta during the measurements. After the start of suction with 25 L/min, the air content remained unchanged during 5 seconds at both sites, followed by an increase after 10 seconds. Stable air contents were reached after 15 and 20 seconds at the right atrium and at the ascending aorta, respectively, with a higher air content at the ascending aorta than at the right atrium (P < .001).
|
| Discussion |
|---|
|
|
|---|
Measuring the content of carbon dioxide or air
11,16 in a cardiothoracic wound cavity may assess the degree of air displacement. In this study the air content was estimated by measuring the oxygen concentration, which is 20.95% in atmospheric air near sea level.
15 The instrument that was used, an oxygen sensor with a heated ceramic element, can assess air displacement more accurately and faster than can commonly used carbon dioxide sensors that use an optical infrared sensor technique to measure 0% to 100% carbon dioxide. The oxygen sensor's accuracy was 1% of the measured value in the range 0.0001% to 100% oxygen, which means that the accuracy increases as the oxygen and the air contents decrease. Moreover, the oxygen sensor requires only a 2-mL gas sample volume and has a response time of less than 2 seconds. The smaller the gas sample needed, the shorter is the response time and the less will gas measurements interfere with carbon dioxide deairing. In contrast, carbon dioxide sensors that use the infrared technique usually have a constant accuracy of approximately ±2% carbon dioxide across the entire range of measurement of 0% to 100% carbon dioxide, a larger required sampling volume, and a longer response time, usually longer than 10 seconds. The fast response modality of the oxygen sensor used enabled us to detect rapid variations of air content with time. Thus we consider the oxygen sensor that we used more suitable for evaluation of low air contents during steady state and during changes than optical infrared carbon dioxide sensors would be.
Because only carbon dioxide flows of 10 L/min or less have so far been reported,
3,11,14,16-20 we did not study carbon dioxide flows higher than 10 L/min. Moreover at 5 L/min almost complete air displacement was already obtained when the gas diffuser was used.
Carbon dioxide insufflation devices
We studied an open-ended tube with an inner diameter of 0.25 inches because tubes of the same size have been used for carbon dioxide insufflation in our clinic as well as in earlier studies.
11,16,20 In this study the air displacement efficiency of the 0.25-inch tube was compared with a new device, a gas diffuser. The study revealed a striking difference in efficiency between the conventional 0.25-inch open-ended tube and the gas diffuser at the studied carbon dioxide flows (Figure 1
). Carbon dioxide insufflation with the 0.25-inch tube resulted in median air contents between 19.5% and 28.3% and between 48.2% and 51.7% at carbon dioxide flows of 5 and 10 L/min, respectively. The insufficient air displacement with the 0.25-inch tube worsened with increased carbon dioxide flow. The 0.25-inch tube's apparent failure to displace air with carbon dioxide was probably due to the turbulence induced by the carbon dioxide jet. The same phenomenon occurs when one tries to fill a pail with water by means of a garden hose. Most of the water splashes out of the pail. By contrast, the pail is quickly filled if the hose is provided with a multi-perforated nozzle, resulting in a reduced flow velocity. With the gas diffuser the median air content was no greater than 1.2% at 5 L/min and no greater than 0.31% at 10 L/min, which indicates minimal turbulence despite a high carbon dioxide flow.
Suction and carbon dioxide flow
The influence of suction has to be considered when aiming at efficient deairing with carbon dioxide insufflation in cardiac surgery. Coronary suction of 1.5 L/min did not deteriorate air displacement in the wound cavity. In vivo, however, continuous suction of almost pure carbon dioxide into the cardiotomy reservoir may affect arterial pCO2.
17,19,20 This may be avoided by minimizing continuous suction of carbon dioxide and by increasing the ventilation of the oxygenator. A rough suction of 25 L/min applied in the wound cavity at carbon dioxide flows of 5 and 10 L/min markedly increased the air content. This was to be expected because the suction markedly exceeded the carbon dioxide inflow. Surprisingly, the air content did not rise toward 100% but stabilized at lower levels. This may be explained by a partial suction of air from above the cavity. In addition, carbon dioxide disperses effectively from the diffuser in multiple directions in the wound cavity and not only toward the suction. The increase of air content to a stable level started first after 5 seconds and before 10 seconds of suction (not adjusted for the response time of the instrument, which is <2 seconds). This delay is most likely explained by the buffering effect of the volume of carbon dioxide within the cavity. A smaller wound cavity will hold a smaller volume of carbon dioxide, which may make it more difficult to keep a low air content when suction is applied. When the rough suction was reduced to equal the carbon dioxide inflow at 10 L/min, the median air content could be kept at no higher than 0.5% in the wound cavity.
These findings suggest that the duration of active use of rough suction at 25 L/min is important if it is higher than the carbon dioxide flow. A short period of suction, up to approximately 5 seconds, does not affect the air content, whereas a longer period of suction creates an increase in the air content, which soon stabilizes. Thus rough suction should be kept as short as possible when used actively inside the wound cavity, provided that it is not continuously sucking fluid. Otherwise the suction should either be shut off or be kept outside the wound cavity.
11 Another option is to set the rough suction equal to or lower than the carbon dioxide inflow. Theoretically the carbon dioxide flow could also be increased to 25 L/min, but this may not be necessary, because as a rule rough suction is only used infrequently and during short periods.
When and where to insufflate carbon dioxide
The wound cavity should be insufflated with carbon dioxide when there is a risk of air entering the circulatory system. If air enters the heart and great vessels, air may be trapped
21 and sooner or later embolize.
8,9 It is important that carbon dioxide is insufflated from incision to closure of the heart and great vessels so that these potential air pockets are filled with carbon dioxide. This is relevant not only for heart and aortic surgery but also when the single-clamp technique is used in conventional coronary artery bypass surgery.
The gas diffuser is preferably placed where it is efficient without interfering with surgery. In this study the diffuser was positioned 5 cm below the opening, adjacent to the diaphragm. During surgery there is little activity at this site, and few if any surgical instruments are present. At this position the gas diffuser provided a high degree of air displacement both at the site of the right atrium and at the ascending aorta. When suction exceeds carbon dioxide inflow, however, the proximity of the diffuser to the location of interest (measuring point) becomes important for air displacement (Figure 2
and 3
).
Conclusions
This study demonstrated that at a carbon dioxide flow of 10 L/min the gas diffuser provided efficient air displacement (
1% remaining air) in a cardiothoracic wound cavity model. By contrast, a conventional 0.25-inch tube failed to do so (19.5%-51.7% remaining air). Coronary suction did not influence carbon dioxide deairing. Rough suction deteriorated air displacement with the gas diffuser when suction exceeded carbon dioxide inflow, only then making the proximity of the gas diffuser to the location of interest important.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. Martens, K. Neumann, C. Sodemann, H. Deschka, G. Wimmer-Greinecker, and A. Moritz Carbon Dioxide Field Flooding Reduces Neurologic Impairment After Open Heart Surgery Ann. Thorac. Surg., February 1, 2008; 85(2): 543 - 547. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Persson and J. van der Linden Can Wound Desiccation Be Averted During Cardiac Surgery? An Experimental Study Anesth. Analg., February 1, 2005; 100(2): 315 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Svenarud, M. Persson, and J. van der Linden Effect of CO2 Insufflation on the Number and Behavior of Air Microemboli in Open-Heart Surgery: A Randomized Clinical Trial Circulation, March 9, 2004; 109(9): 1127 - 1132. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |