J Thorac Cardiovasc Surg 2004;127:1087-1092
© 2004 The American Association for Thoracic Surgery
General thoracic surgery |
a Department of Thoracic Surgery, Saiseikai Central Hospital, Tokyo, Japan
b Nishidai Clinic, Tokyo, Japan
Received for publication June 13, 2003; revisions received July 14, 2003; revisions received July 23, 2003; accepted for publication August 11, 2003.
* Address for reprints: Hiroaki Nomori, MD, Department of Thoracic Surgery, Saiseikai Central Hospital, 1-4-17 Mita, Minato-ku, Tokyo 108-0073, Japan
hnomori{at}qk9.so-net.ne.jp
| Abstract |
|---|
|
|
|---|
METHODS: Preoperative positron emission tomography and computed tomography scans were performed for 564 lymph node stations in 80 patients with peripheral-type lung cancer. The sizes of both the lymph nodes and the metastatic foci within the lymph nodes were measured, and these measurements were compared with those obtained with positron emission tomography scanning. To establish general sizes of metastatic foci within the lymph nodes, 277 metastatic lymph nodes in operative specimens previously resected from another 111 patients with lung cancer were examined as a control.
RESULTS: The sensitivity was significantly higher for positron emission tomography than for computed tomographic scanning (P = .026). The sizes of metastatic foci within lymph nodes that showed false-negative (n = 8) and true-positive (n = 28) with positron emission tomography ranged from 0.5 to 9 mm (3 ± 1 mm) and from 4 to 18 mm (10 ± 3 mm), respectively (P < .001). None of the metastatic foci smaller than 4 mm could be detected with positron emission tomography scanning. The review of the 277 previously resected metastatic lymph nodes showed that 89 (32%) had metastatic foci smaller than 4 mm. The sizes of true-positive (n = 28) and false-positive (n = 10) lymph nodes ranged from 6 to 15 mm (10 ± 2 mm) and from 9 to 16 mm (12 ± 2 mm), respectively (P < .01). None of the false-positive lymph nodes was smaller than 9 mm.
CONCLUSIONS: Although positron emission tomography was superior to computed tomography scanning in lymph node staging in lung cancer, positron emission tomography was unable to distinguish metastatic foci smaller than 4 mm, which were not unusual sizes for lymph node metastases in lung cancer. Positive lymph nodes with positron emission tomography smaller than 9 mm are likely to be true-positive rather than false-positive.
|
Several diagnostic procedures are used for preoperative staging of patients with lung cancer, including computed tomography (CT), magnetic resonance imaging, positron emission tomography (PET), mediastinoscopy, and thoracoscopy. Of the noninvasive procedures, CT remains the standard procedure for N staging in lung cancer. However, CT scanning is not sufficiently sensitive or specific for diagnosing lymph node metastasis1; because size is the only criterion used to differentiate benign from malignant lymph nodes, lymph nodes 1 cm or larger are considered abnormal.
In recent years, fluorodeoxyglucose (FDG)-PET scanning has been used for the staging of lung cancer.2-7 Because of the biological nature of FDG, FDG-PET has been reported to detect metastatic lymph nodes smaller than 1 cm. A meta-analysis by Dwamena and colleagues1 of PET scanning of 514 patients in 14 studies showed that the mean sensitivity and specificity for PET scanning in N staging were 0.79 (range, 0.62 to 0.97) and 0.91 (range, 0.79 to 0.99), respectively, both of which were superior to those for CT scanning, ie, 0.60 (range, 0.25 to 0.89) and 0.77 (range, 0.44 to 0.95), respectively. The detection of small metastatic lymph nodes is, however, limited by the spatial resolution of the PET scanner. In this study, to determine the lower size limit of the metastatic lymph nodes that can be detected by PET scanning, we measured the sizes of metastatic foci in lymph nodes with true-positive (TP) and false-negative (FN) results, and these results were compared with the sizes of metastatic lymph nodes from another large series of lung cancer patients to determine the general frequency of sizes of metastatic lymph nodes. We also measured the sizes of TP and false-positive (FP) lymph nodes to determine the lymph node sizes that gave FP results on PET scan.
| Materials and methods |
|---|
|
|
|---|
|
N staging by PET scanning
PET data were evaluated semiquantitatively on the basis of the contrast ratio (CR), which was obtained as follows. The regions of interest (ROIs) were placed in the lymph nodes and cerebellum. The highest activities in both the lymph node ROI (L) and the cerebellum ROI (C) were measured. The CR was calculated by L/C in each lymph node as an index of FDG uptake. Receiver operating characteristic curves9 were constructed according to the CR value and pathologic diagnosis, and the cutoff value was determined for diagnosis of metastasis.
N staging by CT scanning
Spiral CT was performed by using the ProSeed SA (General Electric Medical Systems, Milwaukee, Wis). The following acquisition parameters were used: high voltage (120 kV), tube load of 160 mA, window level of 500 Hounsfield units (HU), and window width of 1500 HU. The entire thorax was scanned with 1-cm-thick sections at 1 breath hold with maximum inspiration. The criterion of CT definition for suspected metastasis of the lymph node was a short-axis diameter of 1.0 cm or larger.
