J Thorac Cardiovasc Surg 2004;128:770-771
© 2004 The American Association for Thoracic Surgery
Brief communication |
a Cardiothoracic Unit, University Hospital of Wales, Heath Park, Cardiff, United Kingdom
b Wales Institute of Forensic Medicine, University Hospital of Wales, Heath Park, Cardiff, United Kingdom
Received for publication March 31, 2004; accepted for publication April 6, 2004.
* Address for reprints: Heyman Luckraz, FRCS, Cardiothoracic Unit, Block C5, University Hospital of Wales, Heath Park, Cardiff CF14 4XN, United Kingdom
HeymanLuckraz{at}aol.com
Since the experimental work of Koehnlein and Lemperle1 on the use of glue in operations, biologic glues have become part of the armamentarium used in the surgical management of patients undergoing repair of type A aortic dissection, especially where tissues are frail or where added hemostasis is required. It facilitates the reconstruction of the dissected aortic wall layers adjacent to the prosthetic graft. However, the use of tissue glue can be associated with significant risk, namely aortic wall necrosis leading to acute aortic redissection,2 coronary ostial stenosis, cerebral embolism,3 complete heart block, and prosthetic valvular dysfunction. We describe a case in which the glue embolized to the right and left coronary arteries, leading to fatal right ventricular infarction, and also demonstrate the usefulness of postmortem studies with the use of digital technologies for records keeping and educational purposes.
Clinical summary
A 74-year-old woman was referred to our care for emergency repair of type A aortic dissection. The procedure was carried out with deep hypothermic circulatory arrest. The operative findings included a morphologically normal aortic valve with mobile cusps and the proximal extent of the dissection occurring within the noncoronary sinus, with detachment of the noncoronaryright coronary commissure. A tear was also identified on the anteromedial aspect of the proximal arch adjacent to the orifice of the brachiocephalic artery. The false lumen was seen progressing beyond the arch, involving both its convexity and concavity.
The aortic valve was repaired with resuspension of the cusps. Bioglue (Cryolife Inc, Kennesaw, Ga) was injected in the false lumen in the noncoronary sinus and around the false lumen at the level of the sinotubular junction. The hemiarch was replaced with a 24-mm Hemashield gold vascular graft (Boston Scientific Medi-tech, Boston, Mass). Bioglue was inserted within the false lumen at the arch level, and the aortic layers were sealed together. After rewarming the patient, cardiopulmonary bypass was discontinued, with good hemodynamic parameters on 5 µg · kg1 · min1 of dopamine.
The patient made a slow but steady recovery postoperatively until the sixth postoperative day, when she had an episode of chest pain associated with being cold and clammy. She was tachycardic, tachypnoeic, and hypotensive, with electrocardiography showing signs of inferior infarction. An echocardiogram showed a significant pericardial collection, right ventricular akinesia, and moderate mitral regurgitation. She underwent emergency resternotomy, which confirmed 500 mL of bloody pericardial effusion and a distended, akinetic right ventricle. She did not respond to inotropic support, and resuscitative measures were abandoned.
A postmortem examination was requested, with consent for a limited examination of the heart and aorta and no retention of tissues or organs. This showed pale and soft areas within the right ventricular myocardium with mottled, hemorrhagic zones. Further macroscopic examination of the coronary tree revealed occlusion of the origin of the left circumflex artery caused by the Bioglue with distal propagated thrombus (Figure 1). Bioglue emboli were also identified within the right coronary artery and the diagonal branch of the left anterior descending artery. Bioglue from the patient's coronary arteries was compared with freshly prepared Bioglue from the glue tube by means of dissecting microscopy. This examination revealed that they were identical in appearance (amber, translucent, agar-like, solid fragments with crisp margins).
|
Several biologic sealants are now available for use in cardiovascular surgery, with the best studied being gelatin-resorcin-formalin (GRF) glue and a 2-component fibrin sealant. The use of GRF glue has been associated with tissue necrosis, with destruction of the aortic media causing vascular wall thinning and rupture.2 Its use has also been linked with coronary ostial stenosis and fatal cerebral embolism.3 Bioglue, which is composed of bovine serum albumin (45%) and gluteraldehyde (10%), is marketed as being less tissue toxic than GRF (gluteraldehyde 37%), and initial clinical experience was reported by Raanani and colleagues4 in 2001. Polymerization occurs when the 2 components are mixed, reaching bonding strength within 2 minutes. This tissue glue is also described as being more liquid in consistency than the GRF glue. Its potential to penetrate the intravascular circuit is thus higher. Thus it has the potential to embolize, as in the case described. The emboli then activates the clotting cascade directly and also produces stasis, which is one of the triad described by Virchow necessary for thrombus formation and propagation.
This case also illustrates the importance of the use of postmortem studies to elucidate the cause of death, as well as the use of digital imaging by the pathologists, to demonstrate the postmortem findings, obviating the need for tissue retention.5