Measuring the size of lymph nodes
In each lymph node station that was TP or FP with PET scan, the short-axis diameters of the largest lymph nodes were measured on CT scan.
Measuring the size of metastasis
The dissected lymph nodes were examined histologically by using formalin-fixed and paraffin-embedded sections with hematoxylin and eosin staining. Each lymph node was examined with 1 hematoxylin and eosinstained section. In each lymph node station containing metastases, the lymph node with the largest metastatic lesions was selected, and the long-axis diameters of the metastatic foci in the lymph nodes were measured. The sizes of metastatic foci were measured with a microscope.
Control subjects
Another 277 metastatic lymph nodes from 111 patients with lung cancer who had undergone operation with mediastinal lymph node dissection from 1985 to 2001 were also examined as a control in the same way: that is, (1) the lymph nodes with the largest metastatic lesions in each lymph node station were selected, and (2) the long-axis diameters of the metastatic foci were measured. The sizes were classified in 1-mm graduations from the smaller-than-1-mm class to the 10-mm-or-larger class. The histologic types were adenocarcinoma in 72 patients, squamous cell carcinoma in 26, large cell carcinoma in 10, and adenosquamous carcinoma in 3 (Table 2). The pathologic classification was N1 in 36 patients, N2 in 69, and N3 in 6.
|
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
This study showed that the PET scan could not detect lymph nodes that had metastatic foci smaller than 4 mm. Because there was no significant difference of frequency of single-node metastasis between the TP and FN lymph node stations, the sensitivity of PET for N staging could be dependent on the size of metastatic foci rather than on the number of metastatic lymph nodes in each station. The FDG uptake of small lesions may have been underestimated as a result of the spatial resolution of the PET scanner, eg, because of the partial-volume effect. Although Gupta and colleagues3 reported that the sensitivity, specificity, and accuracy of PET scanning for detecting lymph node metastasis showed no difference between lymph nodes smaller than 1 cm, 1 to 3 cm, or larger than 3 cm, they did not define the lower size limit for PET scanning. To examine the general distribution of general sizes of metastatic foci within the lymph nodes, we measured the sizes of 277 metastatic lymph nodes in a large series of lung cancer patients. If 4 mm were the lower limit for PET scanning to detect metastatic foci, 89 (32%) of lymph nodes with metastases smaller than 4 mm would yield FN results with PET scanning. Even if 3 mm were the lower limit, 59 (21%) of lymph nodes with metastases smaller than 3 mm would yield FN results (Table 8). Considering the spatial resolution of the PET scanner, artifacts resulting from respiratory movement, image reconstruction, and biological effects, it is difficult to evaluate such small metastatic foci. Current PET scanners, similar to the POSICAM used in this study, achieve transaxial resolutions of 5 to 6 mm at full width, half maximum. Tumors smaller than 10 mm could therefore not be resolved even by the current scanner because of the partial-volume effect. It has in fact been reported that the current generation of PET scanners are unable to delineate lung cancers smaller than 10 mm.17 This study did show, however, that all metastatic foci 5 mm or larger, except 1 with the lesion 9 mm in size, were detected with PET scanning, apparently exceeding the lower limit of spatial resolution. We consider that metastatic foci smaller than 10 mm can be detected with PET scanning for the following reasons: (1) metastatic tumor cells would have more biological activity than primary tumor cells, causing higher FDG uptake in metastatic foci than in the primary sites, and (2) the immune reaction of macrophages within metastatic lymph nodes might increase FDG uptake, because FDG has been reported to accumulate not only in tumor cells, but also in macrophages.18
In this study, we excluded central-type lung cancer patients for the following reasons: (1) it is difficult for PET scanning to distinguish the primary tumor from adjacent nodal metastasis, and (2) bronchial obstruction by a central tumor can lead to inflammation in the locoregional lymph nodes, with subsequent FDG uptake. Nevertheless, 6 (75%) of 8 FN lymph node stations with PET scanning were located in the hilar region. Although there was no significant difference of accuracy for overall N staging between PET and CT scanning, PET could identify N2 disease more accurately than CT. Vesselle and associates19 also reported that, although PET scanning could correctly differentiate N0 or N1 from N2 or N3 disease, it could not reliably identify N1 disease, with only 6 of 21 cases identified. We therefore believe that PET scanning is useful for determining an indication of operation or neoadjuvant therapy for N2 disease but that it is not reliable for N1 disease.
It has been reported that FP results for lymph nodes can occur with PET scanning because of concurrent inflammatory conditions or centrally located lung cancer.20 Although this study excluded centrally located lung cancers, it is well known that lymph nodes near the primary tumor could become enlarged with an immune reaction, which could become positive with PET scanning because of FDG uptake by reactive macrophages, resulting in an FP result. This study showed that the FP lymph nodes with PET scanning were significantly larger than the TP ones and that none of the FP lymph nodes was smaller than 9 mm. Therefore, we consider that positive lymph nodes with PET scanning smaller than 9 mm are more likely to be TP than FP.
Although PET scanning is superior to CT scanning for N staging in lung cancer, it was unable to detect metastatic foci smaller than 4 mm, which were not unusual sizes for metastatic lymph nodes in lung cancer. Positive lymph nodes with PET scanning that are smaller than 9 mm are more likely to be TP than FP.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. Nomori, T. Mori, K. Ikeda, K. Kawanaka, S. Shiraishi, K. Katahira, and Y. Yamashita Diffusion-weighted magnetic resonance imaging can be used in place of positron emission tomography for N staging of non-small cell lung cancer with fewer false-positive results. J. Thorac. Cardiovasc. Surg., April 1, 2008; 135(4): 816 - 822. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. D. Kozower, B. F. Meyers, C. E. Reed, D. R. Jones, P. A. Decker, and J. B. Putnam Jr Does Positron Emission Tomography Prevent Nontherapeutic Pulmonary Resections for Clinical Stage IA Lung Cancer? Ann. Thorac. Surg., April 1, 2008; 85(4): 1166 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. C. Ung, D. E. Maziak, J. A. Vanderveen, C. A. Smith, K. Gulenchyn, C. Lacchetti, W. K. Evans, and Lung Cancer Disease Site Group of Cancer Care Onta 18Fluorodeoxyglucose Positron Emission Tomography in the Diagnosis and Staging of Lung Cancer: A Systematic Review J Natl Cancer Inst, December 5, 2007; 99(23): 1753 - 1767. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Silvestri, M. K. Gould, M. L. Margolis, L. T. Tanoue, D. McCrory, E. Toloza, and F. Detterbeck Noninvasive Staging of Non-small Cell Lung Cancer: ACCP Evidenced-Based Clinical Practice Guidelines (2nd Edition) Chest, September 1, 2007; 132(3_suppl): 178S - 201S. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Mujoomdar, J. H. M. Austin, R. Malhotra, C. A. Powell, G. D. N. Pearson, M. C. Shiau, and H. Raftopoulos Clinical Predictors of Metastatic Disease to the Brain from Non-Small Cell Lung Carcinoma: Primary Tumor Size, Cell Type, and Lymph Node Metastases Radiology, March 1, 2007; 242(3): 882 - 888. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Massard, E. A. Kochetkova, V. A. Porhanov, and M. Riquet Reply to Belcher and Goldstraw Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 143 - 144. [Full Text] [PDF] |
||||
![]() |
A. Ebihara, H. Nomori, K. Watanabe, T. Ohtsuka, T. Naruke, K. Uno, I. Kuwahira, and K. Eguchi Characteristics of Advantages of Positron Emission Tomography over Computed Tomography for N-staging in Lung Cancer Patients Jpn. J. Clin. Oncol., November 1, 2006; 36(11): 694 - 698. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Massard, X. Ducrocq, E. A. Kochetkova, V. A. Porhanov, and M. Riquet Sampling or node dissection for intraoperative staging of lung cancer: a multicentric cross-sectional study. Eur. J. Cardiothorac. Surg., July 1, 2006; 30(1): 164 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.-i. Watanabe, H. Nomori, T. Ohtsuka, T. Naruke, A. Ebihara, H. Orikasa, K. Yamazaki, K. Uno, T. Kobayashi, and T. Goya [F-18]Fluorodeoxyglucose Positron Emission Tomography Can Predict Pathological Tumor Stage and Proliferative Activity Determined by Ki-67 in Clinical Stage IA Lung Adenocarcinomas Jpn. J. Clin. Oncol., July 1, 2006; 36(7): 403 - 409. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bunyaviroch and R. E. Coleman PET Evaluation of Lung Cancer J. Nucl. Med., March 1, 2006; 47(3): 451 - 469. [Full Text] [PDF] |
||||
![]() |
J. L. Port, R. S. Andrade, M. A. Levin, R. J. Korst, P. C. Lee, D. E. Becker, and N. K. Altorki Positron emission tomographic scanning in the diagnosis and staging of non-small cell lung cancer 2 cm in size or less J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1611 - 1615. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Ohtsuka, H. Nomori, K. Watanabe, T. Naruke, H. Orikasa, K. Yamazaki, K. Suemasu, and K. Uno False-positive Findings on [18F]FDG-PET Caused by Non-neoplastic Cellular Elements After Neoadjuvant Chemoradiotherapy for Non-small Cell Lung Cancer Jpn. J. Clin. Oncol., May 1, 2005; 35(5): 271 - 273. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Nomori, K. Watanabe, T. Ohtsuka, T. Naruke, K. Suemasu, and K. Uno Visual and Semiquantitative Analyses for F-18 Fluorodeoxyglucose PET Scanning in Pulmonary Nodules 1 cm to 3 cm in Size Ann. Thorac. Surg., March 1, 2005; 79(3): 984 - 988. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||